Harboring Contaminants in Repeatedly Reprocessed Pedicle Screws.

Aakash Agarwal1, Christian Schultz2, Anand K Agarwal1, Jeffrey C Wang3, Steve R Garfin4, Neel Anand5.   

Abstract

STUDY
DESIGN: It consisted of evaluation of the pedicle screws for presence of residual nonmicrobial contaminants and tabulation of the minimum steps and time required for reprocessing implants as per guidelines and its comparison with actual practice.
OBJECTIVE: An evaluation of the nonmicrobial contaminants prevalent on the pedicle screws used for spine surgery and the underlying practice cause behind the source.
METHODS: The first component consisted of a random selection of 6 pedicle screws and its assessment using optical microscopy, scanning electron microscopy with energy dispersive spectroscopy, and Fourier transform infrared spectroscopy. The second component consisted of review of implant reprocessing guidelines and its applicability.
RESULTS: Three types of contaminants were identified: corrosion, saccharide of unknown origin, and soap residue mixed with and were mostly present at the interfaces with low permeability. In addition, manufacturer's guideline recommends 19 hours of reprocessing, whereas the real-time observation revealed a turnaround time of 1 hour 17 minutes.
CONCLUSION: Repeatedly reprocessed pedicle screws host corrosion, carbohydrate, fat, and soap, which could be a cause of surgical site infection and inflammatory responses postsurgery. The cause behind it is the impracticality of repeated cleaning and inspection of such devices.

Entities:  

Keywords:  SPD; SSI; implant exposure; preoperative implant handling; sterile processing department; surgical site infection; terminally sterile devices

Year:  2018        PMID: 30984497      PMCID: PMC6448207          DOI: 10.1177/2192568218784298

Source DB:  PubMed          Journal:  Global Spine J        ISSN: 2192-5682


Introduction

Surgical site infection (SSI) is known to occur at the rate of 12.7% following spinal fusion.[1] Pedicle screws are the key implants used in these procedures and have become synonymous with the term spinal fusion. The sterilization processing department (SPD) is at the core of conventional processing of these screws, with daily activities ranging from receiving contaminated surgical instruments and implants, to redistributing them sterilized to the various departments and surgical units in the facility (both on and off campus). The services performed includes decontamination, washing, reassembly in trays or as single items, wrapping, labeling, and sterilization.[2] Efficiency in this process is paramount for the proper surgical management of the patients, consequently avoiding costly delays and SSIs.[3] A caveat with this process is the low ratio of used to reprocessed implants (∼0.03-0.08), resulting in multiple reprocessing life cycle per individual implant before implantation. Thus, on the postoperatively returned implants with low cleanliness assurance in adjunct with highly contaminated instruments (harboring macroscopic human tissues and blood-borne pathogens) deems thorough reprocessing and visual inspection critical to avoid buildup of contaminations or formation of biofilm thereof. However, the most common implant used in spinal fusion, pedicle screws, are multicomponent with lumens, interfaces, and crevices in the range of 0.2 to 1.5 mm. This raises a concern regarding the practicality of a repeated cleaning process, at the heart of which lies the manufacturer’s instruction for cleaning and sterilization. The purpose of this study was to evaluate the pedicle screws retrieved from reprocessed units stored and currently in circulation at surgical centers for presence of residual nonmicrobial contaminants and/or foreign material. In conjunction, multiple manufacturer’s instructions were reviewed to determine the basic steps and the minimum time recommended to clean and sterilize a tray of 164 pedicle screw implants and practicality of such a procedure.

Methods

The study design consisted of 2 components: (1) evaluation of the pedicle screws for presence of residual nonmicrobial contaminants and/or foreign material and (2) tabulation of the minimum steps and time required for reprocessing implants as per manufacturer’s guidelines and its comparison with actual practice. The first component consisted of a random selection of 6 pedicle screws from 4 different trays of cleaned, wrapped, and sterilized implants. The screws were retrieved using clean gloves, followed by its immediate placement inside of clean zip-lock polybags. Each pedicle screw was disassembled and sent for optical microscopy, scanning electron microscopy with energy dispersive spectroscopy, and Fourier transform infrared spectroscopy. The type and the size of contaminants were recorded. The second component consisted of review of 4 major manufacturer’s instructions for reprocessing pedicle screws. Each instruction manual was then converted into a tabulation with the various steps and the minimum time required for each step as per the written guidelines. The common steps were recorded along with the minimum time required in each step. This was then compared with the real-time observations from state of art SPDs on the exact procedure of reprocessing a pedicle screw set.

Results

Three types of contaminants were identified: corrosion, saccharide of unknown origin (biofilm, endotoxins, fatty tissue), and soap residue mixed with fat, each occupying isolated diametrical areas of 1.4 mm, 1.5 mm, and 3.4 mm, respectively (Figures 1 –5 and Table 1). In addition, salt residues were also found at interfaces between the tulip head and shaft (Table 2). The corrosion stains were present on the outer surfaces of the implants, whereas an active corrosion with material erosion was seen at the inner rim of the pedicle screw head (tulip) and some parts of the washer. The saccharides and soap were present in the interfaces with low permeability (interior region of the multipiece assembled device).
Figure 1.

Corrosion on the washer.

Figure 2.

Corrosion on the tulip interface.

Figure 3.

Saccharide of unknown origin.

Figure 4.

Soap residue mixed with fat.

Figure 5.

FTR conforming the results.

Table 1.

Summary of the Results in a Tabulated Form.

SampleCompanySizeResultsEquipment Used for Analysis
1Company A7.5 × 55 mm

Saccharide residue

Stains

Optical microscope

FEI Quanta 3D FEG environmental scanning electron microscope and focused ion beam

Varian Excalibur Series Fourier transform infrared (FTIR) instruments, the FTS-4000 Spectrometer and the UMA-600 Microscope

2Company A7.5 × 45 mm

Soap residue mixed with fat

Stains

3Company B6.5 × 45 mm

Stains. The sample could not be safely disassembled for evaluation

Optical microscope

4Company C7.5 × 45 mm

Corrosion at the tulip interface

Corrosion on the washer surface

Optical microscope

FEI Quanta 3D FEG environmental scanning electron microscope and focused ion beam

5Company D5.5 × 45 mm

Stains

Salts

6Company D7.5 × 55 mm

Stains

Salts

Table 2.

Energy Dispersive X-Ray Spectroscopy Showing Presence of Salt Elements.

ElementWeight (%)Atomic (%)
C K14.0832.64
O K13.2923.13
Al K6.066.25
Si K0.140.14
Cl K3.923.08Salt element
K K8.586.11Salt element
Ti K41.1523.92
V K1.240.68
Fe K0.140.07
Br K11.383.97Salt element
Total100.00
Corrosion on the washer. Corrosion on the tulip interface. Saccharide of unknown origin. Soap residue mixed with fat. FTR conforming the results. Summary of the Results in a Tabulated Form. Saccharide residue Stains Optical microscope FEI Quanta 3D FEG environmental scanning electron microscope and focused ion beam Varian Excalibur Series Fourier transform infrared (FTIR) instruments, the FTS-4000 Spectrometer and the UMA-600 Microscope Soap residue mixed with fat Stains Stains. The sample could not be safely disassembled for evaluation Optical microscope Corrosion at the tulip interface Corrosion on the washer surface Optical microscope FEI Quanta 3D FEG environmental scanning electron microscope and focused ion beam Stains Salts Stains Salts Energy Dispersive X-Ray Spectroscopy Showing Presence of Salt Elements. The tabulated form of manufacturer’s guideline consisted of at least 19 disjoint steps and a minimum of 19 man-hours required for reprocessing an implant tray with 164 pedicle screws (Table 3). In comparison, the real-time observation revealed a substantial lower turnaround time for each set, as it only included the processes of mechanical washing (in the same chamber and along with all the dirty instruments), thermal disinfection, packaging for sterilization, steam sterilization, and drying (Table 3).
Table 3.

Quantification of a Typical Manufacturer’s Reprocessing Guideline.

ProcessStepMinimum Time (Minutes)TypeAllocationFor 164 Devices and 1 Person (Guideline)Real-Time Observation
Inspection10.15VisualPer device24.6N/A
Precleaning21Running waterPer device164N/A
32Cleaner and brushPer device328N/A
40.1WaterAll0.1N/A
51Water jetPer device164N/A
615Ultrasonic cleanerAll15N/A
72DI water jetPer device328N/A
Inspection80.15VisualPer device24.6N/A
Mechanical washer92PrewashAll24
102Wash IAll212
115Wash IIAll5
122RinseAll25
1340DryAll4020
Thermal disinfection142.5At 93°CAll2.58
Inspection150.15VisualPer device24.6N/A
Packaging for sterilization16N/AWrapsAll33
Sterilization174SteamAll44
Drying1820Up to 60 minutesAll2020
Total steps18Total time19 hours1 hour 16 minutes
Quantification of a Typical Manufacturer’s Reprocessing Guideline.

Discussion

Reprocessing and sterilization of orthopedic implants is a labor-intensive process and requires great precision and technical know-how. Ineffective execution can compromise patient’s health along with wasting hundreds of thousands of dollars. Furthermore, the medical device turnaround time has drastically reduced, along with a streak of sophisticated devices being released every year.[4] Alfa et al showed through their study that the screws in the sterilization racks have limited access to the cleaning fluids resulting in insufficient cleaning and rinsing in an automated washer.[5,6] Additionally, their study demonstrated an increase in endotoxin levels post reprocessing. Complementing it was the study from Litrico et al, who reported results on presterile single packed screws and compared it with an older series, which used reprocessed implants, and was performed by the same team.[7] They found that the infection rate was lower with presterile single packed screws compared with the reprocessed implants (2% vs 6%). SSI adds an enormous burden to individuals and society in terms of medications, reoperations, extended stays at the hospital, lost productivity and wages, and emotional and physical trauma afflicted on patients and their families.[8] Our results indicate that implantable pedicle screws, which are reprocessed, harbor contaminants. This could be one of the direct cause of SSI, resulting in additional burden and morbidity to patients involved, which could have been avoided. While the exact source of each contaminant is unknown, presence of foreign material residues on the inner surfaces and at the interfaces of a pedicle screw is unacceptable for an immune-compromised elderly patient. The sources of these contaminants could range from mucous-like deposits from bacteria, biofilms, fatty tissue residues left over from reprocessing the implants with other contaminated instruments, and insufficient rinsing after cleansing with detergent. Previous studies on endoscopes and general orthopedic also indicate that the execution of reprocessing in the health facilities are impractical considering the workload and the intricacies of such devices. Some countries (eg, Japan and Scotland) have banned reprocessing of implants used for spine surgery. In Scotland, for example, the deadline for conversion of all orthopedic units to prepackaged, sterile, single-use implants was December 31, 2007.[9] It was pointed out by the Scottish Health Department that repeatedly reprocessing implants in the hospital is a suboptimal clinical practice. Previous studies have also observed that during reprocessing (mostly instruments) 79% of visual inspections are not performed correctly, 57% of the washer-disinfectors are obsolete or not suitable for performing a validated process, 64% of the reprocessing facility are in need of renovation, and 100% demonstrates a lack of a validated reprocessing method.[10] When categorized by the date of facility establishment, an older facility had a higher number of deficiencies over a newer one. This indicates an existence of resistance in change of standard operation, and hence quality, with regard to changes in technology and accessibility. Nevertheless, the current study observed and collected implant samples from state-of-the art facilities following a strict standard of operation as directed by the hospital administration. In addition, our study did not record or access instrument reprocessing. Therefore, the failure mode here is not the lack of compliance by SPD but the underlying impracticality of repeated cleaning and sterilization of hundreds of small implants with multiple components, each with interface clearances of less than a fraction of millimeter. The medical device turnaround time has drastically reduced, with streak of sophisticated devices being released every year.[4] There already exists evidence that the amount of microscopic carbohydrate residue and endotoxins on any device increases after reprocessing.[5] The increase in more complex medical devices being released to the market undoubtedly necessitates new requirements for handling such devices, and therefore instead of prescribing impractical reprocessing guidelines for SPD, focus should be on providing in a presterile package. The current study demonstrates the risks associated with repeated reprocessing of pedicle screws and the overall inapplicability of manufacturer’s guidelines in a clinical scenario. This information would be crucial for hospitals to reduce liability toward the patients, surgeons, and the administration, by taking appropriate steps to mitigate repeated reprocessing of pedicle screws.

Conclusion

Despite improvements in health care, the practice of reprocessing implants has stayed unaltered. In our evaluation, we discovered corrosion, carbohydrate, fat, and soap on reprocessed pedicle screw implants obtained from reprocessed implant sets in clinical circulation. These results indicate that the reprocessed devices have the potential to not be thoroughly cleaned during reprocessing and prior to sterilization. In par with this is the impracticality of cleaning and inspection methodology prescribed by the manufacturers for cleaning these single-use devices.
  9 in total

1.  Perspectives from sterile processing and perioperative services.

Authors:  Nancy Chobin; Barbara Trattler
Journal:  Mater Manag Health Care       Date:  2003-02

2.  Shifting the sterile processing department paradigm: a mandate for change.

Authors:  Scott C Swanson
Journal:  AORN J       Date:  2008-08       Impact factor: 0.676

3.  The 'pandora's box' dilemma: reprocessing of implantable screws and plates in orthopedic tray sets.

Authors:  Michelle J Alfa
Journal:  Biomed Instrum Technol       Date:  2012

4.  Cleaning efficacy of medical device washers in North American healthcare facilities.

Authors:  M J Alfa; N Olson; A Al-Fadhaly
Journal:  J Hosp Infect       Date:  2009-08-27       Impact factor: 3.926

5.  Hospital uses team approach to improve processes, reduce costs.

Authors:  M A Bell
Journal:  AORN J       Date:  1998-07       Impact factor: 0.676

6.  The impact of surgical-site infections following orthopedic surgery at a community hospital and a university hospital: adverse quality of life, excess length of stay, and extra cost.

Authors:  James D Whitehouse; N Deborah Friedman; Kathryn B Kirkland; William J Richardson; Daniel J Sexton
Journal:  Infect Control Hosp Epidemiol       Date:  2002-04       Impact factor: 3.254

7.  Analysis of Postoperative Thoracolumbar Spine Infections in a Prospective Randomized Controlled Trial Using the Centers for Disease Control Surgical Site Infection Criteria.

Authors:  Shearwood McClelland; Richelle C Takemoto; Baron S Lonner; Tate M Andres; Justin J Park; Pedro A Ricart-Hoffiz; John A Bendo; Jeffrey A Goldstein; Jeffrey M Spivak; Thomas J Errico
Journal:  Int J Spine Surg       Date:  2016-04-21

8.  Single-use instrumentation in posterior lumbar fusion could decrease incidence of surgical site infection: a prospective bi-centric study.

Authors:  Stéphane Litrico; Geoffrey Recanati; Antoine Gennari; Cédric Maillot; Mo Saffarini; Jean-Charles Le Huec
Journal:  Eur J Orthop Surg Traumatol       Date:  2015-09-01

9.  Evaluation of reprocessing medical devices in 14 German regional hospitals and at 27 medical practitioners' offices within the European context - consequences for European harmonization.

Authors:  Birgit Thiede; Axel Kramer
Journal:  GMS Hyg Infect Control       Date:  2013-11-06
  9 in total
  4 in total

1.  Implant Retention or Removal for Management of Surgical Site Infection After Spinal Surgery.

Authors:  Aakash Agarwal; Amey Kelkar; Ashish G Agarwal; Daksh Jayaswal; Christian Schultz; Arvind Jayaswal; Vijay K Goel; Anand K Agarwal; Sandeep Gidvani
Journal:  Global Spine J       Date:  2019-08-11

2.  High Prevalence of Biofilms on Retrieved Implants from Aseptic Pseudarthrosis Cases.

Authors:  Aakash Agarwal; Megan Mooney; Ashish G Agarwal; Daksh Jayaswal; Gayane Saakyan; Vijay Goel; Jeffrey C Wang; Neel Anand; Steve Garfin; Vithal Shendge; Hossein Elgafy
Journal:  Spine Surg Relat Res       Date:  2020-09-23

Review 3.  Updates on Evidence-Based Practices to Reduce Preoperative and Intraoperative Contamination of Implants in Spine Surgery: A Narrative Review.

Authors:  Aakash Agarwal; Boren Lin; Hossein Elgafy; Vijay Goel; Chris Karas; Christian Schultz; Neel Anand; Steve Garfin; Jeffrey Wang; Anand Agarwal
Journal:  Spine Surg Relat Res       Date:  2019-06-21

4.  Implant contamination as a cause of surgical site infection in spinal surgery: are single-use implants a reasonable solution? - a systematic review.

Authors:  Friederike Schömig; Carsten Perka; Matthias Pumberger; Rudolf Ascherl
Journal:  BMC Musculoskelet Disord       Date:  2020-09-25       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.