Andrea Moscatelli1, Francesca Febbo2, Silvia Buratti1, Stefano Pezzato1, Francesca Bagnasco3, Elisabetta Lampugnani1, Halkawt Nuri4, Francesca Buffelli5, Chiara Grasso6, Elio Castagnola7. 1. Neonatal and Pediatric Intensive Care Unit, Department of Surgery and Critical Care, IRCCS Istituto Giannina Gaslini, Genova, Italy. 2. Postgraduate School in Anesthesia and Intensive Care, University of Genova, Genova, Italy. 3. Epidemiology and Biostatistics Unit, IRCCS Istituto Giannina Gaslini, Genova, Italy. 4. Division of Cardiovascular Surgery, Department of Surgery and Critical Care, IRCCS Istituto Giannina Gaslini, Genova, Italy. 5. Fetal-Perinatal Pathology Unit, Mother and Child Department, IRCCS Istituto Giannina Gaslini, PhD Course in Pediatric Sciences, Fetal-Perinatal and Pediatric Pathology, University of Genova, Genova, Italy. 6. Postgraduate School in Anesthesia and Intensive Care, University of Catania, Catania, Italy. 7. Division of Infectious Diseases, Department of Pediatrics, IRCCS Istituto Giannina Gaslini, Genova, Italy.
Abstract
OBJECTIVES: To describe a single institution experience on echo-guided percutaneous bicaval double lumen extracorporeal membrane oxygenation cannulation performed at the bedside by intensivists. DESIGN: Retrospective observational study. SETTING: Extracorporeal membrane oxygenation team of a tertiary care children's hospital. PATIENTS: All patients 0-14 years old undergoing venovenous extracorporeal membrane oxygenation from January 1, 2013, to January 1, 2018. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Thirty children underwent 32 extracorporeal membrane oxygenation runs. Median age at enrollment was 2 months (interquartile range, 0-20.5 mo), 65.6% of the runs (21 patients) were performed in newborns (n = 13, 40.6%) or infants (n = 8, 25%). Median preextracorporeal membrane oxygenation index was 66.9 (interquartile range, 50-85.6). Major comorbidities were present in 50% of patients. All patients were cannulated percutaneously. In two cases cannulation occurred from the left internal jugular vein. Extracorporeal membrane oxygenation was effective in increasing pH, arterial oxygen saturation, PaO2, and lowering PaCO2. The overall differences in pre and postextracorporeal membrane oxygenation values were statistically significant, while stratifying patients according to the cannula diameter (mm)/major diameter of the cannulated internal jugular vein (mm) ratio (> 0.67 or ≤ 0.67), statistical significance was reached only for the highest ratio. Complications were observed in three runs: two cannula tip dislocations in the right atrium and one limited flow in the only case in which an Avalon cannula was not used. In 20 cases (62.5% of 32 runs), the cannulated vessel was patent at follow-up or autopsy. A ratio less than or equal to 0.67 or greater than 0.67 did not influence the occurrence rate of complications, nonpatency of the internal jugular vein, death for intracranial bleeding and death at 30 days from extracorporeal membrane oxygenation discontinuation. Overall cumulative survival at 30 days from extracorporeal membrane oxygenation discontinuation was 60% (95% CI, 40-75), with a survival advantage in the case of ratio greater than 0.67 (65%; 95% CI, 44-80 vs 25%; 95% CI, 0-60). CONCLUSIONS: The described technique proved to be feasible, safe, and effective. Further investigation is needed.
OBJECTIVES: To describe a single institution experience on echo-guided percutaneous bicaval double lumen extracorporeal membrane oxygenation cannulation performed at the bedside by intensivists. DESIGN: Retrospective observational study. SETTING: Extracorporeal membrane oxygenation team of a tertiary care children's hospital. PATIENTS: All patients 0-14 years old undergoing venovenous extracorporeal membrane oxygenation from January 1, 2013, to January 1, 2018. INTERVENTIONS: None. MEASUREMENTS AND MAIN RESULTS: Thirty children underwent 32 extracorporeal membrane oxygenation runs. Median age at enrollment was 2 months (interquartile range, 0-20.5 mo), 65.6% of the runs (21 patients) were performed in newborns (n = 13, 40.6%) or infants (n = 8, 25%). Median preextracorporeal membrane oxygenation index was 66.9 (interquartile range, 50-85.6). Major comorbidities were present in 50% of patients. All patients were cannulated percutaneously. In two cases cannulation occurred from the left internal jugular vein. Extracorporeal membrane oxygenation was effective in increasing pH, arterial oxygen saturation, PaO2, and lowering PaCO2. The overall differences in pre and postextracorporeal membrane oxygenation values were statistically significant, while stratifying patients according to the cannula diameter (mm)/major diameter of the cannulated internal jugular vein (mm) ratio (> 0.67 or ≤ 0.67), statistical significance was reached only for the highest ratio. Complications were observed in three runs: two cannula tip dislocations in the right atrium and one limited flow in the only case in which an Avalon cannula was not used. In 20 cases (62.5% of 32 runs), the cannulated vessel was patent at follow-up or autopsy. A ratio less than or equal to 0.67 or greater than 0.67 did not influence the occurrence rate of complications, nonpatency of the internal jugular vein, death for intracranial bleeding and death at 30 days from extracorporeal membrane oxygenation discontinuation. Overall cumulative survival at 30 days from extracorporeal membrane oxygenation discontinuation was 60% (95% CI, 40-75), with a survival advantage in the case of ratio greater than 0.67 (65%; 95% CI, 44-80 vs 25%; 95% CI, 0-60). CONCLUSIONS: The described technique proved to be feasible, safe, and effective. Further investigation is needed.