Radiographic assessment of overlengthening of the MoPyC radial head prosthesis: a cadaveric study.

Marc Schnetzke1, Frank Feuchtenhofer1, Holger Keil1, Benedict Swartman1, Sven Vetter1, Paul-Alfred Grützner1, Jochen Franke2.   

Abstract

INTRODUCTION: The aim of this study was to evaluate a radiographic measurement method for assessment of overlengthening of the MoPyC radial head prosthesis.
MATERIALS AND METHODS: Seven cadaver specimens were studied in ten stages: native specimen (1), radial head resection (2), and implantation of the MoPyC radial head prostheses (Bioprofile, Tornier, Montbonnot-Saint-Martin, France) in four increasing length (correct length, overlengthening of 1.5 mm, 3 mm, and 5 mm) with an intact medial collateral ligament (MCL 3-6) and following transection of the MCL (7-10). The radiographic measurement method according to Athwal et al. was evaluated to detect overlengthening. Statistical analysis included calculation of the diagnostic accuracy of the radiographic method.
RESULTS: The radiographic measurement method correctly determined the size of the radial head prosthesis within ± 1 mm in 224 of 336 scenarios (67%) and within ± 2 mm in 320 of 336 scenarios (95%). With a threshold value of ≥ 1 mm, the overall diagnostic sensitivity for detecting overlengthening when it was present and was 90% and the specificity was 79%. The sensitivity was higher with increasing size of the prosthesis: in cases with overlengthening of 1.5 mm, the sensitivity was 76%, with an overlengthening of 3 mm, the sensitivity was 95%, and with an overlengthening of 5 mm, the sensitivity was 100%.
CONCLUSION: The radiographic measurement method of Athwal et al. can be used to estimate and to diagnose the magnitude overlengthening of the MoPyC radial head prosthesis. However, the sensitivity is limited (76%) in cases with a small amount of overlengthening of 1.5 mm.

Keywords:  Elbow; Fluoroscopy; Joint gapping; Oversizing; Overstuffing; Radial head prosthesis; Ulnohumeral

Year:  2019        PMID: 30989412     DOI: 10.1007/s00402-019-03187-7

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  2 in total

1.  Fluoroscopy of the Elbow: A Cadaveric Study Defining New Standard Projections to Visualize Important Anatomical Landmarks.

Authors:  Sebastian A Müller; Lars Adolfsson; Cornelia Baum; Magdalena Müller-Gerbl; Andreas M Müller; Daniel Rikli
Journal:  JB JS Open Access       Date:  2021-05-05

2.  Monteggia-like lesions in adults treated with radial head arthroplasty-mid-term follow-up of 27 cases.

Authors:  Matthias Jung; Corinna Groetzner-Schmidt; Felix Porschke; Paul A Grützner; Thorsten Guehring; Marc Schnetzke
Journal:  J Orthop Surg Res       Date:  2020-01-03       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.