Abstract

INTRODUCTION: Bone marrow biopsies are a key diagnostic and monitoring intervention in haematology with manual bone marrow techniques the established method of choice. Powered biopsy devices are now available, but are not widely used in haematology. This study compared the quality of bone marrow trephines obtained with the Jamshidi needle and OnControl powered drill system.
METHODS: Retrospective analysis was undertaken on trephine samples prior to and after implementation of the OnControl drill system. Trephine size and quality were assessed independently by three pathologists and compared between techniques and operators using nonparametric tests.
RESULTS: There were 164 samples assessed (Jamshidi n = 69, OnControl, same site as aspirate n = 48, OnControl, separate site from aspirate n = 47). The assessable and total length were similar between the Jamshidi and OnControl techniques, with increased crush artefact observed with the OnControl drill (P < 0.001). Using a separate puncture site for trephine collection and aspirate did not reduce the artefact seen with the OnControl system (P = 0.274). Smaller samples (P < 0.001) and an increase in crushed (P = 0.009) and connective tissue (P = 0.002) were seen in trephines obtained by nonlaboratory-based trainees, regardless of the needle used or their stage of training, compared to laboratory trainees.
CONCLUSIONS: Trephines obtained by either method had similar assessable areas. The OnControl system was associated with more artefact, a finding in line with previous studies. There was no improvement by sampling the trephine from a separate site to the aspirate. Laboratory-based trainees who reviewed marrow morphology produced trephines with better assessable length than those not based in the laboratory.
© 2019 John Wiley & Sons Ltd.

Entities:  

Keywords:  Jamshidi needle; OnControl bone marrow system; bone marrow biopsy; powered biopsy; quality

Mesh:

Year:  2019        PMID: 30779423     DOI: 10.1111/ijlh.12984

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  2 in total

1.  Comparative analysis of CT guided vertebral biopsy by a conventional bone biopsy needle versus bone biopsy needle with acquisition cradle.

Authors:  Dharmendra Kumar Singh; Tankeshwar Boruah; Anuradha Sharma; Geetika Khanna; Loveneesh G Krishna; Nishith Kumar
Journal:  J Clin Orthop Trauma       Date:  2021-06-02

2.  Bone marrow particle enrichment analysis for the laboratory diagnosis of multiple myeloma: A case study.

Authors:  Congming Zhang; Yanan Zhang
Journal:  J Clin Lab Anal       Date:  2020-06-16       Impact factor: 2.352

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.