N95 filtering facepiece respirators do not reliably afford respiratory protection during chest compression: A simulation study.

Sung Yeon Hwang1, Hee Yoon2, Aerin Yoon3, Taerim Kim1, Guntak Lee1, Kwang Yul Jung1, Joo Hyun Park1, Tae Gun Shin1, Won Chul Cha1, Min Seob Sim1, Seonwoo Kim4.   

Abstract

BACKGROUND: N95 filtering facepiece respirators (N95 respirators) may not provide adequate protection against respiratory infections during chest compression due to inappropriate fitting.
METHODS: This was a single-center simulation study performed from December 1, 2016, to December 31, 2016. Each participant underwent quantitative fit test (QNFT) of N95 respirators according to the Occupational Safety and Health Administration protocol. Adequacy of respirator fit was represented by the fit factor (FF), which is calculated as the number of ambient particles divided by the number inside the respirator. We divided all participants into the group that passed the overall fit test but failed at least one individual exercise (partially passed group [PPG]) and the group that passed all exercises (all passed group [APG]). Then, the participants performed three sessions of continuous chest compressions, each with a duration of 2 min, while undergoing real-time fit testing. The primary outcome was any failure (FF < 100) of the fit test during the three bouts of chest compression.
RESULTS: Forty-four participants passed the QNFT. Overall, 73% (n = 32) of the participants failed at least one of the three sessions of chest compression; the number of participants who failed was significantly higher in the PPG than in the APG (94% vs. 61%; p = 0.02). Approximately 18% (n = 8) of the participants experienced mask fit failures, such as strap slipping.
CONCLUSIONS: Even if the participants passed the QNFT, the N95 respirator did not provide adequate protection against respiratory infections during chest compression.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Chest; Compression; Infection control; Quantitative fit test; Respiratory protective devices

Mesh:

Year:  2019        PMID: 30955924      PMCID: PMC7125572          DOI: 10.1016/j.ajem.2019.03.041

Source DB:  PubMed          Journal:  Am J Emerg Med        ISSN: 0735-6757            Impact factor:   2.469


Introduction

Emergency departments (EDs), the principal portals of entry into healthcare systems, are increasingly required to screen and treat patients with communicable infections [1,2]. In 2015, a large outbreak of Middle East respiratory syndrome (MERS) coronavirus infection occurred after exposure to a single patient in an overcrowded ED of a hospital in South Korea; several individuals, including healthcare workers (HCWs), were infected [3,4]. The World Health Organization and the U.S. Centers for Disease Control and Prevention recommend that HCWs must use a particulate filtering facepiece respirator that is at least as protective as the National Institute for Occupational Safety and Health (NIOSH)-certified N95 filtering facepiece respirator (N95 respirator) or its equivalent when treating patients with airborne infectious diseases [5,6]. However, such facepiece respirators can provide protection only when the face seal fits tightly. Therefore, fit testing is essential; a qualitative or quantitative approach must be used to identify respirators that best suit each individual [7,8]. The quantitative fit test (QNFT) objectively determines the adequacy of respirator fit by measuring leakage around the face seal using the respirator fit tester [9,10]. This device measures the fit factor (FF) (the number of ambient particles divided by the number inside the respirator when simulating eight workplace activities). An overall FF ≥100 is considered the passing level [9]. However, although the overall FF may be ≥100, the FFs for individual exercises may be <100 (for example, during bending). Cardiopulmonary resuscitation (CPR), a common ED procedure, generates infectious aerosols, and this is associated with an increased risk of pathogen transmission to HCWs [11,12]. Some exercises for the conventional QNFT mimic chest compression, which include bending at the waist and head up-and-down movement [9]. However, chest compression during CPR is significantly more dynamic and rapid than QNFT exercises. Therefore, it is unclear whether the protective effects of the N95 respirators will be maintained during chest compression particularly in those who failed at least one individual conventional QNFT exercise. Previously, Shin et al. have evaluated the effects of movements during chest compression on the protective performances of various N95 respirators in the simulated setting [13]. They demonstrated that the FFs of certain respirators decreased during chest compression, thereby seriously compromising respiratory performance. No study has yet explored the stability of N95 respirators in a group that only partially ‘passed’ the QNFT. Thus, we compared the respirator failure (FF < 100) rates during chest compression between a partially passed group (the overall fit factor was adequate, but at least one specific exercise was failed; the PPG) and an all passed group (group that passed all exercises: the APG).

Methods

Study design and setting

This was a single-center simulation study that explored potential issues that may be encountered when wearing an N95 respirator during chest compression. The study was conducted in a laboratory in Samsung Medical Center (a tertiary, university-affiliated, referral hospital located in a large city in Korea) from December 1, 2016, to December 31, 2016. The temperature and humidity of the room were controlled at approximately 23 °C and 30%, respectively, to minimize the impact of environmental factors on outcomes. The institutional review board of our institution approved the study, and a written informed consent was obtained from each participant.

Selection of participants

The inclusion criteria were as follows: HCWs aged ≥20 years, those certified for the delivery of basic life support or advanced cardiovascular life support by the American Heart Association (AHA) or those who had completed our institutional training program, and those who delivered CPR in the clinical field [14]. The exclusion criteria were as follows: HCWs who were pregnant, those with any musculoskeletal diseases that compromised the capacity to deliver chest compression, and those with medical conditions, including asthma, congestive heart failure, or coronary heart disease. Moreover, the participants who failed the fit tests for all three respirators were not included.

Methods of measurements

Preparation for simulation

At the beginning of the simulation, the investigators conducted a brief training session for the participants, which included providing instructions for the overall flow of the study using slides and via demonstration and practice in using the standardized N95 respirator donning technique. They were instructed to complete questionnaires about the demographic characteristics of the participants after the training session. Then, every participant took the QNFT for the N95 respirator. We divided the participants into two groups (PPG and APG) (Fig. 1 ).
Fig. 1

Flow chart of the study. QNFT, quantitative fit test.

Flow chart of the study. QNFT, quantitative fit test.

N95 respirator

Since the MERS epidemic in 2015, all employees in our institution must be fit tested using three N95 respirators: the 1860 and 1870+ (3 M, St. Paul, MN) and the 46,727 (Kimberly Clark, Irving, TX). The best-fitting respirator is identified based on the QNFT FF and comfort of the person who wears such device. We selected respirators based on earlier data available for all participants. However, if the shape of the face had changed because of weight change, plastic surgery, or dental correction, the fit test was repeated.

QNFT

We used the PortaCount Pro+ 8038 Respirator Fit Tester (TSI Inc., Shoreview, MN) for QNFT. The QNFT for the N95 respirator was conducted according to the Occupational Safety and Health Administration (OSHA) protocol [9,15]. While wearing the respirator, eight test exercises were performed in the following order: normal breathing, deep breathing, head side-to-side moving, head up-and-down motion, talking, grimacing, bending over, and normal breathing. For the talking exercise, the participants read identical text prepared in advance. Each exercise was performed for 1 min except for grimacing (15 s). The FF of grimacing was excluded from the final calculation according to the OSHA protocol [9]. An FF > 200 was scored as 200 by the tester. All FFs were continuously monitored and considered passing if the final score was ≥100.

Real-time QNFT during chest compression

After the fit test, the participants were instructed to perform continuous chest compression on a Resusci Anne mannikin (Laerdal Medical, Stavanger, Norway) three times for 2 min each (with 4-min rest between each session) while undergoing further fit testing (Fig. 1); they were not allowed to touch or adjust the mask. If the mask strap loosened, it was re-adjusted during the break. To ensure that the CPR was of high quality, which is in accordance with the 2015 AHA guidelines, all data were collected using a Laerdal PC Skill Reporting System (Laerdal Medical) [16]. One investigator provided feedback to all participants in real time while watching the computer monitor. The participants rested for 20 min after the three sessions of chest compression, followed by an additional compression for 2 min while wearing the same respirator after performing a user-seal-check.

Outcomes

The primary outcome was any failure (FF < 100) of the fit test during the three bouts of chest compression. The secondary outcome was the slipping down of respirator during chest compression.

Data analyses

A priori sample size calculations were made in terms of primary outcome achievement; we assumed an α value of 0.05 for two-sided hypothesis testing and a β error of 0.20 (power = 80%). A preliminary study of 10 participants has revealed a failure rate of 50%. We considered that a 40 percentage point increase in the failure rate was clinically significant; we assumed that the PPG might evidence failure. A total of 42 patients were required to detect this hypothesized failure rate. Standard descriptive statistics were used to present all data. Continuous variables were provided as medians with interquartile ranges (IQRs), and the Wilcoxon rank-sum test was used for comparisons. Categorical data were presented as numbers with percentages and compared using the chi-square test. STATA version 13.0 software (STATA Corporation, College Station, TX) was used to perform all statistical analyses.

Results

Characteristics of the study participants

We recruited 45 participants, of whom 1 was excluded due to failure in the baseline QNFT using all three N95 respirators; 44 participants were ultimately included, of whom 66% (n = 29) were female. The median age of the participants was 31 (IQR: 26.5–36) years. Most participants were nurses (52%, n = 23) with an average clinical experience of 6 (IQR: 3–10) years. Baseline data, such as age, sex, career duration, body mass index, occupation, CPR training, and respirator type, did not significantly differ between the two groups (Table 1 ). The 3M 1870 + N95 respirator (n = 25, 57%) was most frequently used, followed by the Kimberly Clark 46727 (n = 15, 34%) and the 3M 1860 (n = 4, 9%).
Table 1

Baseline characteristics of the participants.

Total (n = 44)Partially passed (n = 16)All passed (n = 28)p-Value
Age (years)31 (26.5–36)34 (26.5–37.5)30.5 (26.5–35)0.34
Sex, female (%)29 (66)10 (63)19 (68)0.72
Career (years)6 (3−10)6.5 (3−12)6 (2.5–8.5)0.31
BMI (kg/m2)20.8 (19.8–23.9)21.0 (19.0–23.2)20.7 (20.2–24.0)0.45
Occupation0.93
 Medical doctor14 (32)5 (31)9 (32)
 Nurse23 (52)8 (50)15 (54)
 Emergency medical technician7 (16)3 (19)4 (14)
CPR training0.94
 ACLS or BLS provider36 (82)13 (81)23 (82)
 Institutional program only8 (18)3 (19)5 (18)
Respirator model0.41
 3M 1870+25 (57)7 (44)18 (64)
 3M 18604 (9)2 (13)2 (7)
 KIMBERLY CLARK 4672715 (34)7 (44)8 (29)

Data were presented as median with interquartile range or n (%). BMI, body mass index; CPR, cardiopulmonary resuscitation; ACLS, advanced cardiovascular life support; BLS, basic life support.

Baseline characteristics of the participants. Data were presented as median with interquartile range or n (%). BMI, body mass index; CPR, cardiopulmonary resuscitation; ACLS, advanced cardiovascular life support; BLS, basic life support.

Quality of chest compression

No significant differences were observed in the quality of chest compression between the two groups except in terms of compression rate during the second cycle (PPG vs. APG: 113.5, IQR: 109–116.5 vs. 108.5, IQR: 105–114, p = 0.04) (Table 2 ).
Table 2

Quality of chest compression

Total (n = 44)Partially passed (n = 16)All passed (n = 28)P-value
First chest compression
Mean rate, counts/min108 (104.5–113)111.5 (107.5–114.5)107 (104–112)0.12
Mean depth, mm54 (50–57)53.5 (48–56.5)54.5 (51–57)0.70
Complete relaxation rate, %100 (100−100)100 (100–100)100 (100–100)0.89
Correct hand position rate, %100 (100–100)100 (100–100)100 (100–100)0.78



Second chest compression
Mean rate, counts/min111 (106–114.5)113.5 (109–116.5)108.5 (105–114)0.04
Mean depth, mm55 (51.5–58)55 (49–58)55 (52–58)0.93
Complete relaxation rate, %100 (100–100)100 (100–100)100 (100–100)0.59
Correct hand position rate, %100 (100–100)100 (94–100)100 (100–100)0.59



Third chest compression
Mean rate, counts/min111 (105.5–114)111.5 (108.5–116)110 (103.5–114)0.11
Mean depth, mm55 (50–57.5)55 (50–57)55 (48.5–57.5)0.86
Complete relaxation rate, %100 (100–100)100 (100–100)100 (100–100)0.18
Correct hand position rate, %100 (100–100)100 (100–100)100 (100–100)0.93



Chest compression after user-seal-check
Mean rate, counts/min111.5 (106–115)110.5 (107–115)113 (106–115)0.73
Mean depth, mm54 (51–58)54 (53–56)54 (50.5–59)0.89
Complete relaxation rate, %100 (100–100)100 (100–100)100 (100–100)0.89
Correct hand position rate, %100 (100–100)99.5 (100–100)100 (100–100)0.23

Data were presented as median with interquartile range or n (%).

Quality of chest compression Data were presented as median with interquartile range or n (%).

Outcome measures

The outcomes are shown in Table 3 and Fig. 2 . Overall, 73% (n = 32) of the participants failed at least one of the three chest compression sessions; the failure rate was significantly higher in the PPG than in the APG (94 vs. 61%; p = 0.02). In total, 18% (n = 8) of the participants experienced respirator failure, such as strap loosening. The overall failure rate of the fit test after user-seal-check was 59% (n = 26), and it was not different between the PPG and APG (63% vs 57%, p = 0.73).
Table 3

Outcomes

Total(n = 44)Partially passed(n = 16)All passed(n = 28)p-Value
Any failure32 (73)15 (94)17 (61)0.02
1st chest compression failure24 (55)12 (75)12 (43)0.04
2nd chest compression failure25 (57)13 (81)12 (43)0.01
3rd chest compression failure27 (61)14 (88)13 (46)<0.01
User-seal-check26 (59)10 (63)16 (57)0.73
Slipping down of strap8 (18)5 (31)3 (11)0.09

Data were presented as n (%).

Fig. 2

Outcomes. The plots represent the median fit factor with interquartile ranges of the baseline QNFT and three sets of chest compression. QNFT, quantitative fit test.

Outcomes Data were presented as n (%). Outcomes. The plots represent the median fit factor with interquartile ranges of the baseline QNFT and three sets of chest compression. QNFT, quantitative fit test.

Discussion

Even when the participants passed the QNFT, N95 respirators did not afford adequate protection during chest compression. Notably, in 94% of the participants in the PPG, the FF decreased to <100 during at least one session of chest compression. Facepiece respirators only work properly when the face seal is tight [17,18]. Body movements during chest compression are both dynamic and intense, and sweat further compromises respirator fit, creating a gap permitting disease transmission. Therefore, our findings are clinically significant in terms of the safety of HCWs who are at high risk for airborne disease transmission during chest compression even when wearing adequately fit-tested N95 respirators. The conventional QNFT is widely used to measure N95 respirator performance in HCWs. However, the exercises performed during the fit test are not similar to those executed in real-world setting. Suen et al. have used a portable aerosol spectrometer to evaluate the performance of N95 respirators during various nursing procedures, including suction and nasogastric tube insertion for 10 min [19]. The average FF decreased significantly from 184.85 to 134.71 after completing the procedures, and the FF fell to <100 in 33% of the participants. This study indicated that N95 respirators may not provide consistent protection against respiratory infection for HCWs. The QNFT failure rates during chest compression after the user-seal-check did not differ between the two groups. Several previous studies have suggested that this user-seal-check alone inadequately evaluates respirator fitting [20,21]. Nevertheless, in clinical practice, all treatments are performed after a user-seal-check alone. Therefore, our results suggest that pre-passing the N95 QNFT did not ensure respiratory safety during chest compressions. It is known that the risk of disease transmission from patients to rescuers during CPR is extremely low [22,23]. One review article has shown that the number of infections acquired during CPR is approximately <1/200,000 [24]. However, previous studies have mainly focused on diseases, such as acquired immune deficiency syndrome, hepatitis B, hepatitis C, and tuberculosis, which are transmitted via mouth-to-mouth ventilation or needlestick injury. HCWs can be infected in various ways when performing CPR. The actual incidences of infection among HCWs due to contact with patients with airborne disease during CPR are not fully elucidated. One retrospective cohort study has reported that one of nine HCWs who participated in cardiac compression developed severe acute respiratory syndrome (SARS) [25]. Of the six HCWs who performed CPR on a patient with MERS, one acquired the infection [11]. Experiences from the outbreaks of highly contagious diseases, such as SARS and MERS, taught us that HCWs should be protected from airborne disease transmission when performing CPR [11,12,26,27]. The current international CPR guidelines do not address the extent of airway protection required by HCWs when performing CPR on patients with suspected or confirmed airborne diseases [22]. Infection-prevention strategies for HCWs tend to take second place in life-threatening situations requiring minimization of no-flow time [11,12]. However, the safety of HCWs is in fact paramount. Mechanical compression devices can be used to minimize HCW participation in CPR. In addition, HCWs engaging in chest compression of patients with airborne diseases could wear powered air-purifying respirators with hoods (PAPRs) rather than N95 respirators. However, this may result in resuscitation difficulties, and both movement and communication are compromised [28,29]. In addition, the protective effects of PAPRs during chest compression have not been explored, and further studies are warranted.

Limitations

Our study had certain limitations. First, we aimed to maximally reflect actual clinical settings. However, we worked in a simulation laboratory, and the outcomes of the present study might differ from those of real-world settings. The participants performed continuous chest compression based on the assumption that the patient had an advanced airway. However, in patients without an advanced airway, chest compressions are briefly paused to provide ventilation [16]. CPR is complex, featuring chest compression, endotracheal intubation, defibrillation, bag-valve ventilation, intravenous line insertion, and drug administration. However, we focused on chest compression only. If the participants carried out other tasks, then the outcomes might have differed. In addition, in the present study, the participants were instructed not to talk as much as possible. We believed that such action was reasonable when providing CPR to patients with air-borne disease. However, HCWs need to talk for communication in real settings, which can loosen the fitting of the N95 respirator. Second, we had only three respirator types available; thus, our results cannot be generalized to other models. Third, although chest compression lasted for 2 min, the FF was obtained after excluding data from the first 20 s because the time was used for ambient purge, ambient sample, and mask purge.

Conclusion

Even in individuals who passed the initial fit test, N95 respirators did not provide adequate protection during chest compression. The participants in the PPG were at particular risk of airborne disease transmission during chest compression. Further study must be conducted to establish specific guidelines about the level of respiratory protection for HCWs during CPR of patients with airborne diseases.

Prior presentations

None.

Funding sources/disclosures

The authors have no potential conflicts of interest or funding sources to declare.

Author contribution

Conceptualization: Yoon H, Hwang SY. Simulation: Yoon H, Hwang S, Yoon A. Writing - original draft: Yoon H, Hwang SY. Writing - review & editing: Yoon H, Hwang SY, Yoon A, Kim T, Lee G, Jung KY, Park JH, Shin TG, Sim MS, Kim S. Study supervision and approval of final manuscript: Yoon H, Hwang SY.
  24 in total

Review 1.  Part 5: Adult Basic Life Support and Cardiopulmonary Resuscitation Quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care.

Authors:  Monica E Kleinman; Erin E Brennan; Zachary D Goldberger; Robert A Swor; Mark Terry; Bentley J Bobrow; Raúl J Gazmuri; Andrew H Travers; Thomas Rea
Journal:  Circulation       Date:  2015-11-03       Impact factor: 29.690

2.  Comparison of two quantitative fit-test methods using N95 filtering facepiece respirators.

Authors:  Margaret Sietsema; Lisa M Brosseau
Journal:  J Occup Environ Hyg       Date:  2016-08       Impact factor: 2.155

3.  Reliability of N95 respirators for respiratory protection before, during, and after nursing procedures.

Authors:  Lorna K P Suen; Lin Yang; Suki S K Ho; Keith H K Fung; Maureen V Boost; Cynthia S T Wu; Cypher H Au-Yeung; Margaret O'Donoghue
Journal:  Am J Infect Control       Date:  2017-05-16       Impact factor: 2.918

4.  Healthcare worker infected with Middle East Respiratory Syndrome during cardiopulmonary resuscitation in Korea, 2015.

Authors:  Hae-Sung Nam; Mi-Yeon Yeon; Jung Wan Park; Jee-Young Hong; Ji Woong Son
Journal:  Epidemiol Health       Date:  2017-11-12

5.  The Utility of Preliminary Patient Evaluation in a Febrile Respiratory Infectious Disease Unit outside the Emergency Department.

Authors:  Jun Sik Kang; Byung Woo Jhun; Hee Yoon; Seong Mi Lim; Eunsil Ko; Joo Hyun Park; Sung Yeon Hwang; Se Uk Lee; Tae Rim Lee; Won Chul Cha; Tae Gun Shin; Min Seob Sim; Ik Joon Jo
Journal:  J Korean Med Sci       Date:  2017-09       Impact factor: 2.153

6.  MERS-CoV outbreak following a single patient exposure in an emergency room in South Korea: an epidemiological outbreak study.

Authors:  Sun Young Cho; Ji-Man Kang; Young Eun Ha; Ga Eun Park; Ji Yeon Lee; Jae-Hoon Ko; Ji Yong Lee; Jong Min Kim; Cheol-In Kang; Ik Joon Jo; Jae Geum Ryu; Jong Rim Choi; Seonwoo Kim; Hee Jae Huh; Chang-Seok Ki; Eun-Suk Kang; Kyong Ran Peck; Hun-Jong Dhong; Jae-Hoon Song; Doo Ryeon Chung; Yae-Jean Kim
Journal:  Lancet       Date:  2016-07-09       Impact factor: 79.321

7.  Sensitivity and specificity of the user-seal-check in determining the fit of N95 respirators.

Authors:  S C Lam; J K L Lee; S Y Yau; C Y C Charm
Journal:  J Hosp Infect       Date:  2011-01-13       Impact factor: 3.926

8.  Possible SARS coronavirus transmission during cardiopulmonary resuscitation.

Authors:  Michael D Christian; Mona Loutfy; L Clifford McDonald; Kennth F Martinez; Mariana Ofner; Tom Wong; Tamara Wallington; Wayne L Gold; Barbara Mederski; Karen Green; Donald E Low
Journal:  Emerg Infect Dis       Date:  2004-02       Impact factor: 6.883

9.  SARS transmission among hospital workers in Hong Kong.

Authors:  Joseph T F Lau; Kitty S Fung; Tze Wai Wong; Jean H Kim; Eric Wong; Sidney Chung; Deborah Ho; Louis Y Chan; S F Lui; Augustine Cheng
Journal:  Emerg Infect Dis       Date:  2004-02       Impact factor: 6.883

10.  Middle East Respiratory Syndrome Coronavirus Superspreading Event Involving 81 Persons, Korea 2015.

Authors:  Myoung-don Oh; Pyoeng Gyun Choe; Hong Sang Oh; Wan Beom Park; Sang-Min Lee; Jinkyeong Park; Sang Kook Lee; Jeong-Sup Song; Nam Joong Kim
Journal:  J Korean Med Sci       Date:  2015-10-16       Impact factor: 2.153

View more
  12 in total

1.  Quantitative Fit Tested N95 Respirator-Alternatives Generated With CT Imaging and 3D Printing: A Response to Potential Shortages During the COVID-19 Pandemic.

Authors:  David H Ballard; Udayabhanu Jammalamadaka; Kathleen W Meacham; Mark J Hoegger; Broc A Burke; Jason A Morris; Alexander R Scott; Zachary O'Connor; Connie Gan; Jesse Hu; Karthik Tappa; Richard L Wahl; Pamela K Woodard
Journal:  Acad Radiol       Date:  2020-11-21       Impact factor: 3.173

2.  Respiratory protection among healthcare workers during cardiopulmonary resuscitation in COVID-19 patients.

Authors:  Kurt Ruetzler; Jacek Smereka; Kobi Ludwin; Anna Drozd; Lukasz Szarpak
Journal:  Am J Emerg Med       Date:  2020-05-11       Impact factor: 2.469

3.  COVID-19 and cardiopulmonary resuscitation: an N95 respirator mask may not be adequate.

Authors:  Patrick Wong; Sharon Gk Ong; Wan Y Lim
Journal:  Br J Anaesth       Date:  2020-05-18       Impact factor: 9.166

4.  Recommendations on cardiopulmonary resuscitation strategy and procedure for novel coronavirus pneumonia.

Authors:  Wei Song; Yuanshui Liu; Yanhong Ouyang; Wenteng Chen; Min Li; Shuming Xianyu; Shengyang Yi
Journal:  Resuscitation       Date:  2020-04-07       Impact factor: 5.262

5.  Effectiveness of Elastomeric Half-Mask Respirators vs N95 Filtering Facepiece Respirators During Simulated Resuscitation: A Nonrandomized Controlled Trial.

Authors:  Andrew J Barros; Costi D Sifri; Taison D Bell; Joshua C Eby; Kyle B Enfield
Journal:  JAMA Netw Open       Date:  2021-03-01

6.  P2/N95 filtering facepiece respirators: Results of a large-scale quantitative mask fit testing program in Australian health care workers.

Authors:  Maxim Milosevic; Raaj Kishore Biswas; Lesley Innes; Martin Ng; Ali Mehmet Darendeliler; Alice Wong; Elizabeth Denney-Wilson
Journal:  Am J Infect Control       Date:  2021-12-29       Impact factor: 4.303

7.  Does a Modified Adhesive Respirator Improve the Face Seal for Health Care Workers Who Previously Failed a Fit Test?: A Pilot Study During the Coronavirus Disease 2019 Pandemic.

Authors:  Richa Wardhan; Meghan M Brennan; Holden L Brown; Trey B Creech
Journal:  A A Pract       Date:  2020-06

8.  A Personal Respirator to Improve Protection for Healthcare Workers Treating COVID-19 (PeRSo).

Authors:  Paul T Elkington; Alexander S Dickinson; Mark N Mavrogordato; Daniel C Spencer; Richard J Gillams; Antonio De Grazia; Sebastian Rosini; Diana J Garay-Baquero; Laura E Diment; Nitin Mahobia; Alexandra Mant; Tom Baynham; Hywel Morgan
Journal:  Front Med Technol       Date:  2021-06-10

9.  Powered Respirators Are Effective, Sustainable, and Cost-Effective Personal Protective Equipment for SARS-CoV-2.

Authors:  Alasdair Munro; Jacqui Prieto; Emmanouil Mentzakis; Mohammed Dibas; Nitin Mahobia; Peter Baker; Sarah Herbert; Trevor Smith; Matthew Hine; Joann Hall; Angie McClarren; Mike Davidson; Julie Brooks; Jane Fisher; David Griffiths; Hywel Morgan; Corrado Giulietti; Saul N Faust; Paul Elkington
Journal:  Front Med Technol       Date:  2021-10-14

10.  Drastic changes in the practice of end-of-life care during the COVID-19 pandemic.

Authors:  Talia K Ben-Jacob; Lars-Kristofer N Peterson
Journal:  J Crit Care       Date:  2021-10-22       Impact factor: 3.425

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.