Acute ascending aortic dissection due to transcatheter patent ductus arteriosus closure in the elderly: An extremely rare complication of transcatheter patent ductus arteriosus closure.

Hiroyuki Yamamoto1, Toshiro Shinke1, Hiromasa Otake1, Hidekazu Tanaka1, Kensuke Matsumoto1, Ken-Ichi Hirata1.   

Abstract

It is well known that patent ductus arteriosus (PDA) in adults, especially in the elderly, differs from that in pediatric patients. A 68-year-old woman with a PDA with focal calcification at the aortic orifice of the ampulla with a minimum diameter of 4.0 mm and length of 14.8 mm, was treated with a 10/8-mm Amplatzer duct occluder (ADO) (St. Jude Medical Corp, St. Paul, MN, USA). After device implantation, systolic blood pressure (BP) increased to approximately 220 mmHg from 130 mmHg. She experienced transient dyspnea from hypertensive heart failure, which improved through continuous infusion of anti-hypertensive agents. She suddenly felt pressure on her chest 12 h post-procedure and collapsed. Surgical thoracotomy revealed an ascending aortic dissection into the pericardial space. In retrospective review, the ADO may have been slightly deformed by fluoroscopy. The complication may have been triggered by the resilience caused by device deformation, damage to the aortic wall due to the aortic side of the device, uneven elasticity of the arterial wall, and uncontrolled excessively high blood pressure. <Learning objective: Although transcatheter patent ductus arteriosus (PDA) closure is an established, safe, and effective procedure when treating PDA of the elderly, wall damage due to the device may occur because of atherosclerotic changes different from that of younger patients and blood pressure will rise after closure. Therefore, it is necessary to carefully select the type and size of the device and to strictly control blood pressure in patients with a history of hypertension.>.

Entities:  

Keywords:  Amplatzer duct occluder; Aortic dissection; Patent ductus arteriosus; Transcatheter closure

Year:  2018        PMID: 30949247      PMCID: PMC6429547          DOI: 10.1016/j.jccase.2018.12.002

Source DB:  PubMed          Journal:  J Cardiol Cases        ISSN: 1878-5409


  3 in total

1.  Intracardiac Echocardiography as a Guide for Transcatheter Closure of Patent Ductus Arteriosus.

Authors:  Hironaga Yoshimoto; Maeda Yasuto; Tadashi Inoue; Yoshiyuki Kagiyama; Yozo Teramachi; Ryuta Takase; Yusuke Koteda; Yoshihiro Fukumoto; Motofumi Iemura; Kenji Suda
Journal:  J Interv Cardiol       Date:  2020-07-30       Impact factor: 2.279

2.  Acute aortic dissection as a late and fatal complication of transcatheter persistent ductus arteriosus occlusion: a case report.

Authors:  David W Baker; David S Celermajer
Journal:  Eur Heart J Case Rep       Date:  2020-12-11

3.  Surgical removal of part of an occluder to treat iatrogenic coarctation of the aorta: a case report.

Authors:  Zhongwei Sun; Dongxu Li; Yabo Wang; Qi An
Journal:  BMC Surg       Date:  2020-01-17       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.