Literature DB >> 30994706

Cyclic fatigue resistance of novel Genius and Edgefile nickel-titanium reciprocating instruments.

Carlos Roberto Emerenciano Bueno1, Marina Tolomei Sandoval Cury1, Ana Maria Veiga Vasques1, Gustavo Sivieri-Araújo1, Rogério Castilho Jacinto1, João Eduardo Gomes-Filho1, Luciano Tavares Angelo Cintra1, Eloi Dezan-Júnior1.   

Abstract

This study aimed to assess the cyclic fatigue resistance of Genius and EdgeFile X1 reciprocating instruments compared with WaveOne Gold Primary. Twenty Genius (Ultradent) 25.04, 20 Genius 30.04, 20 EdgeFile X1 (EdgeEndo) and 20 WaveOne Gold Primary (Dentsply Maillefer) instruments were included in this study and tested in a static cyclic fatigue testing device, which has an artificial stainless steel canal with a 60° angle of curvature and a 5-mm radius of curvature. All instruments were operated in reciprocation mode until fracture occurred. The number of cycles to failure (NCF) was calculated and time to fracture (TF) was recorded in seconds using a digital chronometer. The mean and standard deviations of NCF and TF were calculated for each reciprocating system and the data were subjected to Kruskal-Wallis one-way analysis of variance and to Dunn's test (p < .05) using SigmaPlot software (Systat software, CA, USA). The fractured surfaces of five instruments from each brand were randomly examined and microphotographed by a low-vacuum environmental scanning electron microscopy - SEM (Tabletop Microscope TM3030, Hitachi, Japan) to confirm the cyclic fatigue fracture. EdgeFile exhibited the highest cyclic fatigue resistance, followed by both Genius files (p < .05). Within the limitations of this in vitro study, EdgeFile X1 instruments had significantly higher cyclic fatigue resistance than did Genius and WaveOne Gold Primary instruments. The cyclic fatigue resistance of both Genius files was higher than that of WaveOne Gold Primary.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 30994706     DOI: 10.1590/1807-3107bor-2019.vol33.0028

Source DB:  PubMed          Journal:  Braz Oral Res        ISSN: 1806-8324


  2 in total

1.  A comparative study of cyclic fatigue of 6 endodontic systems. An in vitro study.

Authors:  Jorge Rubio; José-Ignacio Zarzosa; Susana Aranda; Alberto Casino; Antonio Pallarés
Journal:  J Clin Exp Dent       Date:  2022-07-01

2.  Multimethod Assessment of the Cyclic Fatigue Strength of ProGlider, Edge Glide Path and R-Pilot Endodontic Instruments.

Authors:  Jorge N R Martins; Duarte Marques; Isabel Vasconcelos; Sofia Arantes-Oliveira; João Caramês; Francisco Manuel Braz Fernandes
Journal:  Dent J (Basel)       Date:  2022-02-17
  2 in total