Establishing a Ventilator-Heart Lung Machine Communication Bridge to Mitigate Errors when Weaning from Bypass.

Geoffrey Rance1, David Arney2,3, Rithy Srey1, Julian M Goldman2,3, Marco A Zenati1,3.   

Abstract

If a perfusionist weans a patient off the heart lung machine (HLM) and the anesthesiologist has not re-started the ventilator, the patient will become hypoxic. The objective of this project was to create a redundant safety system of verbal and electronic communication to prevent failure to ventilate errors after cardiopulmonary bypass. This objective could be realized by building an electronic communication bridge directly between the HLM and ventilator. A software application was created to retrieve and interpret data from the pump and ventilator and trigger a programmed smart alarm. The software is able to interpret data from the pump and ventilator. When both are off simultaneously (defined as a pump flow of 0 L/min with a respiratory rate of 0 breaths/min), the application will raies an alarm. Communication between a pump and ventilator is possible, enabling the deployment of a safety system that could exist in the operating room (OR) as a standalone alarm. A device dataset can be used to optimize clinical performance of the alarm. The application could also be integrated into smart checklists and computer-assisted OR process models that are currently in development.

Entities:  

Keywords:  cardiopulmonary bypass; patient safety; perfusion; ventilation

Mesh:

Year:  2019        PMID: 30936587      PMCID: PMC6436169     

Source DB:  PubMed          Journal:  J Extra Corpor Technol        ISSN: 0022-1058


  1 in total

1.  Artificial intelligence in cardiothoracic surgery.

Authors:  Roger D Dias; Julie A Shah; Marco A Zenati
Journal:  Minerva Cardioangiol       Date:  2020-09-29       Impact factor: 1.347

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.