BACKGROUND: Sutureless and rapid-deployment bioprostheses (RPDs) are advantageous due to shorter aortic cross-clamp times (ACCs); however, few studies have investigated hemodynamic outcomes in comparison to conventional bioprostheses (CBs).
METHODS: All patients receiving the Perceval (LivaNova, London, United Kingdom) and Intuity (Edwards Lifesciences, Irvine, California, United States) valves from February 2016 to December 2017 were included (n = 61). For controls, patients who underwent aortic valve replacement using CB from 2015 to 2018 (n = 743) were propensity-matched (n = 108). Primary end points were mean gradient at discharge, true internal diameter (ID) after implantation, and paravalvular leakage. Secondary end points were ACCs, cardiopulmonary bypass times (CPBTs), mortality, and pacemaker implantation.
RESULTS: Age was 73.2 years (standard deviation [SD]: 7.6) in the RPD group and 72.9 years (SD: 7.3) in the CB group (p = 0.827). Median logistic EuroSCORE II was 3.4% (Q1: 2; IQ3: 5.3) and 3% (Q1: 1.9; IQ3: 5.6; p = 0.599). While ACCs and CPBT were shorter in RPDs (97 [SD: 31.4] vs. 125.4 minutes [SD: 62.1], p = 0.003; and 76.1 [SD: 25.7] vs. 89.7 minutes [SD: 34.3], p = 0.022), procedural times were similar (p = 0.257). True ID was 21.97 mm (SD: 1.79) in RPDs and 20.15 mm (SD: 1.70) in CB (p < 0.001).RPDs and CB resulted in comparable mean gradients (12.8 mm Hg [SD: 6.4] vs. 13.8 mm Hg [SD: 5.6]; p = 0.387) and rate of paravalvular regurgitation. There were no differences in the rates of pacemaker implantation and mortality.In a subanalysis of RPDs, mean gradient was 15.48 mm Hg (SD: 7.51) in Perceval (n = 21) and 10.79 mm Hg (SD: 4.78) in Intuity (n = 33; p = 0.010).
CONCLUSIONS: RPDs provided comparable hemodynamic performance, although implanted valves were larger in true ID. Whether this provides superior hemodynamic performance during exercise and longer durability needs further investigation. Among RPDs, the Intuity valve provided lower mean gradients at rest than Perceval valve. Thieme. All rights reserved.

Entities:  

Mesh:

Year:  2019        PMID: 30900219     DOI: 10.1055/s-0039-1683426

Source DB:  PubMed          Journal:  Thorac Cardiovasc Surg        ISSN: 0171-6425            Impact factor:   1.827


  3 in total

1.  Sutureless valve and rapid deployment valves: a systematic review and meta-analysis of comparative studies.

Authors:  Campbell D Flynn; Michael L Williams; Adam Chakos; Lucy Hirst; Benjamin Muston; David H Tian
Journal:  Ann Cardiothorac Surg       Date:  2020-09

2.  Direct comparison of rapid deployment versus sutureless aortic valve replacement: a meta-analysis.

Authors:  Suk Ho Sohn; Yoonjin Kang; Ji Seong Kim; Jae Woong Choi; Myoung-Jin Jang; Ho Young Hwang
Journal:  J Thorac Dis       Date:  2021-04       Impact factor: 2.895

Review 3.  Rapid deployment technology versus conventional sutured bioprostheses in aortic valve replacement.

Authors:  Mohammad Yousuf Salmasi; Sruthi Ramaraju; Iqraa Haq; Ryan A B Mohamed; Taimoor Khan; Faruk Oezalp; George Asimakopoulos; Shahzad G Raja
Journal:  J Card Surg       Date:  2022-01-14       Impact factor: 1.778

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.