The accuracy and precision of an open-circuit system to measure oxygen consumption and carbon dioxide production in neonates.

K H Marks, P Coen, J R Kerrigan, N A Francalancia, E E Nardis, M T Snider.   

Abstract

We measured the oxygen consumption, carbon dioxide production, and respiratory quotient during the combustion of a known mass of anhydrous ethanol and methanol to assess the accuracy of an open-circuit flow-through system. Continuous measurements were made of the mass of alcohol burned, the velocity of gas flow through the apparatus, and simultaneous measurements of the fractional concentration of oxygen, carbon dioxide and nitrogen of the inlet and outlet gas using paramagnetic oxygen analyzer, infrared carbon dioxide meter, and mass spectrometer. Standard respiratory and stoichiometric equations were used to calculate the oxygen consumption, carbon dioxide production and RQ for the mass of absolute alcohol combustion per unit time. In a series of 12 consecutive laboratory experiments (on 7 days), the measured values of gas exchange (similar to the rate of respiratory gas exchange by an infant of 1-4 kg) were in agreement within 5% of the true values for ethanol and methanol combustion, confirming the validity of the open-circuit method. The paramagnetic oxygen analyzer and the mass spectrometer gave similar oxygen consumption results and differed very little when the rate of absolute alcohol combustion was used to quantify the accuracy of the complete measurement system. A positive measurement error was observed for the carbon dioxide production results from both the IR meter and mass spectrometer, with the result that the respiratory quotient measurements were 3.4-4.7% higher than the true value. The mass spectrometer gave more precise oxygen consumption results, whereas smaller variance of carbon dioxide production measurements was observed using the infrared CO2 meter.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3099255     DOI: 10.1203/00006450-198701000-00014

Source DB:  PubMed          Journal:  Pediatr Res        ISSN: 0031-3998            Impact factor:   3.756


  3 in total

1.  A new paediatric metabolic monitor.

Authors:  W Weyland; A Weyland; U Fritz; K Redecker; F B Ensink; U Braun
Journal:  Intensive Care Med       Date:  1994       Impact factor: 17.440

2.  Validation of a portable indirect calorimetry system for measurement of energy expenditure in sick preterm infants.

Authors:  G J Shortland; P J Fleming; J H Walter
Journal:  Arch Dis Child       Date:  1992-10       Impact factor: 3.791

3.  Multiple propane gas burn rates procedure to determine accuracy and linearity of indirect calorimetry systems: an experimental assessment of a method.

Authors:  Mohammad Ismail; Fabien Basset; Sanaa A Alsubheen; Angela Loucks-Atkinson; Matthew Atkinson; Thamir Alkanani; Liam P Kelly
Journal:  PeerJ       Date:  2022-08-29       Impact factor: 3.061

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.