AIM: This study aims to propose protocols that enable scoring of mitral valve (MV) in mitral stenosis using the three-dimensional (3D) scoring system.Entities:
Keywords: Mitral stenosis; mitral valve score; real-time three-dimensional echocardiography
Year: 2019 PMID: 31008031 PMCID: PMC6450232 DOI: 10.4103/jcecho.jcecho_58_18
Source DB: PubMed Journal: J Cardiovasc Echogr ISSN: 2211-4122
Figure 1Orientation of mitral valve leaflets (arrows) from its ventricular aspect. Images was rotated to display the aortic valve superiorly to get the optimal anatomical orientation of both leaflets
Figure 2Quad screen of multibeat live three-dimensional image acquired from apical long axis for the assessment of both mitral valve leaflets. The upper 2 images (a and b) are the two-dimensional orthogonal planes created by the three-dimensional probe. The lower left image (c) is the two-dimensional short axis of mitral valve and the lower right view (d) is the corresponding three-dimensional short axis of mitral valve
Figure 3En face view of the proximal third of chordae obtained by zoom three-dimensional images from parasternal long axis showed the separation between the anterior and posterior chords (arrows)
Figure 4Quad screen of cropping of full-volume image obtained from apical long axis using iCrop function for assessment and scoring of the proximal third of chordae. The upper 2 images (a and b) showed the adjustment of box size and position to include the target. The lower left image (c) is the short axis of the proximal third of chordae (arrows) and the lower right image (d) is the corresponding three-dimensional image in short axis
Figure 5Multibeat live three-dimensional image acquired from apical long axis for the assessment of chordal thickness and separation (blue arrows)
Categorization of three-dimensional images in all views using all three-dimensional modalities for scoring of all mitral valve scallops according to the percentage of patients in whom compete assessment was achieved
| 3D modality | MV leaflets | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Apical | PLAX | PSAX | |||||||
| A1/P1 | A2/P2 | A3/P3 | A1/P1 | A2/P2 | A3/P3 | A1/P1 | A2/P2 | A3/P3 | |
| Zoom 3D (%) | |||||||||
| Single beat | I (100) | I (100) | I (100) | III (48) | I (100) | II (86) | I (91) | I (91) | I (91) |
| Multibeat | I (100) | I (100) | I (100) | III (30) | I (100) | II (86) | I (100) | I (100) | I (100) |
| Full volume (%) | |||||||||
| Single beat | II (88.6) | II (88.6) | II (88.6) | II (88.6) | II (88.6) | II (88.6) | III (26) | III (26) | III (26) |
| Multibeat | I (91) | I (91) | I (91) | I (91) | I (91) | I (91) | III (41) | III (41) | III (41) |
| Live 3D (%) | |||||||||
| Single beat | III (0) | III (0) | III (0) | III (0) | III (0) | III (0) | III (17) | III (17) | III (17) |
| Multibeat | III (0) | III (0) | III (0) | III (0) | III (0) | III (0) | I (91) | I (91) | I (91) |
I=Image of optimal assessment, II=Image of suboptimal assessment, III=Image of poor assessment, PSAX=Parasternal short axis, PLAX=Parasternal long axis, 3D=three-dimensional, MV=Mitral valve
Categorization of three-dimensional images in all views using all three-dimensional modalities for scoring of chordal thickness and separation at three levels (proximal, middle, and distal) according to the percentage of patients in whom compete assessment was achieved
| 3D modality | Chordal thickness | Chordal separation | ||||||
|---|---|---|---|---|---|---|---|---|
| Apical | PLAX | Apical | PLAX | |||||
| Proximal and middle | Distal | Proximal and middle | Distal | Proximal and middle | Distal | Proximal and middle | Distal | |
| Zoom 3D (%) | ||||||||
| Single beat | I (91) | I (91) | I (100) | III (0) | I (91) | I (91) | I (100) | III (0) |
| Multibeat | I (100) | I (100) | I (100) | III (0) | I (100) | I (100) | I (100) | III (0) |
| Full volume (%) | ||||||||
| Single beat | I (100) | I (100) | I (100) | III (0) | I (100) | I (100) | I (100) | III (0) |
| Multibeat | I (100) | I (100) | I (100) | III (0) | I (100) | I (100) | I (100) | III (0) |
| Live 3D (%) | ||||||||
| Single beat | II (85.7) | II (85.7) | II (86) | III (0) | III (0) | III (0) | III (0) | III (0) |
| Multibeat | I (91) | I (91) | I (100) | III (0) | III (0) | III (0) | III (0) | III (0) |
I=Image of optimal assessment, II=Image of suboptimal assessment, III=Image of poor assessment, PLAX=Parasternal long axis, 3D=three-dimensional
List of three-dimensional images acquired by all three-dimensional echo modalities with the comparison in-between (acquisition time, analysis time, and image resolution)
| 3D modality | Total ( | Acquisition time (s) | Analysis time (s) | Total time (s) | Resolution (Hz) |
|---|---|---|---|---|---|
| Single-beat zoom 3D | 455 (29.1) | 5±2 | 11±2 | 17±4 | 21±2 |
| Multibeat zoom 3D | 299 (19.1) | 10±2 | 11±2 | 21±4 | 72±5 |
| Single-beat full volume | 455 (29.1) | 5±2 | 25±5 | 30±7 | 6±2 |
| Multibeat full volume | 299 (19.1) | 11±3 | 27±4 | 38±7 | 24±4 |
| Single-beat live 3D | 35 (2.2) | 5±2 | 10±3 | 15±5 | 23±3 |
| Multibeat live 3D | 23 (1.5) | 11±3 | 10±3 | 21±6 | 75±9 |