Force and aspiration on catheters utilized in the ADAPT technique in acute ischemic stroke: A bench top analysis.

Mickey Smith1, Jonathan Pace2, Connie Ju3, Yin C Hu2.   

Abstract

INTRODUCTION: Given the high morbidity and mortality of stroke, there remains a demand for techniques that provide rapid and safe intervention while improving time to recanalization. The direct aspiration first pass technique (ADAPT) uses force and aspiration for clot removal without the aid of separators or retrievers. In this study, we compare the force and aspiration qualities of commercially available catheters.
METHODS: Four different catheters with varying inner diameters were set up in a bench top model to test catheter tip pressure and flow rate. Catheter tip pressure was measured by attaching the catheter to a vacuum pressure gauge and an aspiration pump. The flow rate was calculated by measuring the volume of room temperature water aspirated through each catheter over a given time.
RESULTS: The Microvention Sofia catheter generated the greatest tip force (21.32 g), and the Stryker AXS Catalyst 6 catheter generated the smallest tip force (15.88 g). The Penumbra ACE 068 catheter and Medtronic ARC catheter measured 20.87 g and 16.78 g respectively. The ACE 068 had highest rate of aspiration at 289 mL/min, and the Catalyst 6 catheter had the lowest rate at 214. The Microvention Sofia catheter had the second highest rate while the ARC had the third highest rate, measuring 285 mL/min and 256 mL/min, respectively.
CONCLUSIONS: When using the ADAPT technique, knowledge of the tip force and catheter flow rate of newer catheters with larger distal inner diameters may guide selection of aspiration catheters. While this study demonstrates differences in tip force and flow rate of different commercially available catheters, clinical translation will require further testing and evaluation.
Copyright © 2019 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Aspiration; Catheter; Ischemic stroke; Thrombectomy

Mesh:

Year:  2019        PMID: 30981827     DOI: 10.1016/j.neurad.2019.03.013

Source DB:  PubMed          Journal:  J Neuroradiol        ISSN: 0150-9861            Impact factor:   3.447


  2 in total

1.  Intracranial mechanical thrombectomy of large vessel occlusions in the posterior circulation using SAVE.

Authors:  Volker Maus; Hanna Styczen; Jan Liman; Ilko Maier; Alex Brehm; Ioannis Tsogkas; Marios-Nikos Psychogios
Journal:  BMC Neurol       Date:  2019-08-16       Impact factor: 2.474

2.  Tailored Vessel-Catheter Diameter Ratio in a Direct Aspiration First-Pass Technique: Is It a Matter of Caliber?

Authors:  E Pampana; S Fabiano; G De Rubeis; L Bertaccini; A Stasolla; A Vallone; A Pingi; M Mangiardi; S Anticoli; C Gasperini; E Cotroneo
Journal:  AJNR Am J Neuroradiol       Date:  2021-01-21       Impact factor: 3.825

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.