Literature DB >> 30879719

Ciliary sulcus suture fixation of intraocular lens using an auxiliary device.

Takeshi Sugiura1, Yuichi Kaji2, Yoshikazu Tanaka3.   

Abstract

We developed an injector with a tip that matches the shape of the ciliary sulcus (Ciliary Sulcus Pad Injector) and is used in ciliary sulcus suture fixation of intraocular lenses. The injector is inserted through the corneal incisions and slid along the posterior surface of the iris to the opposite side. When the tip of injector is fitted in the ciliary sulcus, the plunger is pressed, causing the needle to pierce the ciliary sulcus. A suture is inserted in the ciliary sulcus without direct observation. In 91.4% of cases, the needle correctly pierced the ciliary sulcus in a single attempt. The difference between the preoperative expected refractive power and the postoperative spherical equivalent power was -0.63 diopter 6 months after surgery.
Copyright © 2019 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 30879719     DOI: 10.1016/j.jcrs.2019.01.021

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  2 in total

1.  Long-term outcomes of transsclerally sutured intraocular lens correctly fixed in the ciliary sulcus.

Authors:  Takeshi Sugiura; Tohru Sakimoto; Yoshikazu Tanaka; Yasushi Inoue; Tetsuro Oshika
Journal:  BMJ Open Ophthalmol       Date:  2022-04

2.  Flanged intraocular lens fixation via 27-gauge trocars using a double-needle technique decreases surgical wounds without losing its therapeutic effect.

Authors:  Hiroto Ishikawa; Hisashi Fukuyama; Yuki Komuku; Takashi Araki; Fumi Gomi
Journal:  Acta Ophthalmol       Date:  2019-11-17       Impact factor: 3.761

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.