Literature DB >> 30851472

Experience of the New FlowGate2 Device as a Balloon Guide Catheter for Ischemic Stroke Intervention.

Ho Jun Yi1, Jae Hoon Sung2, Min Hyung Lee1, Dong Hoon Lee1.   

Abstract

BACKGROUND: We report the experience of the FlowGate2 (FG2) as a new balloon guide catheter in endovascular stroke intervention.
METHODS: We evaluated the various outcomes and complications of patients with intracranial large artery occlusion undergoing endovascular stroke intervention with FG2 at our center. Baseline characteristics (failure rate of device application, sex, age, risk factors, arterial occlusion sites, and time intervals) were reviewed. Outcomes were evaluated according to National Institutes of Health Stroke Scale score, modified Rankin Scale (mRS) score, number of stent passages required, and Thrombolysis in Cerebral Infarction score. The incidence of hemorrhage, vessel damage, distal emboli, and mortality rate were evaluated as indicators of complications.
RESULTS: Overall, 70 patients were enrolled, except the 2 patients with application failure of FG2. Seventy patients with a median age of 69 years were treated with FG2. Arterial occlusion involved the M1 (50%) and M2 (14.3%) segments, internal carotid artery (25.7%), and posterior circulation (10%). Median value of mRS at 90 days was 2.8, and 37 patients (52.8%) had a mRS score ≤2. The recanalization rate in patients with a Thrombolysis in Cerebral Infarction score of 2b or 3 was 91.4%. The hemorrhage rate was 5.7%, but none were symptomatic. In terms of complications, distal emboli occurred in 4.3% of cases.
CONCLUSIONS: Endovascular stroke intervention with the FG2 is safe and effective with good accessibility and less occurrence of distal emboli. Its trackability, stability, and luminal size make the FG2 suitable for stroke intervention.
Copyright © 2019 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Balloon guide catheter; Intervention; Stents; Stroke; Thrombectomy

Mesh:

Year:  2019        PMID: 30851472     DOI: 10.1016/j.wneu.2019.02.140

Source DB:  PubMed          Journal:  World Neurosurg        ISSN: 1878-8750            Impact factor:   2.104


  5 in total

1.  Balloon guide catheters: use, reject, or randomize?

Authors:  Mayank Goyal; Manon Kappelhof; Johanna M Ospel; Fouzi Bala
Journal:  Neuroradiology       Date:  2021-06-08       Impact factor: 2.804

2.  Efficacy of Asahi Fubuki as a Guiding Catheter for Mechanical Thrombectomy: An Institutional Case Series.

Authors:  L Rinaldo; H Cloft; W Brinjikji
Journal:  AJNR Am J Neuroradiol       Date:  2020-09-24       Impact factor: 3.825

3.  Effectiveness of Anchoring with Balloon Guide Catheter and Stent Retriever in Difficult Mechanical Thrombectomy for Large Vessel Occlusion.

Authors:  Ho Jun Yi; Bum-Tae Kim; Dong-Sung Shin
Journal:  J Korean Neurosurg Soc       Date:  2022-06-22

4.  The Long-Term Functional Effect of Thrombectomy on Patients with Middle Cerebral Artery Occlusion Who Exhibit Moderate to Severe Disability.

Authors:  Ho-Jun Yi; Dong-Hoon Lee; Bo-Young Hong; Seung-Yoon Song; Yeun-Jie Yoo; Mi-Jeong Yoon; Jae-Hoon Sung; Seong-Hoon Lim
Journal:  Medicina (Kaunas)       Date:  2021-05-19       Impact factor: 2.430

Review 5.  Advances in mechanical thrombectomy for acute ischaemic stroke from large vessel occlusions.

Authors:  Xu Guo; Zhongrong Miao
Journal:  Stroke Vasc Neurol       Date:  2021-07-20
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.