Literature DB >> 30947632

Initial surgical and clinical experience with the Nucleus CI532 slim modiolar electrode in the UK.

Manuel Gomez Serrano1, Sharmila Patel1, Robert Harris1, David Selvadurai1.   

Abstract

OBJECTIVE: The goal of this work is to describe the first experience in the UK with the slim pre-curved perimodiolar electrode Nucleus CI532 in a continuous series of patients in terms of surgical and clinical reliability and early performance outcomes.
METHOD: In this retrospective review we describe the complication rate (including electrode array tip fold-over), NRT thresholds, hearing preservation, power efficiency and CI performance outcomes in a continuous series of 40 cochlear implants CI532 performed between October 2016 and November 2017 in 17 adults and 13 children with severe to profound hearing loss.
RESULTS: Preliminary data from these groups reveals some low-frequency hearing preservation in the CI532 group although none of the patients were conventional hearing preservation candidates. NRT thresholds, power efficiency, and BKB sentences in quiet were measured at 3 and 6 months post activation. There were no significant differences in these results. The average BKB score in quiet increases from 22% pre-operatively to 58% at 3 months and 70% at 6 months. In addition, although hearing preservation was not an objective, low-frequency thresholds were preserved in 20% of cases at 3 and 6 months post-operatively. Complications were observed in 5 cases, one case with non-device related aerocoele and four related to the device array: two cases of tip roll over, one case of the electrode array being placed extra-cochlea, and one case with the electrode buckling into the middle ear. The last 2 cases were dealt with per-operatively. DISCUSSION: Our preliminary results with the CI532 implant indicate that it may be reliably placed with standard surgical techniques but care is needed during the deployment of the electrode. Further initial data suggest that switch on and early electrophysiological measures are comparable to the existing CI 512 device. However whilst preliminary, our data suggest that it may be possible to use this electrode for hearing preservation. However, further studies are required to determine its definitive advantage over other electrode designs.
CONCLUSION: CI532 is a reliable device offering good initial results and could be an option for hearing preservation although further studies are required.

Entities:  

Keywords:  CI532; Cochlear Implant; Hearing Preservation; Tip Fold Over

Year:  2019        PMID: 30947632     DOI: 10.1080/14670100.2019.1597461

Source DB:  PubMed          Journal:  Cochlear Implants Int        ISSN: 1467-0100


  8 in total

1.  Molecular aetiology of ski-slope hearing loss and audiological course of cochlear implantees.

Authors:  Yehree Kim; Jin Hee Han; Hyo Soon Yoo; Byung Yoon Choi
Journal:  Eur Arch Otorhinolaryngol       Date:  2022-02-25       Impact factor: 3.236

2.  Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode.

Authors:  Sang-Yeon Lee; Jin Hee Han; Marge Carandang; Yun Jung Bae; Byung Yoon Choi
Journal:  Sci Rep       Date:  2020-10-19       Impact factor: 4.379

3.  Effect of initial switch-on within 24 hours of cochlear implantation using slim modiolar electrodes.

Authors:  Woongsang Sunwoo; Hyoung Won Jeon; Byung Yoon Choi
Journal:  Sci Rep       Date:  2021-11-23       Impact factor: 4.379

4.  Suitable Electrode Choice for Robotic-Assisted Cochlear Implant Surgery: A Systematic Literature Review of Manual Electrode Insertion Adverse Events.

Authors:  Paul Van de Heyning; Peter Roland; Luis Lassaletta; Sumit Agrawal; Marcus Atlas; Wolf-Dieter Baumgartner; Kevin Brown; Marco Caversaccio; Stefan Dazert; Wolfgang Gstoettner; Rudolf Hagen; Abdulrahman Hagr; Greg Eigner Jablonski; Mohan Kameswaran; Vladislav Kuzovkov; Martin Leinung; Yongxin Li; Andreas Loth; Astrid Magele; Robert Mlynski; Joachim Mueller; Lorne Parnes; Andreas Radeloff; Chris Raine; Gunesh Rajan; Joachim Schmutzhard; Henryk Skarzynski; Piotr H Skarzynski; Georg Sprinzl; Hinrich Staecker; Timo Stöver; Dayse Tavora-Viera; Vedat Topsakal; Shin-Ichi Usami; Vincent Van Rompaey; Nora M Weiss; Wilhelm Wimmer; Mario Zernotti; Javier Gavilan
Journal:  Front Surg       Date:  2022-03-24

5.  The influence of Slim Modiolar electrode on residual hearing in pediatric patients.

Authors:  Karolina Haber; Alexandra Neagu; Wiesław Konopka; Katarzyna Amernik; Dan Cristian Gheorghe; Maria Drela; Iwona Wrukowska-Niemczewska; Józef Mierzwiński
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-09-08       Impact factor: 2.503

Review 6.  The Evolution of Care of Cancers of the Head and Neck Region: State of the Science in 2020.

Authors:  Flora Yan; Hannah M Knochelmann; Patrick F Morgan; John M Kaczmar; David M Neskey; Evan M Graboyes; Shaun A Nguyen; Besim Ogretmen; Anand K Sharma; Terry A Day
Journal:  Cancers (Basel)       Date:  2020-06-11       Impact factor: 6.639

7.  Outcomes for a clinically representative cohort of hearing-impaired adults using the Nucleus® CI532 cochlear implant.

Authors:  Matthias Hey; Nicole Neben; Timo Stöver; Uwe Baumann; Alexander Mewes; Tim Liebscher; Mark Schüssler; Antje Aschendorff; Thomas Wesarg; Andreas Büchner; Paula Greenham; Ulrich Hoppe
Journal:  Eur Arch Otorhinolaryngol       Date:  2020-03-05       Impact factor: 2.503

8.  Evaluation of hearing preservation in adults with a slim perimodiolar electrode.

Authors:  Sonja Ludwig; Niklas Riemann; Stefan Hans; Florian Christov; Johannes Maximilian Ludwig; Judith Saxe; Diana Arweiler-Harbeck
Journal:  Eur Arch Otorhinolaryngol       Date:  2021-04-08       Impact factor: 2.503

  8 in total