Literature DB >> 30930015

Cyclic Fatigue and Torsional Failure of EdgeTaper Platinum Endodontic Files at Simulated Body Temperature.

Ahmed Jamleh1, Amjad Alghaihab2, Abdulmohsen Alfadley2, Hussam Alfawaz3, Abdullah Alqedairi3, Khalid Alfouzan2.   

Abstract

INTRODUCTION: The aim of this in vitro study was to compare the cyclic fatigue resistance and torsional resistance of EdgeTaper Platinum (ETP) with those of ProTaper Gold (PTG) at simulated body temperature.
METHODS: ETP and PTG files with #25 tip size were selected. Cyclic fatigue tests were performed in a stainless steel artificial canal until file fracture occurred. The time to fracture was recorded. For the torsional resistance testing, the apical 5 mm of the file was firmly secured with acrylic resin, and the assembly was fixed over torque gauge device. A uniform torsional stress was applied with continuous rotation motion until fracture occurred. The torque at failure was recorded. Both experiments were conducted at 35 ± 1°C. Fractured surfaces were tested via scanning electron microscopy. Statistical analysis was performed using independent Student t test at 5% significance level.
RESULTS: The ETP had superior cyclic fatigue resistance than PTG (P < .01). However, it exhibited lower torsional resistance (P < .05). The ETP and PTG showed typical features of cyclic and torsional fatigue behavior under scanning electron microscopy.
CONCLUSIONS: The ETP exhibited superior resistance to cyclic fatigue but failed to show any improvement in the torsional resistance compared with PTG.
Copyright © 2019 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cyclic fatigue; EdgeTaper Platinum; ProTaper Gold; torsional resistance

Mesh:

Substances:

Year:  2019        PMID: 30930015     DOI: 10.1016/j.joen.2019.02.008

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  3 in total

1.  Effect of NaOCl and EDTA irrigating solutions on the cyclic fatigue resistance of EdgeTaper Platinum instruments.

Authors:  Hussam Alfawaz; Abdullah Alqedairi; Maha Alhamdan; Najla Alkhzim; Shatha Alfarraj; Ahmed Jamleh
Journal:  BMC Oral Health       Date:  2022-05-22       Impact factor: 3.747

2.  A Rare Case of Single-Rooted Mandibular Second Molar with Single Canal.

Authors:  Abdulmohsen Alfadley; Ahmad Alquraishi; Yaser Almazrou; Fahd Aljarbou
Journal:  Case Rep Dent       Date:  2020-06-15

3.  Mechanical Properties of ProTaper Gold, EdgeTaper Platinum, Flex Gold and Pro-T Rotary Systems.

Authors:  Murilo Alcalde; Marco Antonio Hungaro Duarte; Pablo Andrés Amoroso Silva; Pedro Henrique Souza Calefi; Emmanuel Silva; Jussaro Duque; Rodrigo Vivan
Journal:  Eur Endod J       Date:  2020-12
  3 in total