Literature DB >> 31013489

Chronic recording and electrochemical performance of amorphous silicon carbide-coated Utah electrode arrays implanted in rat motor cortex.

Alexandra Joshi-Imre1, Bryan J Black1, Justin Abbott1, Aswini Kanneganti1, Rashed Rihani1, Bitan Chakraborty1, Vindhya R Danda1, Jimin Maeng1, Rohit Sharma2, Loren Rieth2, Sandeep Negi3, Joseph J Pancrazio1, Stuart F Cogan1.   

Abstract

OBJECTIVE: Clinical applications of implantable microelectrode arrays are currently limited by device failure due to, in part, mechanical and electrochemical failure modes. To overcome this challenge, there is significant research interest in the exploration of novel array architectures and encapsulation materials. Amorphous silicon carbide (a-SiC) is biocompatible and corrosion resistant, and has recently been employed as a coating on biomedical devices including planar microelectrode arrays. However, to date, the three-dimensional Utah electrode array (UEA) is the only array architecture which has been approved by the food and drug administration (FDA) for long-term human trials. APPROACH: Here, we demonstrate, for the first time, that UEAs can be fabricated with a-SiC encapsulation and sputtered iridium oxide film (SIROF) electrode coatings, and that such arrays are capable of single-unit recordings over a 30 week implantation period in rat motor cortex. Over the same period, we carried out electrochemical measurements, including voltage transients, cyclic voltammetry, and electrochemical impedance spectroscopy (EIS), to evaluate potential failure modes. Furthermore, we evaluated chronic foreign body response via fluorescence immunohistochemistry following device explantation. MAIN
RESULTS: During the indwelling period, we observed a reduction in active electrode yield percentage from 94.6  ±  5.4 (week 1) to 16.4  ±  11.5% (week 30). While the average active electrode yield showed a steady reduction, it is noteworthy that 3 out of 8 UEAs recorded greater than 60% active electrode yield at all times through 24 weeks and 1 out of 8 UEAs recorded greater than 60% active electrode yield at all times through the whole implantation period. SIGNIFICANCE: In total, these findings further suggest that a-SiC may serve as a mechanically and electrochemically stable device encapsulation alternative to polymeric coatings such as Parylene-C.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31013489     DOI: 10.1088/1741-2552/ab1bc8

Source DB:  PubMed          Journal:  J Neural Eng        ISSN: 1741-2552            Impact factor:   5.379


  3 in total

1.  In Vitro/Ex Vivo Investigation of Modified Utah Electrode Array to Selectively Sense and Pace the Sub-Surface Cardiac His Bundle.

Authors:  Ankur R Shah; Muhammad S Khan; Annie M Hirahara; Matthias Lange; Ravi Ranjan; Derek J Dosdall
Journal:  ACS Biomater Sci Eng       Date:  2020-05-07

Review 2.  A Review: Electrode and Packaging Materials for Neurophysiology Recording Implants.

Authors:  Weiyang Yang; Yan Gong; Wen Li
Journal:  Front Bioeng Biotechnol       Date:  2021-01-14

3.  Focal electrical stimulation on an alcohol disorder model using magnetic resonance imaging-compatible chronic neural monopolar carbon fiber electrodes.

Authors:  Alejandra Lopez-Castro; Diego Angeles-Valdez; Gerardo Rojas-Piloni; Eduardo A Garza-Villarreal
Journal:  Front Neurosci       Date:  2022-09-29       Impact factor: 5.152

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.