The 55Fe half-life measured with a pressurised proportional counter.

S Pommé1, H Stroh2, R Van Ammel2.   

Abstract

The half-life of 55Fe has been measured accurately by following the decay curve of three sources with a large pressurised proportional counter. An argon(90%)-methane(10%) mixture was used as counter gas, at atmospheric pressure (∼1 × 105 Pa) and at enhanced pressures of 5 × 105 Pa and 8 × 105 Pa (for 1 source), respectively. The first measurements were performed in 2001, but the experiment was executed more systematically between 2005 and 2018, covering a period of about 5 half-lives. The residuals from an exponential decay curve were of the order of 0.1% to 0.2% at 1 × 105 Pa, and 0.03% at 5 × 105 Pa and 8 × 105 Pa. The gain of stability with increased gas pressure was due to asymptotically reaching the maximum counting efficiency, resulting in lower sensitivity to pressure variations. The deduced half-life value of T1/2(55Fe) = 1006.70 (15) d or 2.7563 (4) a is more accurate than other data in literature, which are mutually discrepant. It is consistent with previous measurements at JRC with an X-ray defined solid angle counter.
Copyright © 2019 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Calibration; Clearance; Decay constant; Decommissioning; Discrepancy; Half-life

Year:  2019        PMID: 30917331     DOI: 10.1016/j.apradiso.2019.01.008

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  Air humidity and annual oscillations in 90Sr/90Y and 60Co decay rate measurements.

Authors:  S Pommé; K Pelczar; I Kajan
Journal:  Sci Rep       Date:  2022-06-09       Impact factor: 4.996

2.  Fluctuations in measured radioactive decay rates inside a modified Faraday cage: Correlations with space weather.

Authors:  V Milián-Sánchez; F Scholkmann; P Fernández de Córdoba; A Mocholí-Salcedo; F Mocholí; M E Iglesias-Martínez; J C Castro-Palacio; V A Kolombet; V A Panchelyuga; G Verdú
Journal:  Sci Rep       Date:  2020-05-22       Impact factor: 4.379

3.  Role of ambient humidity underestimated in research on correlation between radioactive decay rates and space weather.

Authors:  S Pommé; K Pelczar
Journal:  Sci Rep       Date:  2022-02-15       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.