Ersin Çağrı Şimşek1, Ece Uslu Güvendi2, Aylin Şimşek3, Uğur Kocabaş4, Eser Varış4, Doğaç Çağlar Gürbüz4, Ayhan Aşkin2.
Abstract
BACKGROUND: Although cardiac implantable electronic device (CIED) implantation is considered to be minor surgery, almost 60% of the patients suffer from shoulder-related problems a short time after the procedure. The purpose of this study was to determine the possible effects of the preference of the dominant side for CIED implantation on the ipsilateral superior extremity functions.Entities:
Keywords: arrhythmia; cardiac implantable electronic device; dominant hand side; shoulder impairment; upper extremity disabilities
Year: 2019 PMID: 31007794 PMCID: PMC6457371 DOI: 10.1002/joa3.12156
Source DB: PubMed Journal: J Arrhythm ISSN: 1880-4276
Characteristics of the patients and the cardiac implantable electronic devices
| Ipsilateral dominant‐hand group (IDH), n = 55 | Contralateral dominant‐hand group (CDH), n = 52 |
| |
|---|---|---|---|
| Age (year) | 70 ± 12 | 68 ± 12 | 0.063 |
| Gender(M/F) | 24/31 (43.6%/56.4%) | 19/33 (36.5%/63.5%) | 0.084 |
| Body mass index | 28.9 ± 5.12 | 27.3 ± 3.74 | 0.137 |
| Smoke | 15 (27.2%) | 19 (36.5%) | 0.058 |
| Alcohol | 9 (16.3%) | 6 (11.5%) | 0.626 |
| Diabetes mellitus | 13 (23.6%) | 15 (28.8%) | 0.655 |
| NYHA class | |||
| I‐II | 47 (85.5%) | 40 (77%) | 0.131 |
| III‐IV | 8 (14.5%) | 12 (23%) | |
| CIED type | |||
| ICD | 9 (16.3%) | 41 (78.8%) | <0.001 |
| Pacemaker | 46 (83.7%) | 11 (21.2%) | |
| CIED mode | |||
| VVI | 16 (29.0%) | 30 (61.4%) | <0.001 |
| VDD | 14 (25.4%) | 3 (4.5%) | |
| DDD | 25 (45.6%) | 19 (34.1%) | |
| Number of replacements | |||
| None | 40 (72.7%) | 40 (76.9%) | 0.146 |
| 1 | 12 (21.8%) | 10 (19.2%) | |
| 2< | 3 (5.4%) | 2 (3.9%) | |
| Dominant hand (R/L) | 50/5 (90.9/9.1%) | 46/6 (88.4/11.6%) | |
| Weight of battery (g) | 27 (18‐79) | 79 (18‐92) | <0.001 |
| Battery volume (cc) | 12.1 (8.5‐42.0) | 42 (8.5‐42) | <0.001 |
| Battery‐clavicula distance (cm) | 7.0 ± 2.2 | 6.9 ± 3.0 | 0.419 |
| Incision length (cm) | 4.7 ± 1.6 | 5.6 ± 1.4 | 0.002 |
| Time duration since last implantation (months) | 36.8 ± 22 | 25.8 ± 18 | 0.029 |
| Prolonged immobilization after procedure | 17 (30.9%) | 17 (38.6%) | 0.413 |
| Postprocedure complications | 5 (10.0%) | 4 (9.1%) | |
| Lead complications (dislocation‐extraction) | 1 (1.8%) | 2 (3.6%) | |
| Pocket hematoma | 2 (3.6%) | 1 (1.8%) | |
| Pocket infection | 2 (3.6%) | 1 (1.8%) | |
NYHA, New York heart association; L, left; R, right; CIED, cardiac implantable electronic device; ICD, implantable cardioverter‐defibrillators; VVI, ventricular pacing, ventricular sensing, inhibition; VDD, ventricular pacing, dual sensing, dual response to sensing; DDD, dual pacing, dual sensing, dual response to sensing.
Characteristics of patients and cardiac implantable electronic devices after the patients are divided into two groups according to the presence of disability of the upper extremity
| Patients with disability of the affected extremity, n = 17 | Patients without a disability of the affected extremity, n = 90 | |
|---|---|---|
| Age (year) | 63.0 ± 11.8 | 63.9 ± 12.8 |
| Gender(M/F) | 5/12 (29.4/70.6) | 35/55 (38.8/61.2) |
| Ejection fraction (%) | 37.3 ± 15.9 | 45.0 ± 15.5 |
| CIED type | ||
| ICD | 9 (53) | 41 (45.5) |
| Pacemaker | 8 (47) | 49 (54.5) |
| Number of replacements | ||
| None | 14 (82.4) | 66 (73.4) |
| 1 | 2 (11.8) | 20 (22.2) |
| 2< | 1 (5.8) | 4 (4.4) |
| Side of CIED | ||
| Right | 7 (41.1) | 51 (56.6) |
| Left | 10 (59.9) | 39 (43.4) |
| CIED mode | ||
| VVI | 10 (58.8) | 40 (44.4) |
| VDD | 0 | 14 (15.6) |
| DDD | 7 (41.2) | 36 (40) |
| Number of leads | ||
| 1 | 10 (58.8) | 51 (56.7) |
| 2 | 7 (41.2) | 39 8 (43.3) |
| Weight of battery (g) | 57.9 ± 28.3 | 46.9 ± 26.3 |
| Battery volume (cc) | 29.5 ± 15.2 | 23.9 ± 12.1 |
| Battery‐clavicula distance (cm) | 2.75 ± 0.4 | 2.76 ± 0.4 |
| Incision length (cm) | 4.93 ± 1.3 | 5.19 ± 1.6 |
| Time duration since last implantation (months) | 28.8 ± 21.2 | 34.2 ± 24.3 |
CIED, cardiac implantable electronic device; ICD, implantable cardioverter‐defibrillators; VVI, ventricular pacing, ventricular sensing, inhibition; VDD, ventricular pacing, dual sensing, dual response to sensing; DDD, dual pacing, dual sensing, dual response to sensing.
Functional assessments of the shoulder joint in ipsilateral and contralateral dominant recipients of cardiac implantable electronic devices
| Ipsilateral dominant group (IDH), n = 55 | Contralateral dominant group (CDH), n = 52 |
| |
|---|---|---|---|
| VAS score | |||
| VAS‐rest | 0.384 ± 1.49 | 0.944 ± 1.72 | 0.100 |
| VAS‐activity | 1.307 ± 2.45 | 1.444 ± 2.51 | 0.552 |
| VAS‐sleep | 0.871 ± 2.16 | 0.972 ± 2.26 | 0.957 |
| QuickDASH score | 4.5 (0‐79.5) | 6.8 (0‐90) | 0.212 |
| JAMAR score (kg) | |||
| Right | 34 (16‐95) | 40 (24‐85) | 0.02 |
| Left | 32 (15‐85) | 36 (15‐115) | 0.106 |
| Sulcus sign | 14 (25.4%) | 11 (21.1%) | 0.528 |
| Peri‐incisional paresthesia | 17 (30.9%) | 13 (25%) | 0.565 |
VAS, Visual analogue scale; QuickDASH, quick disability of the arm shoulder and hand questionnaire.
Comparison of maximum isometric grip strength of upper extremities between right and left‐side CIEDs among the right‐dominant recipients
| Right‐side CIED, n = 50 |
| Left‐side CIED, n = 44 |
| Between right‐side and left‐side recipients | |
|---|---|---|---|---|---|
| Jamar‐right (kg) | 34 (12‐95) | 0.056 | 40 (22‐85) | 0.030 | 0.044 |
| Jamar‐left (kg) | 32 (12‐85) | 36 (15‐115) | 0.166 |
CIED, cardiac implantable electronic device.
Indicates the difference between right and left hand in right‐side CIEDs recipients.
Indicates the difference between right and left hand in left‐side CIEDs recipients.
Indicates the difference in right hand between the right‐side and left‐side recipients.
Indicates the difference in left hand between the right‐side and left‐side recipients.
Degrees of reduced range of motion in the shoulder joint in the arm on the side of the cardiac implantable electronic devices compared with contralateral side
| Ipsilateral dominant group (IDH), n = 55 | Contralateral dominant group (CDH), n = 52 | |||||||
|---|---|---|---|---|---|---|---|---|
| ∆ degrees 1‐10 | ∆ degrees 11 ‐20 | ∆ degrees 21‐30 | ∆ degrees ≥‐31 | ∆ degrees 1‐10 | ∆ degrees 11 ‐20 | ∆ degrees 21‐30 | ∆ degrees ≥‐31 | |
| Flexion | 1 (2%) | 2 (4%) | 0 | 2 (4%) | 2 (4%) | 1 (2%) | 0 | 3 (6%) |
| Extension | 1 (2%) | 0 | 0 | 0 | 1 (2%) | 0 | 0 | 0 |
| Abduction | 2 (4%) | 1 (2%) | 0 | 2 (4%) | 1 (2%) | 1 (2%) | 2 (4%) | 3 (6%) |
| Adduction | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Internal rotation | 1 (2%) | 0 | 0 | 1 (2%) | 1 (2%) | 1 (2%) | 0 | 1 (2%) |
| External rotation | 0 | 1 (2%) | 0 | 1 (2%) | 1 (2%) | 1 (2%) | 0 | 1 (2%) |
∆; difference of range of motion between the arms.