Background/Aims: We used 3-dimensional (3D) printing technology to create a new hemostasis simulator for the stomach and investigated its efficacy and realism in endoscopic hemostasis training.Entities:
Keywords: Education; Endoscopy; Hemostasis; Printing; Simulator; three-dimensional
Mesh:
Year: 2019 PMID: 30970441 PMCID: PMC6622563 DOI: 10.5009/gnl18389
Source DB: PubMed Journal: Gut Liver ISSN: 1976-2283 Impact factor: 4.519
Fig. 1Hemostasis simulator composed of a hemoclipping module and injection module. (A) Artificial blood is supplied to the simulator through a pump. Any conventional endoscopic equipment can be used. (B) Two hemostasis modules are incorporated into the main body.
Fig. 2Hemoclipping module. (A) The hemoclipping module is composed of 16 protruding parts. The protruding parts are sealed with a metal pin blocking the artificial blood. (B) When the metal pin is removed from the protruding hole, artificial blood begins pumping in a manner similar to that of actual blood from a vessel. (C) When the hemoclip is properly positioned, the flow of artificial blood stops.
Fig. 3Injection module. (A) The inner part of the injection module has two stacked chambers and the outer part is made of the methacrylic acid ester cured by photoinitiator. (B, C) When the saline is injected into the outer chamber, the flow of artificial blood is diminished and stops if outer chamber pressure is high enough.
Mean Scores of the Answer to the Question for the Hemosta-sis Simulator Using a 7-Point Likert Scale
| Index | Hemoclipping | Injection |
|---|---|---|
| Handling of simulator is similarity with actual case | 6.1±0.4 | 6.4±0.5 |
| Simulator is reasonable for hemostasis training | 6.0±0.5 | 6.4±0.4 |
| Working with the simulator will improves trainee’s skills | 6.3±0.5 | 6.7±0.5 |
| Training with the simulator reduces the risk for patients | 6.1±0.7 | 6.9±0.5 |
| I recommend that the simulator be used in hemostasis training | 6.1±0.7 | 6.9±0.5 |
Data are presented as the mean±SD. Scale, 1 (strongly disagree) to 7 (strongly agree).
Fig. 4Procedure times for each trial for each group. (A) Hemoclipping module and (B) injection module.
Procedure Completing Times with the New Hemostasis Simulator According to the Experience Level of the Endoscopist
| Hemoclipping | Injection | |||
|---|---|---|---|---|
|
|
| |||
| Beginner group (n=11) | Expert group (n=10) | Beginner group (n=11) | Expert group (n=10) | |
| 1st trial | 116.1±15.3 | 25.1±1.2 | 161.3±13.4 | 46±3.1 |
| 2nd trial | 96.4±6.1 | 21.2±1.9 | 118.2±9.3 | 36.4±2 |
| 3rd trial | 65.8±3.3 | 18±1 | 77±10.7 | 33.6±1.6 |
| 4th trial | 54.8±6.9 | 17.7±1 | 61.1±8.2 | 32.9±2.2 |
| 5th trial | 30.5±4.1 | 16.6±1 | 43±7.2 | 32.4±1.5 |
| Mean of 5 trials | 72.7±7.1 | 19.7±1.2 | 92.1±9.8 | 36.3±2 |
Data are presented as mean±SD. Time expressed in seconds.