| Literature DB >> 31016158 |
Sadeq G Kadem1, Majid H Alabbood2.
Abstract
AIM: Ultrasonic scalpel is one of the recent advances in surgical technology. The main limitation of their use is the cost. However, radiofrequency ablation device is a refined type of electrosurgical cautery that can be used with conventional reusable bipolar cautery forceps. This study aims to compare the efficacy of bipolar radiofrequency ablation device that can be used with conventional reusable bipolar cautery forceps for hemostasis and dissection during thyroidectomy with that of ultrasonic scalpel.Entities:
Keywords: Bipolar radiofrequency ablation device; comparative study suture less thyroidectomy; new energy devices; ultrasonic scalpel
Year: 2019 PMID: 31016158 PMCID: PMC6446675 DOI: 10.4103/ijem.IJEM_519_18
Source DB: PubMed Journal: Indian J Endocrinol Metab ISSN: 2230-9500
Figure 1Bipolar radiofrequency ablation device
Basic characteristics of patients
| Patients characteristics | US group ( | BRFAD group ( | |
|---|---|---|---|
| Mean age (y) ± SD | 40.27±10.05 | 40.46±9.74 | 0.460 |
| Gender (%) | |||
| Male | 5 (16.66%) | 6 (20%) | 0.500 |
| Female | 25 (83.33%) | 24 (80%) | 0.125 |
| Total | 30 (100%) | 30 (100%) | |
| Diagnosis (%) | |||
| Toxic goiter | 9 (30%) | 8 (26.66%) | 0.250 |
| Non-toxic goiter | 21 (70%) | 22 (73.33%) | 0.125 |
| Total | 30 (100%) | 30 (100%) | |
| Multinodular goiter | 17 (56.66%) | 18 (60%) | 0.125 |
| Diffuse goiter | 10 (33.33%) | 10 (33.33%) | 1.000 |
| Solitary nodule | 3 (10%) | 2 (6.66%) | 0.375 |
| Total | 30 (100%) | 30 (100%) | |
| Benign | 28 (93.33%) | 28 (93.33%) | 1.000 |
| Papillary cancer | 1 (3.33%) | 1 (3.33%) | 1.000 |
| Follicular cancer | 1 (3.33%) | 1 (3.33%) | 1.000 |
| Total | 30 (100%) | 30 (100%) | |
| Large goiter | 8 (26.66%) | 7 (23.33%) | 0.500 |
| Moderate goiter | 16 (53.33%) | 17 (56.66%) | 0.630 |
| Small goiter | 6 (20%) | 6 (20%) | 1.000 |
| Total | 30 (100%) | 30 (100%) | |
The outcome of procedure in both groups
| Outcome factors | US ( | BRFAD ( | |
|---|---|---|---|
| Mean surgery length (min) ± SD | 65.49±7.78 | 50.13±14.16 | 0.001 |
| Mean hospitalization stay (hr) ± SD | 24.25±2.44 | 24.75±4.197 | 0.320 |
| Mean drain removal duration (hr) ± SD | 24.50±3.44 | 25.00±4821 | 0.417 |
| Mean postoperative drainage volume per drain (ml) ± SD | 63.02±19.91 | 62.50±22.79 | 0.330 |
Technical and financial observations during the study
| Technical and Financial Observations | BRFAD | US |
|---|---|---|
| Approximate cost of the generatora | $11.000 | $30.000 |
| Approximate cost of the hand piece* | $100 (reusable) | $900 (disposable) |
| Size of the jaw of the hand piece | Fine tip, suitable for fine tissues dissection | Relatively large tip, not suitable for fine tissues dissection |
| Heat at the jaw of the hand piece | Remain cold even during firing | Remain hot even after stop firing |
aAccording to the local sale price in Basrah, Iraq