Test-retest reliability of a hip strength assessment system in varsity soccer players.

Gauthier Desmyttere1, Sylvain Gaudet2, Mickael Begon2.   

Abstract

OBJECTIVE: To investigate test-retest reliability of a hip strength assessment system (GroinBar).
DESIGN: Test-retest reliability study.
SETTING: Laboratory. PARTICIPANTS: Twenty asymptomatic varsity soccer players. MAIN OUTCOME MEASURES: Maximal isometric hip strength (adduction, abduction, internal and external rotation, flexion and extension) was assessed using the GroinBar. Intraclass correlation coefficient (ICC) and relative standard error of measurement (SEM) were calculated to evaluate reliability of peak (ICC3,1) (highest peak within 3 trials) and average peak (ICC3,3) (average of 3 trials) force and rate of force development (RFD). Hotelling's T2, were also used to compare bilateral and reciprocal ratios between dominant and non-dominant leg.
RESULTS: ICC for both peak force and RFD values revealed moderate to good reliability (0.53-0.88 and 0.61-0.84, respectively), whereas reliability was good to excellent regarding their average values (0.77-0.95 and 0.81-0.92, respectively). SEM of average peak force and RFD values (4.1-9.4% and 8.2-13.9%, respectively) were lower than that of peak force and RFD values (5.7-13.0% and 10.7-19.1%, respectively). No significant difference was found in bilateral and reciprocal force ratios between dominant and non-dominant leg.
CONCLUSIONS: The GroinBar is a reliable tool to assess hip muscle function in athletic populations and could be used for player screening and follow-up.
Copyright © 2019 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Hip; Muscle function; Reliability; Soccer

Year:  2019        PMID: 30959443     DOI: 10.1016/j.ptsp.2019.03.013

Source DB:  PubMed          Journal:  Phys Ther Sport        ISSN: 1466-853X            Impact factor:   2.365


  2 in total

1.  Reliability of hand-held dynamometer in measuring gluteus medius isometric muscle strength in healthy population.

Authors:  Rabail Rani Soomro; Hossein Karimi; Syed Amir Gillani
Journal:  Pak J Med Sci       Date:  2022 May-Jun       Impact factor: 2.340

2.  Reliability of hip muscle strength measured in principal and intermediate planes of movement.

Authors:  Basilio A M Goncalves; David J Saxby; Adam Kositsky; Rod S Barrett; Laura E Diamond
Journal:  PeerJ       Date:  2021-06-02       Impact factor: 2.984

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.