Literature DB >> 30876781

Effects of decentration and tilt on the optical performance of 6 aspheric intraocular lens designs in a model eye.

Tjundewo Lawu1, Koichiro Mukai2, Hiroyuki Matsushima2, Tadashi Senoo2.   

Abstract

PURPOSE: To compare the effect of decentration and tilt on the optical performance of 6 aspheric intraocular lens (IOL) designs in a model eye.
SETTING: Department of Ophthalmology, Graduate School of Medicine, Dokkyo Medical University, Tochigi, Japan.
DESIGN: Experimental study.
METHODS: In theoretical simulations, the amount of spherical aberration in the IOL was varied to produce residual ocular spherical aberration (range -0.15 to 0.30 μm) at a 6.0 mm entrance pupil. Wavefront aberration analyses were performed with the ZEMAX optical design program (version August 20, 2014) to obtain the ocular root-mean-square values of astigmatism, coma, trefoil, and higher-order aberrations (HOAs) when the IOL was centered on the insertion position and misaligned at a 4.0 mm entrance pupil. The retinal visual images were calculated using the same conditions. Six 20.0 diopter (D) aspheric IOLs and one 20.0 D spherical IOL were used for the experimental studies. Each IOL was inserted in the model eye. The actual alignments were measured using a Scheimpflug camera (EAS-1000). The wavefront aberrations and visual images were gauged using a wavefront analyzer (KR-1W) at several IOL alignments.
RESULTS: Intraocular lens decentration and tilt increased wavefront aberrations and degraded optical performance. Astigmatism, coma, and HOAs generated by misaligned IOLs were related to the amount of spherical aberration correction of the IOLs. The extent of spherical aberration remained unchanged by the amount of misalignment. Experimental model eye results showed trends similar to theoretical results.
CONCLUSIONS: The spherical aberration correction amount in the aspheric IOL design was critical for the astigmatism, coma, and HOAs generated by the IOL misalignment. Additional spherical aberration corrections led to a more sensitive optical performance degradation resulting from IOL misalignment.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Year:  2019        PMID: 30876781     DOI: 10.1016/j.jcrs.2018.10.049

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  16 in total

1.  Comparison of visual outcomes of a diffractive trifocal intraocular lens and a refractive bifocal intraocular lens in eyes with axial myopia: a prospective cohort study.

Authors:  Tong Sun; Yiyun Liu; Yufei Gao; Chuhao Tang; Qianqian Lan; Tingting Yang; Xiaorui Zhao; Hong Qi
Journal:  BMC Ophthalmol       Date:  2022-10-20       Impact factor: 2.086

2.  Effect of decentration, tilt and rotation on the optical quality of various toric intraocular lens designs: a numerical and experimental study.

Authors:  Jesús Pérez-Gracia; Jorge Ares; Francisco J Ávila; Laura Remón
Journal:  Biomed Opt Express       Date:  2022-03-08       Impact factor: 3.562

3.  Stability of Various Types of Aspheric Intraocular Lenses After Implantation: A One-Year Retrospective Study.

Authors:  Yuan Ning; Yushuang Shao; Jiangyue Zhao; Jinsong Zhang; Mingwu Wang; Yu Qin
Journal:  Int J Gen Med       Date:  2021-06-01

4.  Influence of IOL Weight on Long-Term IOL Stability in Highly Myopic Eyes.

Authors:  Yuxi Chen; Jiaqi Meng; Kaiwen Cheng; Qiang Lu; Ling Wei; Yi Lu; Xiangjia Zhu
Journal:  Front Med (Lausanne)       Date:  2022-04-11

5.  Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors.

Authors:  Jesús Pérez-Gracia; Alejandra Varea; Jorge Ares; Juan A Vallés; Laura Remón
Journal:  PLoS One       Date:  2020-05-04       Impact factor: 3.240

6.  Characteristics and Risk Factors of Intraocular Lens Tilt and Decentration of Phacoemulsification After Pars Plana Vitrectomy.

Authors:  Xuhua Tan; Zhenzhen Liu; Xiaoyun Chen; Yi Zhu; Jingmin Xu; Xiaozhang Qiu; Guangyao Yang; Lulu Peng; Xiaoxun Gu; Jiaqing Zhang; Lixia Luo; Yizhi Liu
Journal:  Transl Vis Sci Technol       Date:  2021-03-01       Impact factor: 3.283

7.  Effects of capsular tension ring on surgical outcomes of premium intraocular lens in patients with suspected zonular weakness.

Authors:  Teruyuki Miyoshi; Shuhei Fujie; Hironori Yoshida; Hiroshi Iwamoto; Hideki Tsukamoto; Tetsuro Oshika
Journal:  PLoS One       Date:  2020-02-24       Impact factor: 3.240

8.  Decentration and tilt of plate-haptic multifocal intraocular lenses in myopic eyes.

Authors:  Jiaqi Meng; Wenwen He; Xianfang Rong; Ao Miao; Yi Lu; Xiangjia Zhu
Journal:  Eye Vis (Lond)       Date:  2020-04-07

Review 9.  Effect of Intraocular Lens Tilt and Decentration on Visual Acuity, Dysphotopsia and Wavefront Aberrations.

Authors:  Zahra Ashena; Sundas Maqsood; Syed Naqib Ahmed; Mayank A Nanavaty
Journal:  Vision (Basel)       Date:  2020-09-14

10.  Determinants of intraocular lens tilt and decentration after cataract surgery.

Authors:  Xiaoxun Gu; Xiaoyun Chen; Guangyao Yang; Wei Wang; Wei Xiao; Guangming Jin; Lanhua Wang; Ye Dai; Xiaoting Ruan; Zhenzhen Liu; Lixia Luo; Yizhi Liu
Journal:  Ann Transl Med       Date:  2020-08
View more