Literature DB >> 3084928

Buildup/surface dose and exit dose measurements for a 6-MV linear accelerator.

J A Purdy.   

Abstract

The central axis dose distribution in the buildup region for the Varian Clinac 6/100 6-MV x-ray beam was measured in a polystyrene phantom using a fixed volume (0.5 cm3) parallel-plate ionization chamber (2.4-mm plate separation). Results for the surface dose measurements ranged from approximately 8% of the maximum dose for a 5 X 5 cm field, up to 36% for a 40 X 40 cm field, 100-cm source-skin distance. The effect of a 0.6-cm-thick polycarbonate blocking tray and metal filters on the surface and buildup dose is also reported. In addition, ionization measurements were made to document the dose perturbations caused by the absence of backscattering material at the exit surface of a polystyrene phantom. Exit dose measurements showed a 15% reduction in dose with essentially no scattering material beyond the measurement point. Near full scatter condition could be restored by placing 5-10 mm (depending on field size) of unit density material directly behind the ion chamber's distal surface.

Entities:  

Mesh:

Year:  1986        PMID: 3084928     DOI: 10.1118/1.595908

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  Dose accumulation to assess the validity of treatment plans with reduced margins in radiotherapy of head and neck cancer.

Authors:  Nicholas J Lowther; Steven H Marsh; Robert J W Louwe
Journal:  Phys Imaging Radiat Oncol       Date:  2020-06-10

2.  Surface buildup dose dependence on photon field delivery technique for IMRT.

Authors:  Shigeru Yokoyama; Peter L Roberson; Dale W Litzenberg; Jean M Moran; Benedick A Fraass
Journal:  J Appl Clin Med Phys       Date:  2004-04-01       Impact factor: 2.102

3.  Verification of Treatment Planning Algorithms Using Optically Stimulated Luminescent Dosimeters in a Breast Phantom.

Authors:  Zhenia Gopalakrishnan; Raghuram K Nair; P Raghukumar; Sharika V Menon; Saju Bhasi
Journal:  J Med Phys       Date:  2018 Oct-Dec

4.  Evaluation of six TPS algorithms in computing entrance and exit doses.

Authors:  Yun I Tan; Mohamed Metwaly; Martin Glegg; Shaun Baggarley; Alex Elliott
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

5.  Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.

Authors:  Eric E Klein; Jacqueline Esthappan; Zuofeng Li
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

6.  Skin dose during radiotherapy: a summary and general estimation technique.

Authors:  Stephen F Kry; Susan A Smith; Rita Weathers; Marilyn Stovall
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.