Literature DB >> 30803304

Investigation of load transfer process between external fixator and bone model by experimental and finite element methods.

Xia Zhao1, Jianfeng Li1, Ying Chen1, Chunjing Tao2, Run Ji2.   

Abstract

INTRODUCTION: : Unilateral external fixators are widely used in orthopedics to stabilize fractured bones and to treat limb deformities. One of the main problems is that it is difficult to detect healing status. In addition, whether load transfer progress between the fixator and bone model are the same under axial, torsional, and bending loads has not been studied.
METHODS: : Therefore the main purpose of this study was to detect the load transfer process between the fixator and a bone model by measuring strains on the fixator-bone system during four healing states using experimental and finite element methods. In the experimental method, 20 strain gauges were used to measure strain on the fixator and bone model under three load conditions. Polyacetal slice models with different material properties were used to simulate the callus model during four growth states.
RESULTS: : The results indicate that strain on the bone model increased and strain on the fixator parts decreased with maturation of the callus under axial, bending, and torsional loads. Although all curves showed a similar changing trend, they were slightly different under the three loads. DISCUSSION AND
CONCLUSIONS: : This study provides a useful method to monitor the fracture healing process, and identifies the healing endpoint, detects healing status, and provides useful information for the orthopedist.

Entities:  

Keywords:  Experimental and finite element methods; healing states; load transfer process; unilateral external fixators

Mesh:

Year:  2019        PMID: 30803304     DOI: 10.1177/2280800019826512

Source DB:  PubMed          Journal:  J Appl Biomater Funct Mater        ISSN: 2280-8000            Impact factor:   2.604


  3 in total

1.  Osteosynthesis Metal Plate System for Bone Fixation Using Bicortical Screws: Numerical-Experimental Characterization.

Authors:  Andrea A R Olmos; Aureliano Fertuzinhos; Teresa D Campos; Isabel R Dias; Carlos A Viegas; Fábio A M Pereira; Nguyễn T Quyền; Marcelo F S F de Moura; Andrea Zille; Nuno Dourado
Journal:  Biology (Basel)       Date:  2022-06-20

Review 2.  An engineering review of external fixators.

Authors:  P L N Fernando; Aravinda Abeygunawardane; Pci Wijesinghe; Parakrama Dharmaratne; Pujitha Silva
Journal:  Med Eng Phys       Date:  2021-11-04       Impact factor: 2.242

3.  Direct electromagnetic coupling to determine diagnostic bone fracture stiffness.

Authors:  Jakob G Wolynski; Milan M Ilić; Kevin M Labus; Branislav M Notaroš; Christian M Puttlitz; Kirk C McGilvray
Journal:  Ann Transl Med       Date:  2022-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.