| Literature DB >> 30976727 |
Baharan Kamousi1, Alexander M Grant1, Brad Bachelder1, Jianchun Yi1, Mehdi Hajinoroozi1, Raymond Woo1.
Abstract
OBJECTIVES: To compare the quality of electroencephalography (EEG) signals recorded with a rapid response EEG system and the signals recorded with conventional clinical EEG recordings.Entities:
Keywords: Concordance; Easy EEG; Electroencephalography; Rapid response EEG
Year: 2019 PMID: 30976727 PMCID: PMC6444024 DOI: 10.1016/j.cnp.2019.02.002
Source DB: PubMed Journal: Clin Neurophysiol Pract ISSN: 2467-981X
Fig. 1The rapid response EEG system consisting of a portable EEG recorder and a disposable electrode headband.
Fig. 2Simultaneous recordings in a single subject with three EEG systems: rapid response EEG, conventional EEG #1, and conventional EEG #2.
The tested rapid response EEG system’s electrode pairs.
| Rapid response EEG system channels | Corresponding conventional EEG channels | Rapid response EEG system channels | Corresponding conventional EEG channels |
|---|---|---|---|
| 1–2 | Fp1-F7 | 6–7 | Fp2-F8 |
| 2–3 | F7-T3 | 7–8 | F8-T4 |
| 3–4 | T3-T5 | 8–9 | T4-T6 |
| 4–5 | T5-O1 | 9–10 | T6-O2 |
Fig. 3Comparison of epochs of interest showing the EEG waveforms captured by the tested rapid response EEG (blue traces), Conventional EEG #1 (orange traces), and Conventional EEG #2 (yellow traces).
Wilcoxon signed-rank test analysis of calculated parameters for ICU study of 22 subjects with non-simultaneous EEG recordings with the rapid response EEG and conventional EEG systems.
| Metric | Hjorth Activity | Hjorth Mobility | Hjorth Complexity | Artifact Spike count | Baseline wander | Power of 60 Hz noise | Kurtosis |
|---|---|---|---|---|---|---|---|
| P-value | 0.4264 | 0.0883 | 0.4264 | 0.2912 | 0.1579 | 0.0005 | 0.4455 |
Calculated metrics for all 22 patients recorded consecutively with the rapid response EEG (rapid resp.) and conventional EEG (con.) systems in an ICU environment.
| Hjorth Activity | Hjorth Mobility | Hjorth Complexity | Artifact SpikeCount | Baseline Wander | Power of 60 Hz Noise | Kurtosis | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Subj. | Rapid resp. | Con. | Rapid resp. | Con. | Rapid resp. | Con. | Rapid resp. | Con. | Rapid resp. | Con. | Rapid resp. | Con. | Rapid resp. | Con. |
| 1 | 144 | 270 | 0.60 | 0.47 | 3.15 | 3.58 | 0 | 0.04 | 0.22 | 0.36 | 2.19 | 9.15 | 3.53 | 10.15 |
| 2 | 137 | 3008 | 0.46 | 0.68 | 3.46 | 3.23 | 0.02 | 0.02 | 0.15 | 0.73 | 0.45 | 101.66 | 13.05 | 9.32 |
| 3 | 26 | 2648 | 0.49 | 1.34 | 3.05 | 1.24 | 0.01 | 0.01 | 0.15 | 0.17 | 0.28 | 2197.8 | 3.87 | 4.96 |
| 4 | 1501 | 280 | 0.65 | 1.21 | 3.53 | 1.45 | 0 | 0 | 0.39 | 0.2 | 0.04 | 236.42 | 2.94 | 3.14 |
| 5 | 286 | 106 | 1.01 | 0.59 | 1.60 | 4.32 | 0 | 0 | 0.16 | 0.22 | 0.1 | 52.02 | 2.79 | 2.78 |
| 6 | 483 | 46,925 | 0.43 | 0.59 | 3.99 | 2.86 | 0.01 | 0.01 | 0.33 | 0.69 | 1.17 | 5356.6 | 4.65 | 4.67 |
| 7 | 257 | 310 | 0.87 | 0.81 | 1.92 | 2.45 | 0.02 | 0.01 | 0.3 | 0.34 | 3.31 | 19.38 | 7 | 8.22 |
| 8 | 2071 | 3221 | 0.59 | 0.47 | 3.01 | 5.74 | 0.01 | 0.01 | 1.06 | 1.99 | 11.31 | 4.99 | 4.84 | 6.81 |
| 9 | 40 | 1229 | 0.46 | 1.39 | 3.74 | 1.20 | 0.01 | 0.01 | 0.1 | 0.2 | 0.35 | 1518 | 4.16 | 5.89 |
| 10 | 91 | 14,076 | 0.67 | 1.14 | 2.41 | 1.80 | 0.02 | 0.02 | 0.19 | 1.11 | 9.65 | 6232.4 | 6.23 | 7.09 |
| 11 | 5559 | 3241 | 0.98 | 1.09 | 1.38 | 1.30 | 0.15 | 0.02 | 0.39 | 0.26 | 312.52 | 242.38 | 19.98 | 16.58 |
| 12 | 26,782 | 549 | 0.31 | 0.76 | 4.72 | 2.63 | 0.07 | 0.01 | 0.16 | 0.26 | 0.76 | 118.74 | 9.39 | 4.76 |
| 13 | 38 | 52 | 0.15 | 0.47 | 9.58 | 4.26 | 0 | 0 | 0.19 | 0.21 | 0.01 | 6.86 | 3.6 | 3.07 |
| 14 | 159 | 82 | 0.69 | 0.44 | 2.17 | 3.29 | 0.01 | 0 | 0.32 | 0.25 | 2.53 | 1.43 | 4.37 | 3.36 |
| 15 | 88 | 84 | 0.55 | 0.51 | 2.88 | 3.54 | 0 | 0 | 0.19 | 0.21 | 0.57 | 14.18 | 3.23 | 2.95 |
| 16 | 18 | 57 | 0.50 | 0.95 | 3.04 | 2.33 | 0.01 | 0 | 0.08 | 0.09 | 0.06 | 41.4 | 3.56 | 2.57 |
| 17 | 152 | 3563 | 0.40 | 0.33 | 4.70 | 5.39 | 0.01 | 0.02 | 0.55 | 1.09 | 1.12 | 15.65 | 3.35 | 6.6 |
| 18 | 404 | 336 | 1.03 | 0.33 | 1.56 | 4.75 | 0 | 0.01 | 0.23 | 0.3 | 2.22 | 86.27 | 3.53 | 6.52 |
| 19 | 78 | 379 | 0.65 | 0.71 | 2.34 | 2.18 | 0.01 | 0.01 | 0.19 | 0.14 | 2.86 | 38.48 | 3.88 | 9.12 |
| 20 | 1,205,052 | 592 | 0.81 | 1.01 | 2.42 | 1.52 | 0.02 | 0 | 19.83 | 0.29 | 11.8 | 502.66 | 5.15 | 2.92 |
| 21 | 20 | 106 | 0.30 | 0.57 | 5.58 | 3.35 | 0 | 0.01 | 0.12 | 0.15 | 0.16 | 32.52 | 3.59 | 9.51 |
| 22 | 227 | 67 | 0.42 | 0.31 | 4.56 | 4.71 | 0 | 0 | 0.5 | 0.17 | 0.74 | 0.47 | 5.1 | 3.91 |
| Median | 1927 | 1710 | 0.59 | 0.72 | 3.19 | 3.01 | 0.012 | 0.009 | 0.29 | 0.37 | 2.58 | 529.84 | 4.95 | 5.79 |
*Used trim mean 5% (trimmed mean of the values in X is the mean of X, excluding the highest and lowest K data values, where K = N*(PERCENT(=5)/100)/2 and where N is the number of values in X. Values reported for each subject is the average value of the metric over all 8 channels and entire recording.
Concordance of clinical information obtained with the rapid response EEG and conventional EEG.
| Time | Rapid response EEG Finding | Conventional EEG Finding | |
|---|---|---|---|
| 1 | 21 | Focal seizure in the right occipital region | Moderate diffuse slowing. Generalized rhythmic delta activity with frontal intermittent rhythmic delta activity. |
| 2 | 43 | Focal right hemispheric seizure (12 min 04 s into recording) | Moderate to severe diffuse slowing. Patient was loaded with Keppra 1000 mg prior to conventional EEG. |
| 3 | 134 | Diffuse slowing | Diffuse slowing |
| 4 | 47 | Diffuse slowing, missing data from electrode 8 (on bandage) | Diffuse slowing |
| 5 | 8 | Diffuse slowing | Diffuse slowing |
| 6 | 7 | Diffuse slowing | Diffuse slowing with left frontal epileptiform discharges and 4 electrographic left frontal seizures |
| 7 | 21 | Excessive beta, otherwise mildly slow | Normal awake and asleep EEG |
| 8 | 74 | Slow bilaterally but with excessive movement artifact especially in the posterior channels | Moderate to severe slowing with bifrontal sharply contoured transients |
| 9 | 81 | Diffuse slowing | Diffuse slowing |
| 10 | 115 | Diffuse slowing | Diffuse slowing with Generalized Periodic Discharges |
| 11 | 44 | Generalized periodic discharges on suppressed background with excessive muscle artifact | Myoclonic status epilepticus with intervening burst suppression |
| 12 | 2 | Diffuse slowing | Diffuse slowing |
| 13 | 90 | Diffuse slowing | Diffuse slowing |
| 14 | 25 | Diffuse slowing | Diffuse slowing with additional right fronto-parietal slowing and suppression |
| 15 | 92 | Diffuse slowing | Diffuse slowing |
| 16 | 25 | Diffuse slowing with GPDs | Diffuse slowing with episodes of generalized rhythmic slowing (GRDA), and rare runs of non-specific generalized periodic discharges (GPDs) |
| 17 | 76 | Diffuse slowing | Intermittent epileptiform discharges over the left and right frontal regions, less frequently over the bitemporal regions, and sometimes in the form of lateralized periodic discharges, mild to moderate diffuse slowing with additional intermittent focal slowing over the right |
| 18 | 70 | Diffuse slowing with GPDs | Diffuse slowing, including rhythmic slow patterns with additional intermittent rhythmic slowing and nonspecific periodic discharges over the left frontal region |
| 19 | 76 | Right > left slow and stimulus induced rhythmic activity | Diffuse slowing with stimulus induced rhythmic waveforms upon noxious stimulation |
| 20 | 115 | Diffuse slowing with sharply contoured waveforms (L > R) | Diffuse slowing with brief periods of discontinuities and non-specific blunted generalized discharges, occasionally in runs (Generalized Periodic Discharges) |
| 21 | 33 | Diffuse slowing | Diffuse slowing |
| 22 | 117 | Diffuse slowing | Occasional runs of Generalized Periodic Discharges (GPDs) upon stimulation with diffuse slowing |
Time = duration of rapid response EEG recording.