The influence of cavity preparation and press-fit cup implantation on restoring the hip rotation center.

Philipp Messer-Hannemann1, Johanna Bätz2, Frank Lampe3, Anke Klein4, Klaus Püschel4, Graeme M Campbell2, Michael Morlock2.   

Abstract

BACKGROUND: Reaming of the acetabular cavity and cup implantation directly influence the hip rotation center and contact area between implant and bone. Previous studies have reported on an altered rotation center after total hip arthroplasty, but have not studied the influence of reaming and cup implantation separately. Aim of this study was therefore to analyze the individual influence of acetabular reaming and subsequent cup implantation on the rotation center and how this influences the contact conditions at the bone-implant interface.
METHODS: Acetabular press-fit cups were implanted into the left and right hips of three full cadavers (n = 6). CT scans were performed to calculate the change in hip rotation center after reaming and prior to liner insertion. 3D models of the cups were used to determine the polar gap, the contact conditions and the effective press-fit.
FINDINGS: Reaming the acetabular cavity shifted the rotation center medially (median 5.8 mm, range 4.8-9.1), superiorly (5.3 mm, 3.0-7.0) and posteriorly (2.9 mm, 1.0-5.3). With cup implantation, the rotation center shifted back towards the native position, but no full restoration was observed. The degree of shift increased with the size of polar gap (rs = 0.829, P = .042), which inversely reduced the contact area (rs = 0.886, P = .019).
INTERPRETATION: This study reveals that the dominant factor in hip rotation center restoration is the reaming process, while the cup implantation for a given nominal press-fit has only a small influence. Increasing the press-fit would improve the restoration but bares the danger of insufficient bone coverage and periprosthetic fractures due to the high forces needed.
Copyright © 2019 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Cavity preparation; Contact condition; Cup implantation; Hip arthroplasty; Press-fit; Rotation center

Mesh:

Year:  2019        PMID: 30913461     DOI: 10.1016/j.clinbiomech.2019.03.006

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  2 in total

Review 1.  Cup placement in primary total hip arthroplasty: how to get it right without navigation or robotics.

Authors:  Geert Meermans; George Grammatopoulos; Moritz Innmann; David Beverland
Journal:  EFORT Open Rev       Date:  2022-05-31

2.  Mid-term and long-term results of restoring rotation center in revision hip arthroplasty.

Authors:  Heng Zhang; Jiansheng Zhou; Yang Liu; Jianzhong Guan; Hai Ding; Zhiyan Wang; Qirong Dong
Journal:  J Orthop Surg Res       Date:  2020-04-16       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.