Literature DB >> 31022318

Safety and reliability of a handheld stimulator for neural monitoring during thyroid surgery.

Bradley R Lawson1,2,3, Dipti Kamani1, Mohamed Shama1, Natalia Kyriazidis1, Gregory W Randolph1,4.   

Abstract

OBJECTIVE: The Checkpoint nerve stimulator (Checkpoint Surgical, Cleveland, OH) is a U.S. Food and Drug Administration-approved device for neural localization and monitoring during surgery. Its safety, efficacy, and reliability for neural monitoring during thyroid and parathyroid surgery have not been compared to more standard formats of neural monitoring. STUDY
DESIGN: Retrospective review.
METHODS: Vagal, recurrent, and superior laryngeal nerve monitoring were performed using both the Checkpoint stimulator and Medtronic NIM 3.0 laryngeal electromyography endotracheal tube (Medtronic, Jacksonville, FL) during thyroid and parathyroid surgery. A total of 21 operated sides in 15 patients were included for analysis. Latency and amplitude data for the Checkpoint stimulator were recorded using the NIM monitor and compared to normative endotracheal tube surface electrode data.
RESULTS: Mean amplitude using the Checkpoint stimulator was 574.6 microvolts (μV), 1060.6 μV, and 182.8 μV for the vagus, recurrent laryngeal, and superior laryngeal nerves, respectively. Mean amplitude using standard laryngeal electromyography was 709 μV, 1077.0 μV, and 183.7 μV for the same nerves. Mean latency was significantly shorter with stimulation of the recurrent laryngeal nerve compared to the vagus nerve with both stimulators (P < 0.001). No neural injuries occurred during the study.
CONCLUSION: The Checkpoint stimulator is a safe and reliable alternative to traditional laryngeal electromyography providing equivalent induced electromyography of the vocalis for neural monitoring during thyroid and parathyroid surgery. LEVEL OF EVIDENCE: 4 Laryngoscope, 2019.
© 2019 The American Laryngological, Rhinological and Otological Society, Inc.

Entities:  

Keywords:  Checkpoint stimulator; Endotracheal tube surface electrodes; electromyography; nerve stimulator; neural monitoring; parathyroid surgery; thyroid surgery

Year:  2019        PMID: 31022318     DOI: 10.1002/lary.28027

Source DB:  PubMed          Journal:  Laryngoscope        ISSN: 0023-852X            Impact factor:   3.325


  3 in total

1.  Killian-Jamieson diverticulum safely resected using a manual intraoperative neural monitoring system: a case report.

Authors:  Ryo Ataka; Shigeru Tsunoda; Saori Goto; Tatsuto Nishigori; Shigeo Hisamori; Kazutaka Obama; Yoshiharu Sakai
Journal:  Surg Case Rep       Date:  2020-02-26

Review 2.  Patient safety and quality improvements in parotid surgery.

Authors:  Vidit Talati; Hannah J Brown; Tasher Losenegger; Peter Revenaugh; Samer Al-Khudari
Journal:  World J Otorhinolaryngol Head Neck Surg       Date:  2022-04-27

3.  Serratus anterior dysfunction examination: wall push-up or shoulder flexion resistance test?

Authors:  Ryan Lohre; Bassem Elhassan
Journal:  JSES Int       Date:  2022-05-27
  3 in total