Characterization of Residual Conduction Gaps After HotBalloon-Based Antral Ablation of Atrial Fibrillation - Evidence From Ultra-High-Resolution 3-Dimensional Mapping.

Shiro Nakahara1, Yuichi Hori1, Reiko Fukuda1, Naoki Nishiyama1, Sayuki Kobayashi1, Yoshihiko Sakai1, Isao Taguchi1.   

Abstract

BACKGROUND: The electrophysiological characteristics of residual conduction gaps between the left atrium (LA) and pulmonary veins (PVs) after HotBalloon-based wide antral ablation (HBWA) of atrial fibrillation (AF) remain incompletely understood. This study aimed to characterize the residual gaps by means of ultra-high-resolution mapping.Methods and 
Results: A total of 55 AF patients underwent HBWA by a predetermined protocol (6-shot total-based antral approach). LA-PV maps were created using 64-electrode minibasket catheters. In total, 55 residual gaps were identified among 26 (47%) patients. Residual gaps included 33 left superior (LS: 60%), 10 left inferior (18%), 6 right superior (11%), and 6 right inferior (11%) PVs. Those gaps demonstrated both extremely lower bipolar amplitudes (0.11 [interquartile range: 0.06-0.27] mV) and conduction velocities (0.75±0.27 m/s); however, the length was confined (10.3±4.1 mm) except for the LSPV anterior carina (12.2±2.4 mm) region. Among the carina regions, all gaps had far-field potentials consistently added to the PV potentials. Left atrial appendage pacing to split the far-field activity identified confined gap regions (6.7±1.9 mm). Touch-up ablation eliminated the residual PV potentials in all cases.
CONCLUSIONS: HBWA resulted in a certain degree of residual gap conduction in particular antral regions. These gaps exhibited narrow lengths with lower amplitudes, and often had far-field recordings from the left atrial appendage. Combined with pacing maneuvers, ultra-high-resolution activation maps could both visualize all confined gaps and ensure a bare minimum of touch-up ablations in all patients with gap conduction.

Entities:  

Keywords:  Atrial fibrillation; Conduction gap; Hot balloon ablation; Rhythmia mapping system

Year:  2019        PMID: 30996155     DOI: 10.1253/circj.CJ-18-1051

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  1 in total

1.  Impact of monitoring surface temperature during pulmonary vein isolation in a second-generation hot balloon system.

Authors:  Hiroshi Fukunaga; Yukio Sekiguchi; Sou Asano; Takahiko Nagase; Kohei Tanizaki; Jun Umemura; Junichi Nitta; Mitsuaki Isobe
Journal:  Int J Cardiol Heart Vasc       Date:  2022-02-03
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.