Literature DB >> 30835208

Implantable Multi-Modality Probe for Subdural Simultaneous Measurement of Electrophysiology, Hemodynamics, and Temperature Distribution.

Toshitaka Yamakawa, Takao Inoue, Masatsugu Niwayama, Fumiaki Oka, Hirochika Imoto, Sadahiro Nomura, Michiyasu Suzuki.   

Abstract

OBJECTIVE: The purpose of this paper is to demonstrate how the integration of the multi-channel measurement capabilities of near-infrared spectroscopy (NIRS), electrocorticography (ECoG), and negative temperature coefficient thermistor sensors into a single device compact enough for subdural implantation can provide beneficial information on various aspects of brain cortical activity and prove a powerful medical modality for pre-, intra-, and post-operative diagnoses in neurosurgery.
METHODS: The development of a flexible multi-modal multi-channel probe for the simultaneous measurement of the NIRS, ECoG, and surficial temperature obtained from the cerebral cortex was carried out. Photoelectric bare chips for NIRS channels, miniature temperature-coefficient thermistors for measuring localized temperature variation, and 3-mm-diameter platinum plates for ECoG recording were assembled on a polyimide-based flexible printed circuit to create six channels for each modality. A conformal coating of Parylene-C was applied on all the channels except the ECoG to make the probe surface biocompatible.
RESULTS: As a first-in-human study, the simultaneous measurement capability of the multi-modality probe, with sufficient signal-to-noise ratio and accuracy, to observe pathological neural activities in subjects during surgery and post-operative monitoring, with no complications two weeks since the implantation, was confirmed.
CONCLUSION: The feasibility of using a single device to assess the dynamic pathological activity from three different aspects was determined for human patients. SIGNIFICANCE: The simultaneous and accurate multi-channel recording of electrical, hemodynamic, and thermographic cortical activities in a single device small enough for subdural implantation is likely to have major implications in neurosurgery and neuroscience.

Entities:  

Mesh:

Year:  2019        PMID: 30835208     DOI: 10.1109/TBME.2019.2902189

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Diffuse Optical Tomography Using fNIRS Signals Measured from the Skull Surface of the Macaque Monkey.

Authors:  Ryusuke Hayashi; Okito Yamashita; Toru Yamada; Hiroshi Kawaguchi; Noriyuki Higo
Journal:  Cereb Cortex Commun       Date:  2021-11-10

2.  An Inkjet Printed Flexible Electrocorticography (ECoG) Microelectrode Array on a Thin Parylene-C Film.

Authors:  Yoontae Kim; Stella Alimperti; Paul Choi; Moses Noh
Journal:  Sensors (Basel)       Date:  2022-02-08       Impact factor: 3.576

3.  Epileptic seizure-related changes in electrocorticogram, cortical temperature, and cerebral hemodynamics obtained via an implantable multimodal multichannel probe during preoperative monitoring: illustrative case.

Authors:  Manami Wakuya; Takao Inoue; Hirochika Imoto; Yuich Maruta; Sadahiro Nomura; Michiyasu Suzuki; Toshitaka Yamakawa
Journal:  J Neurosurg Case Lessons       Date:  2022-03-07
  3 in total