| Literature DB >> 30996670 |
O Azzolini1, M T Barrera1, J W Beeman2, F Bellini3,4, M Beretta5,6, M Biassoni6, C Brofferio5,6, C Bucci7, L Canonica7,8,9, S Capelli5,6, L Cardani4, P Carniti5,6, N Casali4, L Cassina5,6, M Clemenza5,6, O Cremonesi6, A Cruciani3,4, A D'Addabbo7, I Dafinei4, S Di Domizio10,11, F Ferroni3,4, L Gironi5,6, A Giuliani12,13, P Gorla7, C Gotti5,6, G Keppel1, M Martinez3,4, S Morganti4, S Nagorny7,14, M Nastasi5,6, S Nisi7, C Nones15, D Orlandi7, L Pagnanini7,14, M Pallavicini10,11, V Palmieri1, L Pattavina7,14, M Pavan5,6, G Pessina6, V Pettinacci3,4, S Pirro7, S Pozzi5,6, E Previtali6, A Puiu5,6, F Reindl4,16,17, C Rusconi7,18, K Schäffner7,14, C Tomei4, M Vignati4, A Zolotarova15.
Abstract
The CUPID-0 detector hosted at the Laboratori Nazionali del Gran Sasso, Italy, is the first large array of enriched scintillating cryogenic detectors for the investigation of 82 Se neutrinoless double-beta decay ( 0 ν β β ). CUPID-0 aims at measuring a background index in the region of interest (RoI) for 0 ν β β at the level of 10 - 3 counts/(keV kg years), the lowest value ever measured using cryogenic detectors. CUPID-0 operates an array of Zn 82 Se scintillating bolometers coupled with bolometric light detectors, with a state of the art technology for background suppression and thorough protocols and procedures for the detector preparation and construction. In this paper, the different phases of the detector design and construction will be presented, from the material selection (for the absorber production) to the new and innovative detector structure. The successful construction of the detector lead to promising preliminary detector performance which is discussed here.Entities:
Year: 2018 PMID: 30996670 PMCID: PMC6435222 DOI: 10.1140/epjc/s10052-018-5896-8
Source DB: PubMed Journal: Eur Phys J C Part Fields ISSN: 1434-6044 Impact factor: 4.590
Fig. 1Schematic view of a single module scintillating bolometer
Fig. 2Rendering of CUPID-0 detector and its single module
Internal radioactive contamination for 2.5 kg of 96.3% enriched Se metal beads and for 2.5 kg of Zn. Limits are computed at 90% C.L.. The measurements were carried out on October 2014
| Chain | Nuclide | ||
|---|---|---|---|
| [ | [ | ||
| < 61 | < 95 | ||
| < 110 | < 36 | ||
| < 110 | < 66 | ||
| < 6200 | < 6200 | ||
| < 3400 | < 4700 | ||
| < 74 | < 91 | ||
| < 990 | < 380 | ||
| < 65 | < 36 | ||
| – | 80 ± 20 | ||
| – | 5200 ± 600 |
Fig. 3Picture of a ZnSe ingot as grown. The two edge of the boule are removed, while the central part is processed for the realization of the final crystal for the CUPID-0 detector
Fig. 4Crystal masses. The Se content for each crystal is shown. Crystal number 4 and 20 are crystal with natural Se isotopic abundance
Fig. 5A high purity Ge slab before applying any procedure
Fig. 6Light detector before its assembly. The side with the antireflective coating is visible. The coating results in a dark internal circle, while 2 mm on the edge are uncoated
Bulk contaminations of the VIKUITI-3M light reflector. The measurement was carried out at the LNGS by means of a mineralization procedure and an ICPMS analysis. The overall mass of VIKUITI-3M used in CUPID-0 is 17 g. The total activity is 0.8 ± 0.2 Bq and 2.9 ± 0.7 Bq for Th and U, respectively
| Chain | Nuclide | VIKUITI-3M activity |
|---|---|---|
| [ | ||
| 49 ± 12 | ||
| 170 ± 50 |
Fig. 7Six out of 9 Ge wafers, diameter of 65.5 and thickness 3.25 mm, installed in an Al holder before neutron irradiation at the MIT Nuclear Reactor Laboratory (MA, USA)
Fig. 8Doped Ge wafer with a 4000 gold deposition. The Au on the two large surface of the wafers serve as Ohmic contacts for reading-out the sensor
Fig. 9Ge Neutron Transmutation Doped sensors for ZnSe (right) and light detector (left). The dimensions for the two sensors are 3 2.8 1 mm and 2 2.8 0.5 mm, respectively
Fig. 10Sensor gluing: (1) Matrix of spring-loaded tips used for the 3 2.8 1 mm NTDs. (2) Placing NTD on the positioning device and held by vacuum. The Spring-loaded tip matrix is lowered on the NTD after dipping in the glue. After that, (3) the ZnSe crystal is placed on the top of the NTD and it is kept in place by a PTFE holder
Fig. 11Photos of the two working areas inside the cleanroom. On the left is shown the station were the Ge-NTD thermal sensors are glued to the absorbers, and on the right is shown the area where the tower are assembled
Radioactive contamination of the polymer-resin used for the 3D-printing of the detector assembling tools (3SP WHITE D7)
| Chain | Nuclide | Activity |
|---|---|---|
| [mBq/kg] | ||
| < 9.3 | ||
| < 10.3 | ||
| < 3.8 | ||
| < 73 | ||
| < 0.25 | ||
| < 5.3 | ||
| 81±35 |
Fig. 12Light detectors assembly: (1) Positioning of the Ge wafer with the PTFE O-shape holders on the mounting template. (2) Positioning of the copper holder and connection of the Ge-NTD gold wires for the detector read-out. (3) Storage of the assembled light detectors in a vacuum box. (4) Fit and tighten of the handling tool and removal from the mounting template
Fig. 13Single tower assembly: a positioning of the first pre-assembled LD on the bottom copper holder of the tower; installation of the first three columns that will host the ZnSe crystal. b Positioning of the ZnSe crystal on the bottom copper frame equipped with three S-shaped PTFE clamps. c Positioning of the VIKUITI-3M reflective foil. d Positioning of the top copper holder equipped with three S-shaped PTFE clamps; connection of the Ge-NTD gold wires in the inner copper pins. e Positioning of the top pre-assembled LD. f Coupling of the pre-assembled LD with the top ZnSe copper holder
Fig. 14CUPID-0 detector installed on the cryogenic system, just below the 10 cm thick Roman Pb shielding
Fig. 15Dilution unit photo. The different thermalization stages are identified by the arrows. A Roman Pb shield is hung on the MC stages by means of a vibrational damping system. The junction board is installed on the Pb stage, this is shown in Fig. 17
Fig. 17Junction boards on the MC for connecting the NbTi-NOMEX cables to the constantan wires from the detectors
Fig. 16NbTi-NOMEX cable from 300 K to the mixing chamber. On the room temperature side they are soldered to Fischer 27-pin connectors on the other end they are soldered to customized Cu-Kapton Zero Insertion Force connectors
Fig. 18Rendering of the double stage mechanical decoupling system. This is installed directly on the cryostat mixing chamber with the top brass anchor and on the bottom part there is a Cu mechanical connection for the detector installation
Fig. 19Block diagram of a readout channel, from the detector to the DAQ
Fig. 20Detectors RMS noise at 5 Hz as a function of the Ge-NTD dynamic impedance for ZnSe (a) and LD (b)
Fig. 21Distribution of the Ge-NTD sensor resistances for ZnSe (a) and LD (b) at the operating conditions
Fig. 22Characteristic load curve of a CUPID-0 ZnSe crystal operated with a Ge-NTD thermal sensor. The figure shows how varying the detector biasing voltage acts on the signal amplitude (red) and on the voltage drop across the sensor resistance (blue)
Fig. 23Characteristic load curve of a CUPID-0 ZnSe crystal operated with a Ge-NTD thermal sensor. The figure shows how a Ge-NTD stands high power dissipation without affecting the sensor operational conditions
Fig. 24Signal-to-noise ratio (SNR) scan for LDs (top) and ZnSe crystals (bottom) varying the voltage bias. It is calculated as the ratio of the filtered pulser amplitude to the . V(Max Amplitude) represents the biasing voltage which gives the maximal pulse amplitude. The scan is performed at bias higher than V(Max Amplitude) because we expect a stronger reduction of the noise compared to the signal amplitude
Fig. 25Distribution of the ZnSe energy resolutions. The FHWM resolution at 0 keV, which is defined as the detector baseline noise, in shown in (a). While the distribution for the 2615 keV -line energy is shown in (b)
Fig. 26Distribution of the ZnSe signal amplitudes for each crystal
Fig. 27Distribution of the LD SNRs. The signal amplitude is evaluated on test pulses generated by a current pulse through a Si resistor coupled to each LD. The amount of dissipated energy is the same for each detector
Summary of the main operating parameters of the CUPID-0 detectors. For the LDs only the median values are reported, given the small spread in the performance. Three detectors (Channel ID 3, 4, 8) have a reduced signal amplitude which prevented us from evaluating the detector energy resolution.
| Channel ID | Name | Tower | Type | Mass | R | R | Noise@5Hz | Signal Amplitude | Rise time | Decay time | FWHM | FWHM |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| [g] | [M | [M | [nV/ | [ | [ms] | [ms] | [keV] | [keV] | ||||
| 1 | CG-01 | 1 | Enriched | 439.40 | 10 | 4.17 | 11.7 | 40 | 10.1 | 24.7 | 2.59 | 21 |
| 2 | CG-13 | 1 | Enriched | 427.86 | 54 | 14.71 | 25.7 | 81 | 12.0 | 37.6 | 2.58 | 35 |
| 3 | CG-28 | 1 | Enriched | 427.00 | 29 | 9.63 | 18.1 | - | 9.5 | 17.3 | – | – |
| 4 | NAT-1 | 1 | Natural | 418.39 | 34 | 11.28 | 26.3 | - | 5.5 | 20.7 | – | – |
| 5 | CG-26 | 1 | Enriched | 408.22 | 7 | 3.94 | 20.9 | 51 | 11.1 | 33.1 | 3.22 | 19 |
| 6 | CG-02 | 2 | Enriched | 441.29 | 16 | 4.90 | 19.3 | 46 | 9.3 | 23.6 | 3.86 | 22 |
| 7 | CG-15 | 2 | Enriched | 469.64 | 25 | 11.25 | 17.2 | 47 | 12.6 | 26.2 | 2.88 | 20 |
| 8 | CG-29 | 2 | Enriched | 480.90 | 23 | 6.42 | 17.7 | - | 8.5 | 13.7 | – | – |
| 9 | CG-14 | 2 | Enriched | 470.59 | 27 | 9.56 | 26.5 | 66 | 10.3 | 21.5 | 3.57 | 20 |
| 10 | CG-16 | 2 | Enriched | 260.52 | 2 | 1.22 | 10.1 | 13 | 9.2 | 26.4 | 6.09 | 19 |
| 11 | CG-03 | 3 | Enriched | 438.65 | 11 | 4.34 | 13.7 | 23 | 10.3 | 27.4 | 4.84 | 22 |
| 12 | CG-20 | 3 | Enriched | 214.62 | 14 | 6.42 | 21.0 | 179 | 18.1 | 48.3 | 0.96 | 15 |
| 13 | CG-23 | 3 | Enriched | 174.89 | 33 | 9.85 | 25.7 | 156 | 12.0 | 34.2 | 1.47 | 14 |
| 14 | Xtra-4 | 3 | Enriched | 409.88 | 63 | 12.69 | 23.6 | 26 | 11.5 | 26.7 | 6.85 | 25 |
| 15 | CG-18 | 3 | Enriched | 410.31 | 77 | 18.13 | 29.0 | 57 | 14.5 | 33.3 | 3.17 | 24 |
| 16 | CG-22 | 3 | Enriched | 418.92 | 65 | 21.26 | 50.7 | 68 | 15.1 | 47.5 | 3.84 | 18 |
| 17 | CG-04 | 4 | Enriched | 442.32 | 150 | 30.78 | 46.1 | 69 | 15.8 | 40.4 | 4.10 | 25 |
| 18 | Xtra-2 | 4 | Enriched | 442.43 | 101 | 23.63 | 62.9 | 73 | 14.4 | 35.7 | 5.05 | 25 |
| 19 | CG-17 | 4 | Enriched | 474.22 | 73 | 20.76 | 32.1 | 36 | 14.0 | 36.6 | 5.11 | 29 |
| 20 | NAT-2 | 4 | Natural | 431.21 | 216 | 41.40 | 63.7 | 38 | 15.9 | 32.9 | 8.41 | 38 |
| 21 | CG-21 | 4 | Enriched | 233.08 | 23 | 10.06 | 27.2 | 61 | 14.6 | 58.1 | 2.40 | 20 |
| 22 | CG-10 | 5 | Enriched | 440.47 | 14 | 6.91 | 17.2 | 68 | 14.4 | 39.3 | 1.94 | 19 |
| 23 | CG-08 | 5 | Enriched | 431.00 | 79 | 28.19 | 53.2 | 84 | 18.1 | 44.0 | 3.36 | 17 |
| 24 | CG-24 | 5 | Enriched | 429.62 | 42 | 14.33 | 26.7 | 62 | 11.9 | 26.1 | 3.06 | 29 |
| 25 | CG-25 | 5 | Enriched | 434.51 | 14 | 5.32 | 18.3 | 35 | 9.7 | 20.5 | 4.84 | 26 |
| 26 | CG-27 | 5 | Enriched | 431.18 | 11 | 5.77 | 13.7 | 21 | 10.6 | 26.8 | 5.36 | 25 |
| ZnSe Median | 28 | 9.95 | 24.6 | 59.3 | 13.5 | 35.7 | 3.47 | 22 | ||||
| LD Median | 6.1 | 4.16 | 6.5 | – | 3.5 | 7.1 | – | – |
Fig. 28Total alpha energy spectrum of all detectors
Evaluated internal radioactive contamination for the CUPID-0 detector. The values reported refer to the overall detector radiopurity (CUPID-0), the lowest (Best) and highest (Worst) measured contamination in a single crystal. Po values refers to the sum of bulk and surface contaminations. Limits are evaluated at 90% C.L.
| Chain | Nuclide | Activity | Activity | Activity |
|---|---|---|---|---|
| [ | [ | [ | ||
| CUPID-0 | Lowest | Highest | ||
| 2.5 ± 0.2 | <0.54 | 8.6 ± 1.2 | ||
| 13.6 ± 0.4 | 2.3 ± 0.8 | 26.9 ± 2.2 | ||
| 10.9 ± 0.3 | 2.1 ± 0.6 | 23.1 ± 0.2 | ||
| 12.2 ± 0.6 | <3.7 | 24.2±3.5 | ||
| 5.1 ± 0.2 | <1.2 | 12.7 ± 1.5 | ||
| 5.3 ± 0.8 | 1.0 ± 2.0 | 14.7±4.3 | ||
| 5.3 ± 0.2 | <2.4 | 16.4±1.7 | ||
| 17.0 ± 0.4 | 3.8 ± 0.9 | 18.4 ± 1.8 | ||
| 17.4 ± 0.4 | 3.4 ± 0.6 | 19.8 ± 1.9 | ||
| 18.8 ± 0.6 | 9.1 ± 0.3 | 45.4 ± 2.6 |