Literature DB >> 30982763

High Failure at a Minimum 5-Year Follow-Up in Primary Total Hip Arthroplasty Using a Modular Femoral Trunnion.

Cindy R Nahhas1, Paul H Yi2, Mario Moric1, Rajeev Puri3, Joshua J Jacobs1, Scott M Sporer1.   

Abstract

BACKGROUND: Modular femoral trunnions enable the surgeon to independently adjust offset, leg length, and anteversion in total hip arthroplasty (THA). However, modularity may result in an increased risk of fretting and corrosion along with a higher risk of implant dissociation or fracture. The purpose of this study is to evaluate mid-term survivorship of THAs using a cementless modular system.
METHODS: A consecutive series of 221 patients who underwent a primary THA using the ALFA II modular stem by a single surgeon between 2002 and 2004 were reviewed. Survivorship of the ALFA II modular hip system was evaluated at a minimum of 5 years postoperatively.
RESULTS: Of the 221 patients, 28 (12.7%) died from causes unrelated to the surgery before adequate follow-up, and 64 (29.0%) patients were lost to follow-up. The remaining 129 patients had a mean 6.5-year (range: 5-8 years) follow-up. All-cause survivorship of the modular stem system was 81% (95% confidence interval = 69-90) at a mean 6.5-year follow-up. Of the 25 (19.4%) cases requiring revision surgery, 52.0% was for dissociation of the modular components, 32.0% was for fracture of the prosthesis, 12.0% was for instability/multiple dislocations, and 4.0% was for chronic septic THA. Body mass index (odds ratio = 1.080) and offset (odds ratio = 1.254) were independent risk factors for mechanical failures of the modular stem system.
CONCLUSION: The modular stem hip system of interest in this study demonstrates a high failure rate at mid-term follow-up, and we caution against the use of similar designs in primary THAs. Published by Elsevier Inc.

Entities:  

Keywords:  cementless; modular neck; modularity; primary total hip arthroplasty; survivorship

Mesh:

Year:  2019        PMID: 30982763     DOI: 10.1016/j.arth.2019.03.033

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  2 in total

1.  Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years' follow-up.

Authors:  Sverrir Kiernan; Bart Kaptein; Carl Flivik; Martin Sundberg; Gunnar Flivik
Journal:  Acta Orthop       Date:  2020-12-10       Impact factor: 3.717

2.  Trunnion fracture of femoral prosthesis following a large metal-on-metal uncemented total hip arthroplasty: a case report.

Authors:  Krishna Kiran Eachempati; Chandra Sekhar Dannana; Sunil Apsingi; Vinay Kishore Ponnala; Giridhar Boyapati; Apurve Parameswaran
Journal:  Arthroplasty       Date:  2020-10-28