Literature DB >> 30866744

An experimental study on the digital precision of internal fixation via the sinus tarsi approach for calcaneal fractures.

Jing Xu1,2, Zaopeng He3,2, Guodong Zhang4,2, Yongshao Chen5, Chunrong Wei3, Yudong Zheng3, Haibin Lin4, Wei Li3, Wenhua Huang1.   

Abstract

OBJECTIVE: To investigate the effectiveness and accuracy of internal fixation of calcaneal fractures with a three-dimensional (3-D) printing navigation module via the sinus approach.
METHODS: Eleven lateral lower extremity specimens were used in the experiment and divided into the digital design (DD) group ( n = 11) and the real surgery (RS) group ( n = 11). For the DD group, thin-section computed tomography (CT) scans, virtual fracture modeling, virtual bone plate fixation, sinus occlusal module design, and navigation module design were performed for the cadaver specimens. A 3-D navigation module was printed using a 3-D printer. For the RS group, the sinus approach incision was made, and the navigation module was used to assist the placement of screw fixation and anatomically locked bone plate fixation. Thin-layer CT scans, 3-D reconstructions, and reconstruction of the nail paths were performed in the RS group and 3-D registered with the DD group. 3-D coordinate values for the screw entry and exit points in the RS and DD groups were recorded, and the corresponding offset values of the screw points were calculated. The number of qualified nails at different accuracy levels was counted. The χ2 test was used to compare the data for the DD and RS groups to obtain the accuracy level for both groups when p > 0.05, which is the critical value for experimental accuracy.
RESULTS: A total of 11 bone plates were placed and 120 screws were locked. None of them protruded from the inferior articular surface. The screw entry and exit point offsets were 1.71 ± 0.11 mm and 3.10 ± 0.19 mm, respectively. When the accuracy requirement for the entry point was 3.8 mm, there was no statistically significant difference between the two groups ( p > 0.05). Similarly, the accuracy of the exit point of the screw could only reach 6.6 mm.
CONCLUSION: Internal fixation of calcaneal fractures via the sinus tarsi approach with an anatomical locking plate based on the 3-D printing navigation module can achieve accurate screw placement and good overall internal fixation.

Entities:  

Keywords:  3-D printing; Calcaneal fracture; internal fixation; locking bone plate; navigation

Mesh:

Year:  2019        PMID: 30866744     DOI: 10.1177/2309499019834072

Source DB:  PubMed          Journal:  J Orthop Surg (Hong Kong)        ISSN: 1022-5536            Impact factor:   1.118


  2 in total

1.  Comparison of the modified sinus tarsi approach versus the extensile lateral approach for displaced intra-articular calcaneal fractures.

Authors:  Chao Ma; Jiaju Zhao; Yong Zhang; Nan Yi; Jupu Zhou; Zhicheng Zuo; Bo Jiang
Journal:  Ann Transl Med       Date:  2021-04

2.  Biomechanical testing of three coracoclavicular ligament reconstruction techniques with a 3D printing navigation template for clavicle-coracoid drilling.

Authors:  Ji Qi; Shijie Fu; Ruiyue Ping; Kai Wu; Ziyu Feng; Yanxiao Xu; Xiaoguang Guo; Dingkun Lin; Lei Zhang
Journal:  Ann Transl Med       Date:  2021-07
  2 in total