| Literature DB >> 30876607 |
Yusheng Wang1, Xiaoqi Wang1, Wei Lu1, Qingpan Yuan1, Yourong Zheng1, Bo Yao2.
Abstract
In this work, a thin film PET based gold electrode (PGE) glucose sensor was fabricated through ultraviolet mediated chemical plating technique. Compared with the most existing wearable thin film gold electrode sensors, the PGE-glucose sensor is simple, low cost and minimum instrumentation was required. The proposed glucose sensor revealed a sensitivity of 22.05 μA mM-1 cm-2 in a linear range from 0.02 to 1.11 mM with a low detection limit of 2.7 μM (S/N = 3). The PGE-glucose sensor had a good selectivity and was not interfered by lactic acid, urea, acetaminophen, uric acid, dopamine and ascorbic acid. Additionally, it exhibited good reproducibility and long-term stability over four weeks. Finally, the disposable PGE-glucose sensor was successfully applied to determine the glucose in sweat and commercial beverage accurately.Entities:
Keywords: Chemical plating; Glucose sensor; PET; Plastic gold electrode; Sweat
Mesh:
Substances:
Year: 2019 PMID: 30876607 DOI: 10.1016/j.talanta.2019.01.104
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057