Three-Dimensional Corrective Osteotomy for Malunited Fractures of the Upper Extremity Using Patient-Matched Instruments: A Prospective, Multicenter, Open-Label, Single-Arm Trial.

Kunihiro Oka1,2, Hiroyuki Tanaka1, Kiyoshi Okada3, Wataru Sahara1, Akira Myoui3, Tomomi Yamada3, Michiro Yamamoto4, Shigeru Kurimoto4, Hitoshi Hirata4, Tsuyoshi Murase1.   

Abstract

BACKGROUND: Medical image processing has facilitated simulation of 3-dimensional (3-D) corrective osteotomy, and 3-D rapid prototyping technology has further enabled the manufacturing of patient-matched surgical guides and implants (patient-matched instruments, or PMIs). However, 3-D corrective osteotomy using these technologies has not been the standard procedure. We aimed to prospectively verify the efficacy and safety of PMIs in corrective osteotomy for deformities of the upper extremity.
METHODS: We enrolled 16 patients with a total of 17 bone deformities in the upper extremity. Eight patients had distal radial malunion; 5, distal humeral malunion; and 3, forearm diaphyseal malunion. All cases underwent 3-D corrective osteotomy with PMIs. The primary end point was the residual maximum deformity angle (MDA), which was calculated from 2 deformity angles-1 on the anteroposterior and 1 on the lateral postoperative radiograph. Secondary end points included the deformity angle on radiographs, 3-D error between the preoperative planning model and the postoperative result, range of motion, grip strength, pain measured with a visual analog scale (VAS), patient satisfaction, and Disabilities of the Arm, Shoulder and Hand (DASH) score.
RESULTS: The average MDA significantly improved from 25.5° preoperatively to 3.3° at the final follow-up (p < 0.001). The angular deformity was within 5° in all cases, except for 1 with distal radial malunion who had a higher angle on the anteroposterior radiograph. The error between the correction seen on the postoperative 3-D bone model and the planned correction was <1° and <1 mm. Flexion and extension of the wrist and pronation of the forearm of the patients treated for distal radial malunion improved significantly, and pronation improved for those treated for forearm diaphyseal malunion. The average VAS score, grip strength, and DASH score significantly improved as well. Of the 16 patients, 15 were very satisfied or satisfied with the outcomes.
CONCLUSIONS: Corrective osteotomy using PMIs achieved accurate correction and good functional recovery in the upper extremity. Although our study was limited to cases without any deformity on the contralateral side, 3-D corrective osteotomy using PMIs resolved treatment challenges for complex deformities in upper extremities. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Year:  2019        PMID: 30994589     DOI: 10.2106/JBJS.18.00765

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  8 in total

1.  Intra-articular opening wedge osteotomy for varus ankle arthritis with computer-assisted planning and patient-specific surgical guides: a retrospective case series.

Authors:  Xin-Long Ma; Jian-Xiong Ma; Xing-Wen Zhao; Yu-Ren Du; Ying Wang; Hao-Hao Bai; Bin Lu
Journal:  BMC Musculoskelet Disord       Date:  2022-05-21       Impact factor: 2.562

2.  Three-dimensional printing technology for patient-matched instrument in treatment of cubitus varus deformity: A case report.

Authors:  Nithid Sri-Utenchai; Nachapan Pengrung; Korakod Srikong; Chedtha Puncreobutr; Boonrat Lohwongwatana; Paphon Sa-Ngasoongsong
Journal:  World J Orthop       Date:  2021-05-18

3.  Patient-specific plate for navigation and fixation of the distal radius: a case series.

Authors:  Johannes G G Dobbe; Abbas Peymani; Hendrika A L Roos; Maikel Beerens; Geert J Streekstra; Simon D Strackee
Journal:  Int J Comput Assist Radiol Surg       Date:  2021-02-11       Impact factor: 2.924

4.  Three-dimensional printing in paediatric orthopaedic surgery.

Authors:  Sven Goetstouwers; Dagmar Kempink; Bertram The; Denise Eygendaal; Bart van Oirschot; Christiaan Ja van Bergen
Journal:  World J Orthop       Date:  2022-01-18

5.  3D accuracy and clinical outcomes of corrective osteotomies with patient-specific instruments in complex upper extremity deformities: an approach for investigation and correlation.

Authors:  Heide Delbrück; David Christian Weber; Jörg Eschweiler; Frank Hildebrand
Journal:  Eur J Med Res       Date:  2022-10-08       Impact factor: 4.981

6.  Preoperative psychometric properties of visual analog scale asessments for function, pain, and strength compared with legacy upper extremity outcome measures in glenohumeral osteoarthritis.

Authors:  Alexander Beletsky; Yining Lu; Benedict U Nwachukwu; Evan Polce; Bhargavi Maheshwer; Jorge Chahla; Brian Forsythe; Brian J Cole; Nikhil N Verma
Journal:  JSES Int       Date:  2020-06-17

7.  Psychometric properties of visual analog scale assessments for function, pain, and strength compared with disease-specific upper extremity outcome measures in rotator cuff repair.

Authors:  Alexander Beletsky; Benedict U Nwachukwu; Tomás Gorodischer; Jorge Chahla; Brian Forsythe; Brian J Cole; Nikhil N Verma
Journal:  JSES Int       Date:  2020-05-23

8.  Outcomes of 3-D corrective osteotomies for paediatric malunited both-bone forearm fractures.

Authors:  Kasper C Roth; Eline M van Es; Gerald A Kraan; Jan A N Verhaar; Filip Stockmans; Joost W Colaris
Journal:  J Hand Surg Eur Vol       Date:  2021-07-14