Intra- and Interdevice Deviation of Optical Coherence Tomography Angiography.

Stefan Trachsler1, Arthur Emanuel Baston2, Marcel Menke1.   

Abstract

PURPOSE: To compare 4 optical coherence tomography-angiography (OCT-A) devices for foveal avascular zone (FAZ) measurements in healthy subjects.
METHODS: The central retinas of 24 eyes of 12 healthy subjects were scanned with 4 different OCT-A devices (Optovue RTVue-XR, Zeiss Cirrus 5000-HD-OCT, a prototype Spectralis OCT2, Heidelberg Engineering, and Topcon DRI-OCT Triton Swept-source OCT). For the Topcon, Zeiss, and Optovue devices, 3-mm and 6-mm scans were performed. The Heidelberg device only provided 4-mm scans. En-face OCT-A images of the superficial and deep capillary plexus of the macular area were generated. The FAZ areas were measured and compared.
RESULTS: Twenty-four healthy eyes were included. OCT-A devices showed significant differences in FAZ measurements. The Zeiss OCT-A device measured the smallest values for foveal avascular area (mean 218.7 mm2), followed by the Optovue device (229.6 mm2), the Topcon device (239.3 mm2), and the Heidelberg device (250.4 mm2). Differences were statistically significant for following devices: Heidelberg versus Optovue (p < 0.001), Heidelberg versus Zeiss (p < 0.001), Topcon versus Zeiss (p < 0.001), and Optovue versus Zeiss (p = 0.046). For the Optovue device, FAZ measurements were significantly different between 3 mm (mean 220 mm2) and 6 mm (mean 239.3 mm2, p = 0.007) scans. All other devices showed no significant difference within scan modes.
CONCLUSION: Current OCT-A devices provide images that allow such measurements, but values showed significant differences between devices and, for the Optovue instrument, even within scan modes. The data for OCTA measurements cannot be transferred interchangeably between the devices. Therefore, a patient should always be measured with the same device. Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2019        PMID: 30919402     DOI: 10.1055/a-0747-5333

Source DB:  PubMed          Journal:  Klin Monbl Augenheilkd        ISSN: 0023-2165            Impact factor:   0.700


  4 in total

1.  Comparison between two multimodal imaging platforms: Nidek Mirante and Heidelberg Spectralis.

Authors:  Kimberly Spooner; Long Phan; Mariano Cozzi; Thomas Hong; Giovanni Staurenghi; Eugenia Chu; Andrew A Chang
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-01-06       Impact factor: 3.117

2.  Dynamics of Retinal Vessel Loss After Acute Optic Neuritis in Patients With Relapsing Multiple Sclerosis.

Authors:  Lilian Aly; Christina Noll; Rebecca Wicklein; Elisabeth Wolf; Eva Feodora Romahn; Josphine Wauschkuhn; Sami Hosari; Christian Mardin; Achim Berthele; Bernhard Hemmer; Thomas Korn; Benjamin Knier
Journal:  Neurol Neuroimmunol Neuroinflamm       Date:  2022-03-17

3.  Correlation analysis of physical fitness and retinal microvasculature by OCT angiography in healthy adults.

Authors:  Pieter Nelis; Boris Schmitz; Andreas Klose; Florian Rolfes; Maged Alnawaiseh; Michael Krüger; Nicole Eter; Stefan-Martin Brand; Florian Alten
Journal:  PLoS One       Date:  2019-12-03       Impact factor: 3.240

4.  Assessing the Influence of OCT-A Device and Scan Size on Retinal Vascular Metrics.

Authors:  Jessica A Kraker; Bisola S Omoba; Jenna A Cava; Taly Gilat Schmidt; Toco Y Chui; Richard B Rosen; Judy E Kim; Joseph Carroll; Rachel E Linderman
Journal:  Transl Vis Sci Technol       Date:  2020-10-07       Impact factor: 3.283

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.