Deformation and retentive force following in vitro cyclic fatigue of cobalt-chrome and aryl ketone polymer (AKP) clasps.

Ali Marie1, Andrew Keeling1, T Paul Hyde1, Brian R Nattress1, Sue Pavitt1, Ryan J Murphy2, Timothy J Shary2, Sean Dillon1, Cecilie Osnes1, David J Wood3.   

Abstract

OBJECTIVE: To compare the retention force of individual clasps made from cobalt chromium (CoCr) or new aryl ketone polymer (AKP) material, Ultaire™ AKP, following prolonged fatigue testing along ideal and non-ideal paths of removal and to assess 3D deformation of the active and passive clasp tips.
METHODS: CoCr and AKP clasps were manufactured in their standard, respective processes, digitally scanned prior to testing, then cycled 15,000 times over an e.max analogue crown in artificial saliva. Retentive load was measured in situ, as a function of cycles. Clasps were rescanned to assess deformation and along with their antagonists subjected to SEM to assess localised wear.
RESULTS: Distortion of the CoCr clasps was consistently larger than Ultaire™ AKP clasps, irrespective of removal path. CoCr clasps had significantly higher retentive forces than AKP clasps, for both removal paths. Ultaire™ AKP clasps showed a lower but relatively constant retentive force. The non-ideal path of removal affected retentive forces for both clasp materials. SEM showed localised removal of glaze for e.max crowns used with CoCr clasps. SIGNIFICANCE: Ultaire™ AKP clasps showed significantly less permanent deformation and lower retentive force than CoCr clasps. Unlike CoCr, the Ultaire™ AKP clasps did not work harden, nor had as large a reduction in retentive force and accompanying permanent deformation; the retentive force for the Ultaire™ AKP clasps was consistent over 15,000 cycles of fatigue mimicking prolonged clinical use. The AKP material was more robust; showing minimal deformation even in non-ideal paths of removal, as many patients would routinely use.
Copyright © 2019. Published by Elsevier Inc.

Entities:  

Keywords:  Aryl ketone polymer; Clasp; Cobalt chrome; Deformation; Denture; Fatigue; RPD; Removable partial denture; Retentive force

Year:  2019        PMID: 30948229     DOI: 10.1016/j.dental.2019.02.028

Source DB:  PubMed          Journal:  Dent Mater        ISSN: 0109-5641            Impact factor:   5.304


  6 in total

1.  Moldova Region Population Study of Social Cases Affected by Partial Edentation.

Authors:  Florinel Cosmin Bida; Doriana Agop-Forna; Cosmin Ionuţ Creţu; Walid Edlibi Al Hage; Norina Consuela Forna
Journal:  Curr Health Sci J       Date:  2022-03-31

2.  Comparison between Integrated and Parallel Interlock Designs of an Extracoronal Attachment-retained Distal Extension Removable Partial Dentures: A Clinical Trial.

Authors:  Maria R Reslan; Essam Osman; Lucette Segaan; Mohammad Rayyan; Christelle Joukhadar; Mohamed Fattouh
Journal:  J Int Soc Prev Community Dent       Date:  2021-01-30

3.  Retention Assessment of High Performance Poly-etheretherketone Removable Partial Denture Frameworks Constructed by Various Techniques (in vitro Study).

Authors:  Nesreen El Mekawy; Mohamed Elgamal
Journal:  J Dent (Shiraz)       Date:  2021-12

4.  Performance of Graphene-Based and Polyether-Ether-Ketone Polymers as Removable Partial Denture Esthetic Clasp Materials after Cyclic Fatigue.

Authors:  Mostafa Omran Hussein
Journal:  Polymers (Basel)       Date:  2022-07-23       Impact factor: 4.967

5.  Examination of inter-rater and intra-rater reliability during retentive force measurement of different clasps using the developed small-sized retentive force measurement device.

Authors:  Hitoshi Akiyama; Maiko Sakamoto; Ryoichi Akama; Jun Takei
Journal:  BMC Oral Health       Date:  2020-08-20       Impact factor: 2.757

6.  Is the high-performance thermoplastic polyetheretherketone indicated as a clasp material for removable dental prostheses?

Authors:  Danka Micovic; Felicitas Mayinger; Sebastian Bauer; Malgorzata Roos; Marlis Eichberger; Bogna Stawarczyk
Journal:  Clin Oral Investig       Date:  2020-10-07       Impact factor: 3.573

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.