| Literature DB >> 30855238 |
Sanghoon Song1, Roberto Alonso-Mori1, Matthieu Chollet1, Yiping Feng1, James M Glownia1, Henrik T Lemke1, Marcin Sikorski1, Diling Zhu1, Stefan Moeller1, Hae Ja Lee1, Mark S Hunter1, Gabriella Carini2, Kai Tiedtke3, Ulf Jastrow3, Andrey Sorokin3, Mathias Richter4, Shigeki Owada5, Kensuke Tono5, Norio Saito5, Takahiro Tanaka5, Masahiro Kato5, Makina Yabashi5, Aymeric Robert1.
Abstract
X-ray free-electron lasers provide intense pulses of coherent X-rays with a short pulse duration. These sources are chaotic by nature and therefore, to be used at their full potential, require that every X-ray pulse is characterized in terms of various relevant properties such as intensity, photon energy, position and timing. Diagnostics are for example installed on an X-ray beamline to specifically monitor the intensity of individual X-ray pulses. To date, these can however only provide a single-shot value of the relative number of photons per shot. Here are reported measurements made in January 2015 of the absolute number of photons in the hard X-ray regime at LCLS which is typically 3.5 × 1011 photons shot-1 between 6 and 9.5 keV at the X-ray Pump-Probe instrument. Moreover, an average transmission of ≉62% of the hard X-ray beamline over this energy range is measured and the third-harmonic content of ≉0.47% below 9 keV is characterized. open access.Entities:
Keywords: FEL; absolute intensity; hard X-rays
Year: 2019 PMID: 30855238 PMCID: PMC6412180 DOI: 10.1107/S1600577519000250
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616
Figure 1Schematic of the experimental set-up showing the various areas of interest, i.e. undulator hall, front-end enclosure (FEE) and XPP instrument hutch, and highlighting the location of relevant optical and detector components, such as the hard X-ray offset mirrors system (HOMS) and X-ray gas monitor detector (XGMD).
Figure 2Average absolute number of photons per shot as a function of X-ray photon energy E X-ray which is normalized by the FEE GD. The X-ray photon energy dependence of the number of photons, , is provided by circles at the XPP sample location under typical experimental conditions. The same quantity circles 〈N p〉HOMS is extrapolated to the location downstream of the HOMS and is indicated by squares. This is achieved by correcting by the transmission of all components located between directly downstream of the HOMS and the XPP sample location. Details are provided in Appendix A . The dashed line displays the photon energy dependence of the number of photons for a 1 mJ X-ray beam. Error bars of circles are contained within the symbol size. The gray area surrounding the squares indicates a total 10% error margin in transmission correction as a result of the uncertainty of the instrument optics (thickness of one diamond and one kapton window) and calibration error of the FEE GD (Moeller et al., 2011 ▸).
X-ray photon energy dependence of: the average number of photons per shot at the XPP sample location and directly downstream of the HOMS which are normalized by the FEE GD, the HOMS total transmission and the third- to first-harmonic content ratio
| 〈 | ||||
|---|---|---|---|---|
| Photon energy | (1011 photons shot−1) | Transmission | Third-harmonic | |
| (keV) | XPP | HOMS | (%) | ratio (%) |
| 6 | 2.71 ± 0.12 | 5.80 ± 0.58 | 57.7 ± 5.8 | 0.47 ± 0.07 |
| 7 | 3.35 ± 0.14 | 5.39 ± 0.54 | 61.5 ± 6.2 | 0.47 ± 0.04 |
| 8 | 3.51 ± 0.15 | 4.84 ± 0.48 | 61.0 ± 6.1 | 0.46 ± 0.08 |
| 9 | 3.45 ± 0.18 | 4.31 ± 0.43 | 65.3 ± 6.5 | 0.01 ± 0.04 |
| 9.5 | 3.42 ± 0.14 | 4.14 ± 0.41 | 66.0 ± 6.6 | 0.06 ± 0.04 |
Figure 3Flux measured simultaneously by the gas detector (GD) (squares) and the XGMD (circles) as a function of horizontal beam size, as defined by the horizontal slit gap. This measurement was performed at E X-ray = 6 keV. Solid and dotted lines are simulations of the transmission of a 2D Gaussian beam profile through a one-dimensional slit. The dashed line indicates the average value of the flux measured in the saturation region by the XGMD above ∼420 µm. The dashed area highlights the transmission loss from the HOMS related to its effective active length. The error bars originate from the inherent SASE intensity fluctuation of the beam.
Figure 4Summary of the outcome of the analysis showing the X-ray photon energy dependence of various parameters of the HOMS: the effective geometrical transmission [(a), squares], effective reflectivity [(b), circles] and total transmission [(c), triangles]. The theoretical effective reflectivity of the HOMS is indicated in (b) by the dashed line. Error bars of the squares are contained within the symbol size. The gray area surrounding circles and triangles indicates a total 10% error margin in transmission correction as a result of the uncertainty of the instrument optics (thickness of one diamond and one kapton window) and calibration error of the FEE GD (Moeller et al., 2011 ▸).
Figure 5Summary of the X-ray flux measurements for various X-ray photon energies as a function of silicon thickness used for beam attenuation. The measurements obtained by the XGMD and the radiometer are plotted in (a) and (b), respectively. The various solid, dashed and dotted lines are the result of simulations using equation (1) with a third- to first-harmonic ratio r 3 of 0, 0.1 and 1%, respectively. The solid red line indicates the fit to the data for each E X-ray, whose results are displayed in Fig. 6 ▸ and are summarized in Table 1 ▸. Each data set has been arbitrarily shifted in the vertical direction for clarity.
Figure 6First harmonic X-ray photon energy dependence of the third- to first-harmonic ratio r 3. The dashed line indicates the expected behavior of the HOMS if both angles had an incidence angle of 1.35 mrad. The solid line is a fit to the data and provides an incidence angle of 1.426 ± 0.01 mrad.
Components from the front-end enclosure to the detectors in the XPP instrument that can contribute to the transmission of the X-ray beam
| Distance from HOMS (m) | Name | Details | Nominal thickness (µm) | Measured thickness (µm) | Transmission at 6 keV (%) | Transmission at 9.5 keV (%) |
|---|---|---|---|---|---|---|
| −27.6 | FEE slit | – | – | – | – | – |
| −25.6 | FEE gas detector | – | – | – | – | – |
| −11.3 | HOMS M1 | – | – | – | – | – |
| 0 | HOMS M2 | – | – | – | – | – |
| 33.9 | IPM target | Si3N4 | 1 | – | 99.3 | 99.8 |
| 36.8 | Attenuator | Si | – | – | – | – |
| 36.2 | Exit window of nominal beamline | Diamond | 100 ± 10% | – | 69.2 | 91.3 |
| Laser in-coupling chamber | Vacuum | – | – | – | – | |
| Exit window of laser in-coupling | Kapton | 25.4 ± 5% | – | 95.0 | 98.7 | |
| Beam path | Air | – | 1.25 × 105 | 71.5 | 91.8 | |
| Entrance window of XGMD | Diamond | 100 | 109 ± 3 | 66.9 | 90.5 | |
| 37.2 | XGMD | – | – | – | – | – |
| Exit window of XGMD | Diamond | 100 | 111 ± 3 | 66.4 | 90.4 | |
| Beam path | Air | – | 1.71 × 105 | 95.5 | 98.8 | |
| Beam path window | Kapton | 128 ± 5% | – | 77.4 | 93.8 | |
| Beam path | Helium or vacuum | – | 1.52 × 106 | 99.3 | 99.6 | |
| Beam path window | Kapton | 52 ± 5% | – | 90.1 | 97.4 | |
| Beam path | Air | – | 9.94 × 105 | 76.6 | 99.1 | |
| Entrance window of radiometer | Diamond | 100 | 95 ± 3 | 70.4 | 91.7 | |
| 40.2 | Radiometer | – | – | – | – | – |