| Literature DB >> 30834686 |
Shantel Lewis1, Tracey Pieterse1, Heather Lawrence1.
Abstract
INTRODUCTION: Digital radiography lacks visual clues of exposure techniques used to obtain radiographs, therefore manufacturers have included exposure indicators (EIs). EIs provides feedback about exposure techniques used and evaluating EIs will yield much needed information about exposure trends used in digital radiography.Entities:
Keywords: Computed radiography; detector dose indicators; digital radiography; exposure creep exposure indicator
Mesh:
Year: 2019 PMID: 30834686 PMCID: PMC6399191 DOI: 10.1002/jmrs.317
Source DB: PubMed Journal: J Med Radiat Sci ISSN: 2051-3895
Factors causing variations in the exposure indicator.25, 31, 37, 38, 39
| Factor | Explanation |
|---|---|
| Extraneous exposure information, including scatter | In computed radiography (CR) it is suggested that an optimal size CR cassette be used to eliminate extraneous exposure information. If a larger CR cassette must be used, the area of interest must be centred to the CR cassette with four‐sided collimation equally distributed from the cassette borders. With incorrect centring, the collimation field would have to be larger to include all the anatomy that is needed, resulting in a wider histogram |
| Exposure field recognition error | Occurs if collimation margins are not detected |
| Unexpected material in field | Widen histogram analysis due to densities that would usually not be included in look up table (LUT) |
| Extreme underexposure or overexposure | Histogram analysis error due to excessive quantum mottle or saturation |
| Delay in processing | CR cassettes must be moved away from the radiation area and be processed as soon as possible to prevent errors in the histogram analysis of the fog‐altered pixel values |
| Part selection from workstation menu | If the anatomical part selected from the workstation menu is incorrect, the LUT used to rescale the histogram data will be incorrect. The LUT for each anatomical part has a specific optimum gray‐scale and brightness level. |
Additional factors affecting exposure indicators.13
| Factor | Explanation |
|---|---|
| Patient gender | Female patients had a higher variant exposure indicator (EI) which is congruent with Lanca and Silva's |
| Time/date of exposure | X‐rays taken outside of normal working hours also showed an increase in the EI variations |
| Grid usage | When grids were used the EI obtained varied from that obtained when no grid was used for the same exam |
| Presence of implant or prosthesis | The presence of artefacts, implants or prostheses will widen the histogram and display a varied EI value |
Retrospective data collection criteria
| Date |
| Time |
| Age |
| Gender |
| Examination |
| Projection (anterior‐posterior/lateral/oblique, etc) |
| Presence of prosthesis/artefact (Yes/No) |
| Effective four‐sided collimation (before post‐processing (Yes/No) |
| Grid (Yes/No) |
| Any noise/saturation present |
| Repeated (Yes/No) |
| kVp given (if available) |
| Milliamperes second given (if available) |
| Actual exposure indicator (EI) |
| The manufacturer recommended EIs that were supplied to research sites (minimum and maximum) |
Factors affecting exposure indicators
| Factor | Actual exposure indicators within, lower or greater than manufacturer recommended range | ||||
|---|---|---|---|---|---|
| Chi‐square | |||||
| df |
| Pearson chi‐square value | Asymptotic significance (2‐sided) | Phi | |
| Normal or after hours | 2 | 1421 | 0.431 |
| 0.01 |
| Gender | 2 | 1420 | 1.89 |
| 0.03 |
| Examination | 50 | 1419 | 170.71 |
| 0.34 |
| Projection | 36 | 1421 | 76.38 |
| 0.23 |
| Presence of prosthesis | 2 | 1420 | 9.79 |
| 0.08 |
| Effective four sided collimation before post‐ processing | 2 | 1415 | 12.64 |
| 0.09 |
| Grid | 2 | 1419 | 6.75 |
| 0.06 |
Bold values indicate Significance was accepted if P < 0.05.