Literature DB >> 30885874

Side Slit Guide Pipe for Precise Placement of Depth Electrodes.

Ayataka Fujimoto1, Yosuke Masuda2, Naoki Ichikawa2, Keishiro Sato2, Shimpei Baba2, Shinji Itamura2, Mitsuyo Nishimura2, Hideo Enoki2, Tohru Okanishi2.   

Abstract

BACKGROUND: Using a stereotactic technique, surgeons can accurately place a depth electrode (DE), but sometimes the DE deviates from the intended target due to movement of the electrode or leakage of cerebrospinal fluid when placing the electrode. If DEs can be anchored before removing the catheter insertion guide pipe, more accurate placement may be possible.
METHODS: We made a side slit guide pipe. When the DEs were anchored to the dura or the edge of the burr hole, the DE did not move when the guide pipe was removed. We measured the distance between the planned target and the tip of the electrode in 13 patients (3 female and 10 male patients; age range, 7-43 years; mean age 23.0 years; median age 27 years) with medically intractable epilepsy who underwent DE placement with stereotactic neuronavigation guidance.
RESULTS: There were 30 DEs implanted. The mean distance from the planned target to the tip of the DE was 0.570 mm (range, 0.3-1.2 mm; median 0.5 mm; SD 0.212). The mean distance from the planned target to the tip of the DE with dural anchoring was 0.467 mm (range, 0.3-0.6 mm; median 0.45 mm; SD 0.121) and with burr hole edge anchoring was 0.596 mm (range, 0.3-1.2 mm; median 0.50 mm; SD 0.224; P = 0.205).
CONCLUSIONS: DEs can be anchored using the side slit guide pipe for more precise placement.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Accurate placement; Depth electrode; Neuronavigation; Side slit guide pipe; Stereoelectroencephalography

Mesh:

Year:  2019        PMID: 30885874     DOI: 10.1016/j.wneu.2019.03.067

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  2 in total

1.  Primary Experiences with Robot-assisted Navigation-based Frameless Stereo-electroencephalography: Higher Accuracy than Neuronavigation-guided Manual Adjustment.

Authors:  Yuichiro Kojima; Takehiro Uda; Toshiyuki Kawashima; Saya Koh; Masato Hattori; Yuki Mito; Noritsugu Kunihiro; Shohei Ikeda; Ryoko Umaba; Takeo Goto
Journal:  Neurol Med Chir (Tokyo)       Date:  2022-05-25       Impact factor: 2.036

2.  Postoperative Pneumocephalus on Computed Tomography Might Predict Post-Corpus Callosotomy Chemical Meningitis.

Authors:  Ayataka Fujimoto; Keisuke Hatano; Toshiki Nozaki; Keishiro Sato; Hideo Enoki; Tohru Okanishi
Journal:  Brain Sci       Date:  2021-05-15
  2 in total