Literature DB >> 31005179

Ultra-low dose chest computed tomography: Effect of iterative reconstruction levels on image quality.

Mercy Afadzi1, Elisabeth Kirkeby Lysvik2, Hilde Kjernlie Andersen2, Anne Catrine T Martinsen3.   

Abstract

PURPOSE: To optimize image quality and radiation dose of chest CT with respect to various iterative reconstruction levels, detector collimations and body sizes.
METHOD: A Kyoto Kagaku Lungman with and without extensions was scanned using fixed ultra-low doses of 0.25, 0.49 and 0.74 mGy CTDIvol, and collimations of 40 and 80 mm. Images were reconstructed with the lung kernel, filtered back projection (FBP) and different ASIR-V levels (10-100%). Contrast-to-noise ratios (CNR) were calculated for 12 mm simulated lesions of different densities in the lung. Image noise, signal-to-noise ratios (SNR), variations in Hounsfield units (HU), noise power spectrum (NPS) and noise texture deviations (NTD) were evaluated for all reconstructions. NTD was calculated as percentage of pixels outside 3 standard deviations to evaluate IR-specific artefacts.
RESULTS: Compared to the FBP, image noise reduced (5-55%) with ASIR-V levels irrespective of dose or collimation. SNR correlated positively (r ≥ 0.925, p ≤ 0.001) with ASIR-V levels at all doses, collimations, and phantom sizes. ASIR-V enhanced the CNR of the lesion with the lowest contrast from 12.7-42.1 (0-100% ASIR-V) at 0.74 mGy with 40 mm collimation. As expected, higher SNR and CNR were measured in the smaller phantom than the bigger phantom. Uniform HU were observed between FBP and ASIR-V levels at all doses, collimations, and phantom sizes. NPS curves left-shifted towards lower frequencies at increasing levels of ASIR-V irrespective of collimation. A positive correlation (r ≥ 0.946, p ≥ 0.001) was observed between NTD and ASIR-V levels. NTD of the FBP was not significantly (p ≤ 0.087) different from NTD of ASIR-V ≤ 20%. The data from the NPS and NTD indicates a blotchier and coarser noise texture at higher levels of ASIR-V, especially at 100% ASIR-V.
CONCLUSION: In comparison with the FBP technique, ASIR-V enhanced quantitative image quality parameters at all ultra-low doses tested. Moreover, the use of ASIR-V showed consistency with body size and collimation. Hence, ASIR-V may be useful for improving image quality of chest CT at ultra-low doses.
Copyright © 2019 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  ASIR-V; Chest CT; Image quality; Iterative reconstruction; Ultra-low dose

Mesh:

Year:  2019        PMID: 31005179     DOI: 10.1016/j.ejrad.2019.02.021

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  5 in total

Review 1.  Quantitative inspiratory-expiratory chest CT to evaluate pulmonary involvement in pediatric hematopoietic stem-cell transplantation patients.

Authors:  Francesca Pennati; Laura L Walkup; Anuj Chhabra; Christopher Towe; Kasiani Myers; Andrea Aliverti; Jason C Woods
Journal:  Pediatr Pulmonol       Date:  2020-12-23

2.  Optimization of imaging parameters in chest CT for COVID-19 patients: an experimental phantom study.

Authors:  Yantao Niu; Shunxing Huang; Huan Zhang; Shuo Li; Xiaoting Li; Zhibin Lv; Shuo Yan; Wei Fan; Yanlong Zhai; Eddy Wong; Kexin Wang; Zongrui Zhang; Budong Chen; Ruming Xie; Junfang Xian
Journal:  Quant Imaging Med Surg       Date:  2021-01

3.  Effect of adaptive statistical iterative reconstruction-V (ASiR-V) levels on ultra-low-dose CT radiomics quantification in pulmonary nodules.

Authors:  Kai Ye; Min Chen; Qiao Zhu; Yuliu Lu; Huishu Yuan
Journal:  Quant Imaging Med Surg       Date:  2021-06

4.  Image Quality and Pulmonary Nodule Detectability at Low-dose Computed Tomography (low kVp and mAs): A phantom study.

Authors:  Sepideh Iranmakani; Amir Reza Jahanshahi; Parinaz Mehnati; Tohid Mortezazadeh; Davood Khezerloo
Journal:  J Med Signals Sens       Date:  2021-12-28

5.  Three-stage segmentation of lung region from CT images using deep neural networks.

Authors:  Michael Osadebey; Hilde K Andersen; Dag Waaler; Kristian Fossaa; Anne C T Martinsen; Marius Pedersen
Journal:  BMC Med Imaging       Date:  2021-07-15       Impact factor: 1.930

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.