Atomic force microscopy and energy dispersive X-ray spectrophotometry analysis of reciprocating and continuous rotary nickel-titanium instruments following root canal retreatment.

Bulem Üreyen Kaya1, Cevat Emre Erik1, Gülsen Kiraz2.   

Abstract

The aim was to examine the effect of retreatment process on the surface roughness and nickel titanium (NiTi) composition of ProTaper Universal Retreatment (PTUR; consists of 3 files; D1, D2, D3) and WaveOne Gold (WOG) (primary) instruments. Twenty extracted mandibular molar teeth with severe curved (30-40°) mesial roots were selected and divided into two groups (n = 10) based on the instrument used for the removal of the root canal filling. Before and after using the instruments in two canals, they were subjected to atomic force microscopy (AFM) and energy dispersive X-ray spectrophotometry (EDX) analysis. The EDX analysis data and roughness average (Ra) and root mean square (RMS) values were analyzed statistically using a one-way analysis of variance and post hoc Tukey's test at the 5% significant level. There was no significant difference between the new and used D1 and D2 PTUR and WOG instruments in terms of the Ni composition (p > .05). The Ti contents of the used D2 and D3 PTUR instruments were lower those of the new instruments (p < .05). In both new and used instruments, PTUR and WOG have no difference in terms of Ra and RMS values. (p > .05). The Ra and RMS values of the PTUR and WOG systems significantly increased after removal of the root canal filling (p < .05). The use of PTUR and WOG instruments for removal of root canal filling in severely curved root canals affected the surface topography of the files. The NiTi composition of the WOG instruments was unaffected by the retreatment process.
© 2019 Wiley Periodicals, Inc.

Entities:  

Keywords:  atomic force microscopy; energy dispersive X-ray spectrophotometry; nickel-titanium instruments; root canal retreatment; surface roughness

Mesh:

Substances:

Year:  2019        PMID: 30916441     DOI: 10.1002/jemt.23264

Source DB:  PubMed          Journal:  Microsc Res Tech        ISSN: 1059-910X            Impact factor:   2.769


  3 in total

1.  Microstructural, microchemical, and mechanical changes associated with the clinical reuse of two nickel-titanium endodontic instruments.

Authors:  Felipe Augusto Restrepo-Restrepo; Viviana Andrea Holguín-Vásquez; Syldana Julieth Cañas-Jiménez; Paula Andrea Villa-Machado; Sara Ochoa-Soto; Claudia Patricia Ossa-Orozco; Sergio Iván Tobón-Arroyave
Journal:  Dent Res J (Isfahan)       Date:  2021-06-22

2.  Effectiveness and safety of rotary and reciprocating kinematics for retreatment of curved root canals: a systematic review of in vitro studies.

Authors:  Lucas Pinho Simões; Alexandre Henrique Dos Reis-Prado; Carlos Roberto Emerenciano Bueno; Ana Cecília Diniz Viana; Marco Antônio Húngaro Duarte; Luciano Tavares Angelo Cintra; Cleidiel Aparecido Araújo Lemos; Francine Benetti
Journal:  Restor Dent Endod       Date:  2022-04-06

Review 3.  Metallurgical Tests in Endodontics: A Narrative Review.

Authors:  Alessio Zanza; Marco Seracchiani; Rodolfo Reda; Gabriele Miccoli; Luca Testarelli; Dario Di Nardo
Journal:  Bioengineering (Basel)       Date:  2022-01-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.