| Literature DB >> 31011608 |
Somiya Lamrabat1,2, Amina Lyagoubi1,2, Ouardia Bouayadi1,2, Jalila Elmalki1,2, Khadija Elkouche1,2, Mohammed Bensalah1,2, Mohammed Choukri1,2.
Abstract
BACKGROUND: The aim of this work is to present the results of a comparative study between the ALT assay on Integra 400 plus Roche Diagnostic versus Architect ci8200 of Abbott Diagnostic.Entities:
Keywords: Abbott diagnostic; Alanine aminotransferase; Comparison; Concordance; Roche diagnostic
Year: 2019 PMID: 31011608 PMCID: PMC6462767 DOI: 10.1016/j.plabm.2019.e00121
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Comparison of the characteristics of the two methods for the determination of Alanine aminotransferase. (ND: not disclosed).
| Architect ci8200 | Integra 400 plus | |
|---|---|---|
| Principle | Spectrophotometric | Spectrophotometric |
| Analytical target | ALT | ALT |
| Reagent stability if it’s uncapped and onboard | 27 days | 12 weeks |
| Samples | Serum, plasma | Serum, plasma |
| Measuring range (U/L) | 2–942 | 5–700 |
| Reference value (U/L) | 0–55 | Man: 0-41 |
| Intra-serial precision (CV [%]) | 1.5 | ND |
| Total precision (CV [%]) | 3.6 | 3.2 |
| Dosing time (min) | 9 | 10 |
| Certificate of analysis | YES | YES |
| Ec declaration of conformity | YES | YES |
| Safety data sheet | YES | YES |
Results of Pearson’s correlation analysis.
| Sample size, n | 200 |
| Correlation coefficient r | 0,9998 |
| 95% confidence interval for r | 0,9998 to 0,9999 |
| P- value | <0,0001 |
Fig. 1Bland-Altman diagram of the difference between Architect ci8200 and Integra 400 plus in ALT assay.
Results of the regression analysis. (NA: not applied).
| Passing-Bablok | Deming | |
|---|---|---|
| Intercept A | −0,5625 | −0,7503 |
| 95% CI | −0,6667 to −0,3563 | −0,9450 to −0,5555 |
| SlopeB | 0,9917 | 0,9938 |
| 95% CI | 0,9875 to 0,9952 | 0,9897 to 0,9979 |
| Cusum de linéarité | 0,07 | NA |
Fig. 2Correlation diagram according to Passing and Bablok. N = 200 (Intersection: -0,5625 Slope: 0,9917).