Stefan Kurath-Koller1, Sabrina Schweintzger2, Gernot Grangl2, Ante Burmas2, Andreas Gamillscheg2, Martin Koestenberger2.
Abstract
BACKGROUND: To report first clinical experience on three cases of congenital complete heart block and the use of a pacemaker system with a maximum lower rate interval of 95 beats per minute.Entities:
Keywords: Cardiac output; Children; Congenital heart block; Pacemaker
Mesh:
Year: 2019 PMID: 31018851 PMCID: PMC6480703 DOI: 10.1186/s12887-019-1494-7
Source DB: PubMed Journal: BMC Pediatr ISSN: 1471-2431 Impact factor: 2.125
Demographic data on infants with Kora pacemaker system
| Patient 1 | Patient 2 | Patient 3 | |
|---|---|---|---|
| a) Demographic Data | |||
| Sex | F | F | M |
| GA (wks + d) | 37 + 0 | 36 + 4 | 38 + 0 |
| BW (kg) | 2,8 | 3 | 2,9 |
| APGAR | 7/8/9 | 8/8/8 | 7/8/9 |
| Fetal HR (bpm) | 50 | 64 | 54 |
| pp HR (bpm) | 55 | 60 | 52 |
| pp AoVTI (cm) | 26,1 | 21,8 | 23,9 |
| Age at Implant (d) | 2 | 142 | 14 |
| Weight at Implant (kg) | 2,8 | 5,1 | 3,1 |
| Follow Up (a) | 2,9 | 2,2 | 2,3 |
| Echocardiography | unremarkable | unremarkable | unremarkable |
| b.) Initial programming | |||
| Model | Kora 100 SR | Kora 100 SR | Kora 250 SR |
| Mode | VVIR | VVIR | VVIR |
| Lower rate interval | 90 | 95 | 90 |
| Intrinsic rhythm (bpm) | 50 | 69 | 58 |
| Polarity | unipolar | unipolar | bipolar |
| Impedance (Ω) | 341 | 510 | 900 |
| R-wave (mV) | 7,9 | 7,75 | 15 |
| Threshold (V/ms) | 1,5/0,5 | 1,75/0,85 | 0,75/0,35 |
| Impulse (ms) | 0,5 | 0,85 | 0,35 |
| Amplitude (V) | 4 | 3,5 | 3 |
Leg.: F female, M male, GA gestational age, a years, wks weeks, d days, kg kilogram, bpm beats per minute, Ω Ohm, mV millivolt, ms milliseconds, V Volt, pp post partum, AoVTI aortic velocity time integral, HR heart rate, Implant Implantation
Fig. 1Circadian heart rate diagrams of our patients. Legend: x-axis shows time of day as hh:mm; y-axis shows heart rate as beats per minute. Diagrams were recorded at 2.5, 2 and 1.9 years respectively