G S Sarkisov1, P D McChesney1, S Stein1, S Pronko1, H Bluem2, G Lednum2.
Abstract
To increase the sensitivity of optical interferometry for diagnostic of gas/plasma jet density with weak refractivity, we realize several multi-pass probing setups with the visualization field. These optical setups are based on a modified Michelson interferometer with multi-pass and multi-angle propagation of the probing beam through the transparent inhomogeneity. In comparison to the standard single-pass Mach-Zehnder setup, the multi-pass interferometer allows for an increase in the measurement sensitivity of up to 8 times. This technique can be applied to a transparent axially symmetrical gas/plasma jet with low refractivity.Year: 2019 PMID: 30831685 DOI: 10.1063/1.5081830
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523