Literature DB >> 30908684

Effect of cavity preparation and bone mineral density on bone-interface densification and bone-implant contact during press-fit implantation of hip stems.

Johanna Bätz1, Philipp Messer-Hannemann1, Frank Lampe2, Anke Klein3, Klaus Püschel3, Michael M Morlock1, Graeme M Campbell1.   

Abstract

Implant loosening and periprosthetic fracture are two major revision causes for uncemented hip stems. The chosen method of cavity preparation could play a key role for both failure mechanisms. The aim of this study was to determine the dependence of the broach type as well as patient bone mineral density (BMD) on densification and contact conditions at the bone-implant interface. Hip stems were implanted into cadaveric femora using compaction, blunt extraction or sharp extraction broaches with computed tomography scans performed prior to broaching, after broaching and after stem implantation. Proximal periprosthetic bone densification as well as press-fit, contact area and stem seating relative to the last broach were determined. Median bone densification was higher with the compaction and blunt extraction broaches compared to sharp extraction broaches (181% and 177%, respectively, p = 0.002). The bone densification of femora prepared with compaction broaching increased with higher BMD (R2  = 0.183, p = 0.037), while stem seating decreased with higher BMD for all broach types (R2  = 0.259, p = 0.001). Incomplete seated prostheses were associated with smaller press-fit and bone-implant contact area (R2  = 0.249, p = 0.001; R2  = 0.287, p < 0.001). Clinical Significance: The results suggest that compaction broaching maximizes bone densification in patients with higher bone density. However, there appears to be an increased risk of insufficient stem seating in high-density bone that could limit the benefits for primary stability. For lower quality bone, the broach type appears to play a lesser role, but care must be taken to limit extensive stem seating which might increase periprosthetic fracture risk.
© 2019 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 37:1580-1589, 2019. © 2019 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.

Entities:  

Keywords:  QCT analysis; bone-implant contact; femoral bone preparation; osseodensification; press-fit

Year:  2019        PMID: 30908684     DOI: 10.1002/jor.24288

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  4 in total

1.  Uncemented total hip arthroplasty; increased risk of early periprosthetic fracture requiring revision surgery in elderly females.

Authors:  S R Hopman; T S de Windt; J H J van Erp; J E J Bekkers; A de Gast
Journal:  J Orthop       Date:  2021-04-02

2.  Co-Cr-Mo-Cu alloys for clinical implants with osteogenic effect by increasing bone induction, formation and development in a rabbit model.

Authors:  Jingzhu Duan; Yang Yang; Erlin Zhang; Huan Wang
Journal:  Burns Trauma       Date:  2020-12-21

3.  Topology Optimisation for Compliant Hip Implant Design and Reduced Strain Shielding.

Authors:  Nathanael Tan; Richard J van Arkel
Journal:  Materials (Basel)       Date:  2021-11-25       Impact factor: 3.623

4.  Variability in Femoral Preparation and Implantation Between Surgeons Using Manual and Powered Impaction in Total Hip Arthroplasty.

Authors:  Tobias Konow; Johanna Bätz; David Beverland; Tim Board; Frank Lampe; Klaus Püschel; Michael M Morlock
Journal:  Arthroplast Today       Date:  2022-01-20
  4 in total