Flow analysis with digital subtraction angiography: 1. Description of a simplified flow model.

J R Hesselink, K H Chang, K J Chung, L Abbate.   

Abstract

An inexpensive, simplified flow model for use with digital subtraction angiographic (DSA) equipment is described. System tests show that the model gives reproducible data with a standard deviation of +/- 2.6%. Flow analysis also reveals an appropriate response to flow rates of 20-240 ml per min. The flow model can be used to investigate many of the variables and sources of error in DSA flow measurements.

Mesh:

Year:  1986        PMID: 3085447

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  1 in total

1.  Estimation of 2-D blood flow velocity map from cine-angiograms: algorithm using overlapping block set and illustration of vortex flow in abdominal aneurysm.

Authors:  N Sugimoto; C Uyama; T Sugahara; Y Yanagihara
Journal:  Med Biol Eng Comput       Date:  1994-07       Impact factor: 2.602