Kezhong Wang1, Christopher L Frewin2, Dorna Esrafilzadeh3, Changchun Yu1, Caiyun Wang1, Joseph J Pancrazio2, Mario Romero-Ortega2, Rouhollah Jalili4, Gordon Wallace1.
Abstract
Fabrication of flexible and free-standing graphene-fiber- (GF-) based microelectrode arrays with a thin platinum coating, acting as a current collector, results in a structure with low impedance, high surface area, and excellent electrochemical properties. This modification results in a strong synergistic effect between these two constituents leading to a robust and superior hybrid material with better performance than either graphene electrodes or Pt electrodes. The low impedance and porous structure of the GF results in an unrivalled charge injection capacity of 10.34 mC cm-2 with the ability to record and detect neuronal activity. Furthermore, the thin Pt layer transfers the collected signals along the microelectrode efficiently. In vivo studies show that microelectrodes implanted in the rat cerebral cortex can detect neuronal activity with remarkably high signal-to-noise ratio (SNR) of 9.2 dB in an area as small as an individual neuron.Entities:
Keywords: brain-machine interface; graphene fiber; neural interface; neural stimulation and recording; soft neural microelectrode
Year: 2019 PMID: 30803072 DOI: 10.1002/adma.201805867
Source DB: PubMed Journal: Adv Mater ISSN: 0935-9648 Impact factor: 30.849