Computer Simulations of Continuous Flow Peritoneal Dialysis Using the 3-Pore Model-A First Experience.

Carl M Öberg1, Giedre Martuseviciene2.   

Abstract

Background:Continuous flow peritoneal dialysis (CFPD) is performed using a continuous flux of dialysis fluid via double or dual-lumen PD catheters, allowing a higher dialysate flow rate (DFR) than conventional treatments. While small clinical studies have revealed greatly improved clearances using CFPD, the inability to predict ultrafiltration (UF) may confer a risk of potentially harmful overfill. Here we performed physiological studies of CFPD in silico using the extended 3-pore model.Method:A 9-h CFPD session was simulated for: slow (dialysate to plasma creatinine [D/P crea] < 0.6), fast (D/P crea > 0.8) and average (0.6 ≤ D/P crea ≤ 0.8) transporters using 1.36%, 2.27%, or 3.86% glucose solutions. To avoid overfill, we applied a practical equation, based on the principle of mass-balance, to predict the UF rate during CFPD treatment.
Results: Increasing DFR > 100 mL/min evoked substantial increments in small- and middle-molecule clearances, being 2 - 5 times higher compared with a 4-h continuous ambulatory PD (CAPD) exchange, with improvements typically being smaller for average and slow transporters. Improved UF rates, exceeding 10 mL/min, were achieved for all transport types. The β2-microglobulin clearance was strongly dependent on the UF rate and increased between 60% and 130% as a function of DFR. Lastly, we tested novel intermittent-continuous regimes as an alternative strategy to prevent overfill, being effective for 1.36% and 2.27%, but not for 3.86% glucose.
Conclusion: While we find substantial increments in solute and water clearance with CFPD, previous studies have shown similar improvements using high-volume tidal automated PD (APD). Lastly, the current in silico results need confirmation by studies in vivo.
Copyright © 2019 International Society for Peritoneal Dialysis.

Entities:  

Keywords:  Metabolic cost; automated peritoneal dialysis; dialysate flow rate

Mesh:

Substances:

Year:  2019        PMID: 30846606     DOI: 10.3747/pdi.2018.00225

Source DB:  PubMed          Journal:  Perit Dial Int        ISSN: 0896-8608            Impact factor:   1.756


  1 in total

1.  Novel Method for Osmotic Conductance to Glucose in Peritoneal Dialysis.

Authors:  Giedre Martus; Karin Bergling; Ole Simonsen; Eric Goffin; Johann Morelle; Carl M Öberg
Journal:  Kidney Int Rep       Date:  2020-09-19
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.