Literature DB >> 30942506

Validation of a novel smartphone application-enabled, patient-operated skin barrier device.

Erin E Grinich1, Anuja V Shah2, Eric L Simpson2.   

Abstract

BACKGROUND: Transepidermal water loss (TEWL) and surface capacitance measure skin barrier permeability and stratum corneum (SC) hydration, respectively, and are frequently utilized in atopic dermatitis clinical trials. Many barrier devices are costly and often used only in the academic setting. GPSkin is a low-cost, patient-operated device that measures both TEWL and SC hydration. This study aimed to test the reliability of GPSkin and assess its correlation with current industry standards.
MATERIALS AND METHODS: GPSkin was compared to the Biox AquaFlux (TEWL) and Courage-Khazaka Corneometer (SC hydration). Participants with healthy skin (n = 50) collected measurements with GPSkin in Trial 1 without any device education and in Trial 2 with additional instruction. In Trial 2, the investigator also performed measurements with GPSkin. Spearman's coefficients (rs ) were performed to assess device correlation. Intraclass correlation coefficients (ICC) were calculated to determine reliability.
RESULTS: Overall, GPSkin was moderately correlated with current industry device measurements for TEWL (Trial 1 rs :0.48; Trial 2 rs :0.40 participant, 0.34 investigator) and SC hydration (Trial 1 rs :0.63; Trial 2 rs :0.45). GPSkin demonstrated "good" test-retest reliability for both TEWL (ICC: 0.89) and SC hydration (ICC: 0.85) measurements when participants were provided with some device education. There was no difference in reliability between participants provided with device education and investigators.
CONCLUSION: Based on these findings, we concluded that GPSkin provides reasonably precise and reliable measurements of SC hydration and TEWL as compared to current devices.
© 2019 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  atopic dermatitis; barrier; capacitance; device; hydration; transepidermal water loss

Mesh:

Year:  2019        PMID: 30942506     DOI: 10.1111/srt.12692

Source DB:  PubMed          Journal:  Skin Res Technol        ISSN: 0909-752X            Impact factor:   2.365


  4 in total

1.  Reliable, low-cost, fully integrated hydration sensors for monitoring and diagnosis of inflammatory skin diseases in any environment.

Authors:  Surabhi R Madhvapathy; Heling Wang; Jessy Kong; Michael Zhang; Jong Yoon Lee; Jun Bin Park; Hokyung Jang; Zhaoqian Xie; Jingyue Cao; Raudel Avila; Chen Wei; Vincent D'Angelo; Jason Zhu; Ha Uk Chung; Sarah Coughlin; Manish Patel; Joshua Winograd; Jaeman Lim; Anthony Banks; Shuai Xu; Yonggang Huang; John A Rogers
Journal:  Sci Adv       Date:  2020-12-04       Impact factor: 14.136

2.  Value of GPSkin for the measurement of skin barrier impairment and for monitoring of rosacea treatment in daily practice.

Authors:  Jade G M Logger; Rieke J B Driessen; Elke M G J de Jong; Piet E J van Erp
Journal:  Skin Res Technol       Date:  2020-06-23       Impact factor: 2.365

3.  Novel technology at hand to measure skin hydration by Biodisplay smartphone touch screen panel.

Authors:  YoungHwan Choi; Se Jin Oh; Jong Hee Lee
Journal:  Sci Rep       Date:  2021-09-30       Impact factor: 4.379

4.  Fine Wrinkle Treatment and Hydration on the Facial Dermis Using HydroToxin Mixture of MicroBotox and MicroHyaluronic Acid.

Authors:  Jong Seo Kim
Journal:  Aesthet Surg J       Date:  2021-05-18       Impact factor: 4.283

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.